Model-based synthetic fuzzy logic controller for indirect blood pressure measurement

被引:19
作者
Wang, JJ
Lin, CT [1 ]
Liu, SH
Wen, ZC
机构
[1] I Shou Univ, Dept Biomed Engn, Kaohsiung, Taiwan
[2] Natl Chiao Tung Univ, Dept Elect & Control Engn, Hsinchu 300, Taiwan
[3] Yuanpei Inst Sci & Technol, Dept Biomed Engn, Hsinchu 300, Taiwan
[4] Hsinchu Hosp, Dept Hlth Execut Yuan, Hsinchu 300, Taiwan
来源
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS PART B-CYBERNETICS | 2002年 / 32卷 / 03期
关键词
compliance; fuzzy logic control; mean arterial pressure; oscillometry; tonometer;
D O I
10.1109/TSMCB.2002.999807
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this paper, a new measurement system for the non-invasive monitoring of the continuous blood pressure waveform in the radial artery is presented. The proposed system comprises a model-based fuzzy logic controller, an arterial tonometer and a micro syringe device. The flexible diaphragm tonometer is to register the continuous blood pressure waveform. To obtain accurate measurement without distortion, the tonometer's mean chamber pressure must be kept equal to the mean arterial pressure (MAP), the so-called optimal coupling condition, such that the arterial vessel has the maximum compliance. Since the MAP cannot be measured directly, to keep the optimal coupling condition becomes a tracking control problem with unknown desired trajectory. To solve this dilemma, a model-based fuzzy logic controller is designed to compensate the change of MAP by applying a counter pressure on the tonometer chamber through the micro syringe device. The proposed controller consists of a model-based predictor and a synthetic fuzzy logic controller (SFLC). The model-based predictor is to estimate the MAPs changing tendency based on the identified arterial pressure-volume model. The SFLC is composed of three subcontrollers, each of which is a simple fuzzy logic controller, for processing the three changing states of the MAP: ascending, descending and stabilizing states, respectively. Simulation results show that, for the MAP with changing rates of +/-10, +/-20 or +/-30 mm Hg/min, the model-based SFLC can beat-to-beat adjust the tonometer's chamber pressure only with a mean square error of 1.9, 2.2, or 2.8 nun Hg, respectively.
引用
收藏
页码:306 / 315
页数:10
相关论文
共 39 条
[1]  
ANEDERSON BDO, 1989, OPTIMAL CONTROL LINE
[2]  
[Anonymous], 1993, System Modeling and Identification
[3]   Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure [J].
Baker, PD ;
Westenskow, DR ;
Kuck, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :271-278
[4]   NONINVASIVE ASSESSMENT OF MECHANICAL-PROPERTIES OF PERIPHERAL ARTERIES [J].
BUNTIN, CM ;
SILVER, FH .
ANNALS OF BIOMEDICAL ENGINEERING, 1990, 18 (05) :549-566
[5]   IMPROVED TRACKING OF LIMB OCCLUSION PRESSURE FOR SURGICAL TOURNIQUETS [J].
BUSSANI, CR ;
MCEWEN, JA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (04) :221-229
[6]   On equivalence of approximate reasoning results using different interpretations of fuzzy if-then rules [J].
Czogala, E ;
Leski, J .
FUZZY SETS AND SYSTEMS, 2001, 117 (02) :279-296
[7]   MECHANICS OF THE OCCLUSIVE ARM CUFF AND ITS APPLICATION AS A VOLUME SENSOR [J].
DRZEWIECKI, G ;
BANSAL, V ;
KARAM, E ;
HOOD, R ;
APPLE, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (07) :704-708
[8]  
Drzewiecki G., 1996, Professional Program Proceedings. ELECTRO '96 (Cat. No.96CH35926), P61, DOI 10.1109/ELECTR.1996.501212
[9]   THEORY OF THE OSCILLOMETRIC MAXIMUM AND THE SYSTOLIC AND DIASTOLIC DETECTION RATIOS [J].
DRZEWIECKI, G ;
HOOD, R ;
APPLE, H .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) :88-96
[10]  
Eckerle J.S., 1988, ENCY MED DEVICES INS