Potential errors in axial alignment using intramedullary instrumentation for total knee arthroplasty

被引:40
作者
Nuño-Siebrecht, N [1 ]
Tanzer, M [1 ]
Bobyn, JD [1 ]
机构
[1] Univ Bologna, Fac Engn, DEIS, Bologna, Italy
关键词
total knee arthroplasty; alignment; intramedullary instrumentation;
D O I
10.1016/S0883-5403(00)90352-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Accurate restoration of normal limb alignment is crucial for the long-term survivorship of total knee arthroplasty (TKA). A mathematical model was used to evaluate the maximum error in varus and valgus alignment that could occur when cutting the tibia or femur during TKA using intramedullary (IM) guides of varying length and diameter. Minor deviations in the insertion point of IM instrumentation during TKA can result in malalignment of several degrees. This error can be minimized by careful attention to the entry point of the IM instrumentation or by increasing the IM rod diameter and length used during primary TKA.
引用
收藏
页码:228 / 230
页数:3
相关论文
共 3 条
[1]  
BERTIN KC, 1989, CONTROVERSIES TOTAL, P175
[2]  
ECKHOFF DG, 1995, CLIN ORTHOP RELAT R, P176
[3]  
THORNHILL TS, 1991, BRIS MYER Z, P7