New valves - New dangers? 22 valves (38 probes) designed in the 'nineties in ultralong-term tests (365 days)

被引:7
作者
Oikonomou, J [1 ]
Aschoff, A [1 ]
Hashemi, B [1 ]
Kunze, S [1 ]
机构
[1] Univ Heidelberg, Dept Neurosurg, D-69120 Heidelberg, Germany
关键词
bench-test; cerebrospinal fluid; cerebrospinal fluid shunts; gravitational valves; hydrocephalus; CSF overdrainage;
D O I
10.1055/s-2008-1072307
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
In the nineties, more than 30 valves have been launched, most are "Clones", but some introduce new concepts (Diamond, Siphon-Guard, PaediGAV). Of the 22 designs we tested. 20 have no previous tests and 3 are low-cost valves from Mexico (Dewimed), Zimbabwe (Harare-shunt) and England (Nottingham Shunt). 27 were tested using non-stop-perfusions for 365 days, 8 for 168-294 and 3 for 2 days. Results: Ranking of the mean relative deviation: adjustable valves 19%, gravitational 34%, simple ball 51%, diaphragm 73%, distal slit 74% and autoregulating valves 121%. Ranking of physiological flow properties: gravitational > Equiflow(TM) > autoregulating > adjustable > ball > diaphragm > distal slit. Safety: The Equiflow were highly susceptible to external pressure. Conclusion: Most valves designed in the nineties showed inaccuracy, long-term-drifts, safety deficits and hydraulic mismanagement similar to more than 400 previously tested probes. With regard to accuracy and drift the ball designs were superior, but simple and adjustable ball valves are at risk of overdrainage. The gravitational ball valves (Cordis GCA, Miethke valves) showed the closest relation to physiological flow requirements.
引用
收藏
页码:23 / 26
页数:4
相关论文
共 10 条
[1]  
ASCHOFF A, 1995, THESIS U HEIDELBERG
[2]   Hydrodynamic properties of hydrocephalus shunts: United Kingdom shunt evaluation laboratory [J].
Czosnyka, M ;
Czosnyka, Z ;
Whitehouse, H ;
Pickard, JD .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1997, 62 (01) :43-50
[3]  
EKSTEDT J, 1980, INTRACRANIAL PRESSUR, V4, P483
[4]  
FRUH K, 1994, THESIS U HEIDELBERG
[5]  
HASHEMI B, 1996, EUR J PEDIAT SURG S, V1
[6]  
KEITH HD, 1981, 223795064 USPH FDA U
[7]  
RICHARD KE, 1989, 32832 RI DFG ABSCHL
[8]  
Schubert W, 1979, Z Exp Chir, V12, P379
[9]  
SCHULTE C, 1996, CHILD NERV SYST, V12, P503
[10]   TESTING THE HYDROCEPHALUS SHUNT VALVE - LONG-TERM BENCH TEST-RESULTS OF VARIOUS NEW AND EXPLANTED VALVES - THE NEED FOR A MODEL FOR TESTING VALVES UNDER PHYSIOLOGICAL CONDITIONS [J].
TROST, HA ;
HEISSLER, HE ;
CLAUSSEN, G ;
GAAB, MR .
EUROPEAN JOURNAL OF PEDIATRIC SURGERY, 1991, 1 :38-40