Comparison of three commercially available ektacytometers with different shearing geometries

被引:73
作者
Baskurt, Oguz K. [1 ]
Hardeman, M. R. [2 ]
Uyuklu, Mehmet [1 ]
Ulker, Pinar [1 ]
Cengiz, Melike [1 ]
Nemeth, Norbert [3 ]
Shin, Sehyun [4 ]
Alexy, Tamas [5 ]
Meiselman, Herbert J. [5 ]
机构
[1] Akdeniz Univ, Fac Med, Dept Physiol, TR-07058 Antalya, Turkey
[2] Univ Amsterdam, Acad Med Ctr, Dept Physiol, NL-1105 AZ Amsterdam, Netherlands
[3] Univ Debrecen, Inst Surg, Dept Operat Tech & Surg Res, Med & Hlth Sci Ctr, Debrecen, Hungary
[4] Korea Univ, Dept Mech Engn, Seoul, South Korea
[5] Univ So Calif, Keck Sch Med, Dept Physiol & Biophys, Los Angeles, CA 90033 USA
关键词
Erythrocyte deformability; ektacytometry; BLOOD-CELL DEFORMABILITY; ERYTHROCYTE DEFORMABILITY; LASER DIFFRACTION; ANALYZER LORCA; INSTRUMENT;
D O I
10.3233/BIR-2009-0536
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In December 2008, the International Society for Clinical Hemorheology organized a workshop to evaluate and compare three ektacytometer instruments for measuring deformability of red blood cells (RBC): LORCA (Laser-assisted Optical Rotational Cell Analyzer, RR Mechatronics, Hoorn, The Netherlands), Rheodyn SSD (Myrenne GmbH, Roetgen, Germany) and RheoScan-D (RheoMeditech, Seoul, Korea). Intra-assay reproducibility and biological variation were determined using normal RBC, and cells with reduced deformability (i.e., 0.001-0.02% glutaradehyde (GA), 48 degrees C heat treatment) were employed as either the only RBC present or as a sub-population. Standardized difference values were used as measure of the power to detect differences between normal and treated cells. Salient results include: (1) All instruments had intra-assay variations below 5% for shear stress (SS) > 1 Pa but a sharp increase was found for Rheodyn SSD and RheoScan-D at lower SS; (2) Biological variation was similar and markedly increased for SS < 3-5 Pa; (3) All instruments detected GA-treated RBC with maximal power at 1-3 Pa, the presence of 10% or 40% GA-modified cells, and the effects of heat treatment. It is concluded that the LORCA, Rheodyn SSD and RheoScan-D all have acceptable precision and power for detecting reduced RBC deformability due to GA treatment or heat treatment, and that the SS range selected for the measurement of deformability is an important determinant of an instrument's power.
引用
收藏
页码:251 / 264
页数:14
相关论文
共 15 条
[1]  
Baskurt OK, 1996, CLIN HEMORHEOL, V16, P753
[2]  
Baskurt OK, 2000, CLIN HEMORHEOL MICRO, V23, P43
[3]   Cellular determinants of low-shear blood viscosity [J].
Baskurt, OK ;
Meiselman, HJ .
BIORHEOLOGY, 1997, 34 (03) :235-247
[4]  
Bohren C.F., 1983, Absorption and Scattering of Light by Small Particles
[5]  
Cohen J., 1988, Statistical power analysis for the behavioural sciences, V2nd
[6]  
Dobbe JGG, 2004, BIORHEOLOGY, V41, P65
[7]  
Hardeman MR, 2007, BIOMED HEALTH RES, V69, P242
[8]  
HARDEMAN MR, 1994, CLIN HEMORHEOL, V14, P605
[9]  
HARDEMAN MR, 1994, CLIN HEMORHEOL, V14, P619
[10]  
JOHNSON RM, 1989, METHOD ENZYMOL, V173, P35