Investigation of oesophageal photoplethysmographic signals and blood oxygen saturation measurements in cardiothoracic surgery patients

被引:49
作者
Kyriacou, PA
Powell, S
Langford, RM
Jones, DP
机构
[1] Univ London Queen Mary & Westfield Coll, Dept Engn, London E1 4NS, England
[2] St Bartholomews Hosp Barts & The London NHS Trust, Anaesthet Lab, London EC1A 7BE, England
关键词
photoplethysmography (PPG); oesophagus; blood oxygen saturation; cardiac surgery;
D O I
10.1088/0967-3334/23/3/305
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Pulse oximeter probes attached to the finger may fail to estimate blood oxygen saturation (SpO(2)) in patients with compromised peripheral perfusion (e.g. hypothermic cardiopulmonary bypass surgery). The measurement of SpO(2) from a central organ such as the oesophagus is suggested as an alternative to overcome this problem. A reflectance oesophageal pulse oximeter probe and a processing system implemented in LabVIEW were developed. The system was evaluated in clinical measurements on 50 cardiothoracic surgery patients. Oesophageal photoplethysmographic (PPG) signals with large amplitudes and high signal-to-noise ratios were measured from various depths within the oesophagus from all the cardiothoracic patients. The oesophageal PPG amplitudes from these patients were in good agreement with previous oesophageal PPG amplitude measurements from healthy anaesthetized patients. The oesophageal pulse oximeter SpO(2) results agreed well with the estimated arterial oxygen saturation (SaO(2)) values inferred from the oxygen tension obtained by blood gas analysis. The mean (+/-SD) of the differences between the oesophageal pulse oximeter SpO(2) readings and those from blood gas analysis was 0.02 +/- 0.88%. Also, the oesophageal pulse oximeter was found to be reliable and accurate in five cases of poor peripheral perfusion when a commercial finger pulse oximeter probe failed to estimate oxygen saturation values for at least 10 min. These results suggest that the arterial blood circulation to the oesophagus is less subject to vasoconstriction and decreased PPG amplitudes than are the peripheral sites used for pulse oximetry such as the finger. It is concluded that oesophageal SpO(2) monitoring may be of clinical value.
引用
收藏
页码:533 / 545
页数:13
相关论文
共 13 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]  
CHALLONER AVJ, 1979, NONINVASIVE PHYSL ME, P125
[3]   A COMPARISON OF THE PERFORMANCE OF 20 PULSE OXIMETERS UNDER CONDITIONS OF POOR PERFUSION [J].
CLAYTON, DG ;
WEBB, RK ;
RALSTON, AC ;
DUTHIE, D ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (01) :3-10
[4]   PHOTO-ELECTRIC PLETHYSMOGRAPHY AS A MONITORING DEVICE IN ANESTHESIA - APPLICATION AND INTERPRETATION [J].
DORLAS, JC ;
NIJBOER, JA .
BRITISH JOURNAL OF ANAESTHESIA, 1985, 57 (05) :524-530
[5]   PHOTOPLETHYSMOGRAPHIC EVALUATION OF THE RELATIONSHIP BETWEEN SKIN REFLECTANCE AND SKIN BLOOD-VOLUME [J].
HIGGINS, JL ;
FRONEK, A .
JOURNAL OF BIOMEDICAL ENGINEERING, 1986, 8 (02) :130-136
[6]   A system for investigating oesophageal photoplethysmographic signals in anaesthetised patients [J].
Kyriacou, PA ;
Moye, AR ;
Gregg, A ;
Choi, DMA ;
Langford, RM ;
Jones, DP .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (05) :639-643
[7]   Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation [J].
Kyriacou, PA ;
Moye, AR ;
Choi, DMA ;
Langford, RM ;
Jones, DP .
PHYSIOLOGICAL MEASUREMENT, 2001, 22 (01) :223-232
[8]  
LINDGREN L, 1991, HARVARD LIBR BULL, V2, P7
[9]   NONINVASIVE PULSE OXIMETRY UTILIZING SKIN REFLECTANCE PHOTOPLETHYSMOGRAPHY [J].
MENDELSON, Y ;
OCHS, BD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) :798-805
[10]  
NIJBOER J A, 1981, Clinical Physics and Physiological Measurement, V2, P205, DOI 10.1088/0143-0815/2/3/004