Chemical shift imaging at 3 Tesla: effect of echo time on assessing bone marrow abnormalities

被引:20
作者
Del Grande, F. [1 ,2 ]
Subhawong, Ty [3 ]
Flammang, A. [4 ]
Fayad, L. M. [1 ]
机构
[1] Johns Hopkins Med Inst, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
[2] Osped Reg Lugano, Serv Radiol, CH-6900 Lugano, TI, Switzerland
[3] Univ Miami, Miller Sch Med, Dept Radiol, Miami, FL 33136 USA
[4] Siemens Med Solut USA Inc, Malvern, PA 19355 USA
关键词
Echo time; Bone marrow; Chemical shift imaging; 3 Tesla MRI; COMPRESSION FRACTURES; ADRENAL MASSES; IN-PHASE; BENIGN; FAT; DIFFERENTIATION;
D O I
10.1007/s00256-014-1883-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Our purpose is to test the effect of varied in-phase (IP) and opposed-phase (OP) sequence order on characterizing marrow signal changes at 3T. The study was HIPAA compliant and IRB approved. Informed consent was waived. At 3T, IP and OP sequences were acquired in three patients with biopsy-proven osteosarcomas, using two methods: approach 1 (OP acquisition before IP acquisition) and approach 2 (OP after IP). Signal intensity (SI) measurements in 12 locations of biopsy-proven osteosarcoma and in six locations with normal bone marrow were performed independently by two experienced musculoskeletal radiologists. The signal intensity ratio (SIR) was measured within the marrow where there was T1 signal lower than skeletal muscle. A SIR < 20 % on the OP compared with IP imaging was considered positive for marrow replacement, while SIR > = 20 % was considered negative. Interobserver agreement was measured by the Lin concordance correlation coefficient (CCC). In 75 % (18/24) of locations within the biopsy-proven tumors, the SIR was > 20 % (SI drop more than 20 % in OP compared to IP) using approach 2 and in 100 % (24/24) of the locations the SIR was < 20 % (SI drop less than 20 % in OP compared to IP) using approach 1, indicating a high percentage of false-negative results by approach 2, and no false-negative results with approach 1. There was good agreement between observer measurement (CCC = 0.96). At 3T, the OP sequence should be acquired prior to the IP sequence, because susceptibility artifacts on a later-acquired OP sequence may lead to an erroneous interpretation of marrow signal abnormalities.
引用
收藏
页码:1139 / 1147
页数:9
相关论文
共 13 条
[11]   CHARACTERIZATION OF ADRENAL MASSES WITH CHEMICAL-SHIFT MR-IMAGING - HOW TO SELECT ECHO TIMES [J].
TSUSHIMA, Y ;
DEAN, PB .
RADIOLOGY, 1995, 195 (01) :285-286
[12]   Benign and malignant processes: Normal values and differentiation with chemical shift MR imaging in vertebral marrow [J].
Zajick, DC ;
Morrison, WB ;
Schweitzer, ME ;
Parellada, JA ;
Carrino, JA .
RADIOLOGY, 2005, 237 (02) :590-596
[13]   Value of opposed-phase gradient-echo technique in distinguishing between benign and malignant vertebral lesions [J].