Effects of ProTaper Universal, ProTaper Next, and HyFlex Instruments on Crack Formation in Dentin

被引:74
作者
Capar, Ismail Davut [1 ]
Arslan, Hakan [3 ]
Akcay, Merve [2 ]
Uysal, Banu [1 ]
机构
[1] Izmir Katip Celebi Univ, Fac Dent, Dept Endodont, TR-35620 Izmir, Turkey
[2] Izmir Katip Celebi Univ, Fac Dent, Dept Pedodont, TR-35620 Izmir, Turkey
[3] Ataturk Univ, Fac Dent, Dept Endodont, Erzurum, Turkey
关键词
Cracks; controlled memory; dentinal damage; ProTaper Next; ProTaper Universal; root canal instrumentation; root fracture; rotary nickel-titanium instruments; NITI ROTARY INSTRUMENTS; SELF-ADJUSTING FILE; CANAL PREPARATION; DEFECTS; HAND;
D O I
10.1016/j.joen.2014.02.026
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The aim of the present study was to investigate the incidence of cracks in root dentin after root canal preparation with ProTaper Next (Dentsply Maillefer, Ballaigues, Switzerland), HyFlex (Coltene-Whaledent, Allstetten, Switzerland), and ProTaper Universal (Dentsply Maillefer) rotary instruments. Methods: One-hundred mandibular premolars were selected. Twenty-five teeth were left unprepared and served as a negative control; another 25 teeth were in-strumented with the ProTaper Universal system up to size F4 as a positive control, and the remaining 50 teeth were shaped with the following experimental groups with an apical size 40 file: ProTaper Next X4 and HyFlex 40/0.4. After root canal preparation, all of the roots were sectioned perpendicular to the long axis at 2, 4, 6, and 8 mm from the apex, and the sections were then observed under a stereomicroscope. The absence/presence of cracks was recorded, and the data were analyzed with a chi-square test. The significance level was set at P = .05. Results: No cracks were observed in the negative control group. Vertical root fractures were not observed in any of the groups. The ProTaper Next and HyFlex instruments caused fewer cracks (28%) than the ProTaper Universal instrument (56%) (P < .05). However, there were no significant differences in crack formation between the ProTaper Next and HyFlex groups (P > .05). Conclusions: Within the limitations of this in vitro study, all of the instrumentation systems used in this study created cracks in the root dentin. The ProTaper Next and HyFlex instruments tended to cause fewer dentinal cracks compared with the ProTaper Universal instrument.
引用
收藏
页码:1482 / 1484
页数:3
相关论文
共 19 条
  • [1] Ashwinkumar V., 2013, Int Endod J
  • [2] Vertical Root Fractures and Dentin Defects: Effects of Root Canal Preparation, Filling, and Mechanical Cycling
    Barreto, Mirela Sangoi
    Moraes, Rafael do Amaral
    da Rosa, Ricardo Abreu
    Cunha Moreira, Carlos Heitor
    Reis So, Marcus Vinicius
    Souza Bier, Carlos Alexandre
    [J]. JOURNAL OF ENDODONTICS, 2012, 38 (08) : 1135 - 1139
  • [3] Smooth flexible versus active tapered shaft design using NiTi rotary instruments
    Bergmans, L
    Van Cleynenbreugel, J
    Beullens, M
    Wevers, M
    Van Meerbeek, B
    Lambrechts, P
    [J]. INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (10) : 820 - 828
  • [4] Capar ID, 2014, J ENDOD
  • [5] Gambarini G., 2005, ENDOD TOPICS, V10, P179, DOI DOI 10.1111/j.1601-1546.2005.00119.x
  • [6] Effects of Self-Adjusting File, Mtwo, and Pro Taper on the Root Canal Wall
    Hin, Ellemieke S.
    Wu, Min-Kai
    Wesselink, Paul R.
    Shemesh, Hagay
    [J]. JOURNAL OF ENDODONTICS, 2013, 39 (02) : 262 - 264
  • [7] Potential Relationship between Design of Nickel-Titanium Rotary Instruments and Vertical Root Fracture
    Kim, Hyeon-Cheol
    Lee, Min-Ho
    Yum, Jiwan
    Versluis, Antheunis
    Lee, Chan-Joo
    Kim, Byung-Min
    [J]. JOURNAL OF ENDODONTICS, 2010, 36 (07) : 1195 - 1199
  • [8] Fracture strength of tooth roots following canal preparation by hand and rotary instrumentation
    Lam, PPS
    Palamara, JEA
    Messer, HH
    [J]. JOURNAL OF ENDODONTICS, 2005, 31 (07) : 529 - 532
  • [9] The Incidence of Root Microcracks Caused by 3 Different Single-file Systems versus the ProTaper System
    Liu, Rui
    Hou, Ben Xiang
    Wesselink, Paul R.
    Wu, Min-Kai
    Shemesh, Hagay
    [J]. JOURNAL OF ENDODONTICS, 2013, 39 (08) : 1054 - 1056
  • [10] Incidence of Apical Root Cracks and Apical Dentinal Detachments after Canal Preparation with Hand and Rotary Files at Different Instrumentation Lengths
    Liu, Rui
    Kaiwar, Anjali
    Shemesh, Hagay
    Wesselink, Paul R.
    Hou, Benxiang
    Wu, Min-Kai
    [J]. JOURNAL OF ENDODONTICS, 2013, 39 (01) : 129 - 132