Comparison of the radioiodinated serum albumin (RISA) dilution technique with direct volumetric measurements in animal models of peritoneal dialysis

被引:0
|
作者
Van Biesen, W
Devriese, AS
Carlsson, O
Van Landschoot, M
Dheuvaert, T
Lameire, NH
机构
[1] State Univ Ghent Hosp, Div Renal, Dept Internal Med, B-9000 Ghent, Belgium
[2] Lund Univ, Dept Nephrol & Physiol, Lund, Sweden
来源
PERITONEAL DIALYSIS INTERNATIONAL | 2002年 / 22卷 / 03期
关键词
animal model; transport characteristics; volume measurements;
D O I
暂无
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: Rat models of peritoneal dialysis (PD) are useful for studying the physiology of peritoneal transport and evaluating new osmotic agents. Intraperitoneal (IP) solute concentrations and their evolution over time are easy to measure, but IP volume (IPV) is not. Direct volumetric measurements are the "gold standard;' but they are expensive and do not allow for repetitive measurements in the same animal. The indicator dilution technique is therefore used as an alternative. However, that technique is based on assumptions that are not always valid. The present study compares direct volume measurement with the indicator dilution technique [radioiodinated serum albumin (RISA)] to determine the IPV over time curves in a rat model of PD. Methods: In series 1, 17 Sprague-Dawley rats were instilled IP with 25 mL 1.36% glucose dialysate through a Teflon catheter. In 9 animals, 0.35 mL dialysate was sampled and discarded at time points 0, 3,15, 30, 60,180, and 240 minutes. In the other 8 animals, no sampling was performed. At 240 minutes, all 12 animals were humanely killed, and direct volumetric measurements of IPV were performed. In series 2, rats were instilled IP with 25 mL 1.36% glucose dialysate containing 18.5 kBq I-131 RISA. In 9 animals, dialysate was sampled at 0, 3, 15, 30, 60, 90, 120,180, and 240 minutes for the construction of the RISA concentration-over-time curve, and to calculate the elimination constant K-e. At 30, 60, 180, and 240 minutes, dialysate was sampled in 6 different animals (total n = 24) to calculate IPV using the RISA dilution technique. Immediately afterward, the animals were humanely killed, and direct volumetric measurements of IPV were performed. Results: In series 1, after 240 minutes' dwell time, the IPV was lower in the sampled animals as compared with the non sampled animals (27.11 +/- 1.85 mL vs 30.75 +/- 0.59 mL, p = 0.001). In series 2, the evolution of RISA activity in the dialysate over time was described by piece-wise linear regression, yielding 3288 - 8.2T counts (cts) for T < 52.72 minutes and 2973 - 1.99T counts for T > 52.72 minutes. The IPV was better predicted with a K-e that took into account the disappearance of RISA by sampling than with a K-e that took into account disappearance of RISA only by absorption. Conclusions: If indicator dilution techniques are used to measure IPV, attention must be paid to the disappearance of the osmotic agent and the marker by multiple sampling. The best way to meet that goal is to use micropipettes to avoid large sample volumes.
引用
收藏
页码:316 / 322
页数:7
相关论文
共 7 条
  • [1] Comparison of the radioiodinated serum albumin (RISA) dilution technique with direct volumetric measurements in animal models of peritoneal dialysis. (vol 22, pg 316, 2002)
    Van Biesen, W
    De Vriese, AS
    Carlsson, O
    Van Landschoot, M
    Dheuvaert, T
    Lameire, NH
    PERITONEAL DIALYSIS INTERNATIONAL, 2002, 22 (06): : 739 - 739
  • [2] INTRAPERITONEAL FLUID VOLUME CHANGES DURING PERITONEAL-DIALYSIS IN THE RAT - INDICATOR DILUTION VS VOLUMETRIC MEASUREMENTS
    ZAKARIA, E
    RIPPE, B
    BLOOD PURIFICATION, 1995, 13 (05) : 255 - 270
  • [3] Comparison of Transpulmonary Thermodilution and Ultrasound Dilution Technique: Novel Insights into Volumetric Parameters from an Animal Model
    Martin Boehne
    Florian Schmidt
    Lars Witt
    Harald Köditz
    Michael Sasse
    Robert Sümpelmann
    Harald Bertram
    Armin Wessel
    Wilhelm Alexander Osthaus
    Pediatric Cardiology, 2012, 33 : 625 - 632
  • [4] Comparison of Transpulmonary Thermodilution and Ultrasound Dilution Technique: Novel Insights into Volumetric Parameters from an Animal Model
    Boehne, Martin
    Schmidt, Florian
    Witt, Lars
    Koeditz, Harald
    Sasse, Michael
    Suempelmann, Robert
    Bertram, Harald
    Wessel, Armin
    Osthaus, Wilhelm Alexander
    PEDIATRIC CARDIOLOGY, 2012, 33 (04) : 625 - 632
  • [5] COMPARISON OF TRANSPULMONARY THERMODILUTION AND ULTRASOUND DILUTION TECHNIQUE: NOVEL INSIGHTS IN VOLUMETRIC PARAMETERS FROM A JUVENILE ANIMAL MODEL
    Boehne, M.
    Schmidt, F.
    Witt, L.
    Koeditz, H.
    Suempelmann, R.
    Wessel, A.
    Sasse, M.
    Osthaus, W. A.
    INTENSIVE CARE MEDICINE, 2011, 37 : S337 - S337
  • [6] AMNIOTIC-FLUID VOLUME ASSESSMENT - COMPARISON OF ULTRASONOGRAPHIC ESTIMATES VERSUS DIRECT MEASUREMENTS WITH A DYE-DILUTION TECHNIQUE IN HUMAN-PREGNANCY
    DILDY, GA
    LIRA, N
    MOISE, KJ
    RIDDLE, GD
    DETER, RL
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1992, 167 (04) : 986 - 994
  • [7] COMPARISON OF SERUM-CALCIUM MEASUREMENTS WITH RESPECT TO 5 MODELS OF ATOMIC-ABSORPTION SPECTROMETERS USING NBS-AACC CALCIUM REFERENCE METHOD AND ISOTOPE-DILUTION MASS-SPECTROMETRY AS DEFINITIVE METHOD
    COPELAND, BE
    GRISLEY, DW
    CASELLA, J
    BAILEY, H
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1976, 66 (04) : 619 - 633