Simple transient extension chamber to permit full mass scans with electrothermal vaporization inductively coupled plasma mass spectrometry

被引:16
作者
Langer, D [1 ]
Holcombe, JA [1 ]
机构
[1] Univ Texas, Dept Chem & Biochem, Austin, TX 78712 USA
关键词
ETV-ICPMS; electrothermal vaporizer; mass scanning;
D O I
10.1366/0003702991945489
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The design for an in-line transient extension (TEx) chamber was developed to provide a simple means of lengthening an electrothermal vaporizer (ETV) signal for the purpose of obtaining a full mass scan from a single ETV firing with inductively coupled plasma mass spectrometry (ICP-MS) detection. The TEx chamber is simply a flask (e.g., 100-500 mt) with a mated joint allowing the ETV-generated aerosol to enter the chamber in a pulse and exit the chamber exponentially diluted. Characterization of the TEx chamber was completed with the use of the following metals: Ba, Be, Ce, Co, In, Mg, Pb, Th, and Tl. A decrease in peak area of 20 +/- 4% was witnessed when the TEx chamber was put in-line between the ETV and the ICP-MS for all metals tested. Comparison of the peak areas obtained with and without the addition of 10 mu L of 100 ppm NaCl to the sample without the TEx chamber in-line resulted in increases from 8 to 40%, depending on the metal. Repetition of the experiment with the TEx chamber in line showed no appreciable difference in results. This outcome indicates that dilution in the TEx chamber does not influence the behavior (i.e., enhancement effect) of the carrier. Interrupting the carrier gas flow for up to 2 min after the TEx chamber was "loaded" with the aerosol caused decreases in peak area from 25 to 35%, depending on the length of time the sample was held in the chamber. This result is most likely due to settling/loss of the aerosol particles in the chamber. Addition of 10 mu L of 100 ppm NaCl to the sample showed a decrease in peak area of 30-43% when the sample was held for up to 2 min in the chamber, slightly higher than when no NaCl was present. A comparison of full mass scans taken with a concentric nebulizer and the ETV with an in-line TEx chamber using identical data collection rates and times showed that the TEx chamber can be used successfully as a method for both qualitative and quantitative identification of trace elements in an unknown solution. Detection limits for the nine metals tested with the TEx chamber in line are given.
引用
收藏
页码:1244 / 1250
页数:7
相关论文
共 6 条
[1]   INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY USING CONTINUUM FLOW ION EXTRACTION [J].
GRAY, AL ;
DATE, AR .
ANALYST, 1983, 108 (1290) :1033-1050
[2]  
Gregoire DC, 1997, CAN J ANAL SCI SPECT, V42, P1
[3]   Thermophoretic collection and analysis of submicrometer Ag particles emitted from a graphite tube-type electrothermal vaporizer [J].
Langer, D ;
Holcombe, JA .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :582-588
[4]  
MAHONEY PP, 1999, IN PRESS ANAL CHEM
[5]   The contribution of inductively coupled plasma mass spectrometry to biomedical research [J].
Patriarca, M .
MICROCHEMICAL JOURNAL, 1996, 54 (03) :262-271
[6]   PARTICLE-SIZE DISTRIBUTION OF SAMPLE TRANSPORTED FROM AN ELECTROTHERMAL VAPORIZER TO AN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER [J].
SPARKS, CM ;
HOLCOMBE, J ;
PINKSTON, TL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (13) :1607-1615