Hemodynamic support with the pulsatile catheter pump in a sheep model of acute heart failure

被引:8
作者
Li, Zhicheng
Gu, Y. John
Ye, Qing
Cheng, Shaofei
Wang, Weijun
Tang, Min
Zhao, Xiaogang
Rakhorst, Gerhard
Chen, Changzhi
机构
[1] Shanghai Jiao Tong Univ, Ren Ji Hosp, Dept Cardiothorac Surg, Sch Med, Shanghai 200127, Peoples R China
[2] Univ Groningen, Med Ctr, Dept Biomed Engn, Groningen, Netherlands
关键词
left ventricular assist device; catheter pump; heart failure; sheep;
D O I
10.1111/j.1525-1594.2006.00316.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study was aimed to mimic clinical heart failure (HF) conditions and to assess the effect of pulsatilecatheter (PUCA) pump support on hemodynamics and tissue perfusion in a sheep model of acute HF. In 14 sheep, HF was induced by partial occluding the middle left circumflex coronary artery combined with pacemaker-induced tachycardia. PUCA pump was then activated to support the HF for 3 h. Hemodynamic parameters were recorded at baseline, HF, and then every 30 min during experiments. Blood samples were taken in carotid artery (CA), pulmonary artery (PA), and coronary sinus (CS) for the determination of oxygen saturation (SO2) and lactate concentration as markers of tissue perfusion. Results showed that HF model was induced successfully in 10 sheep and failed in four sheep due to refractory ventricular fibrillation. PUCA pump support was successful in seven out of 10 sheep for 3 h. Three cases failed due to technical problems. After HF (n = 10), cardiac output (CO) was decreased from 3.7 +/- 0.5 to 2.0 +/- 0.5 L/min (P < 0.001). Mean arterial pressure (MAP) was lowered from 116.1 +/- 14.2 to 68.1 +/- 14.7 mm Hg (P < 0.001). In seven sheep supported with PUCA pump, MAP rose from 68.9 +/- 15.2 to 94.7 +/- 14.7 mm Hg (P = 0.005), systolic blood pressure increased from 86.6 +/- 17.0 to 112.6 +/- 17.1 mm Hg (P = 0.009), and diastolic blood pressure increased from 57.7 +/- 12.6 to 79.9 +/- 13.9 mm Hg (P = 0.011). CO remained at about 2.0 L/min. SO2 in CA, PA, and CS decreased significantly after HF (P < 0.001), with an increase after support (compared with HF, P < 0.001, 0.066 and 0.114, respectively). Lactate concentrations increased gradually in CA, PA, and CS toward the end of experiments without difference among different sampling sites. This HF model in sheep is simple, easy to manipulate, reproducible and reflecting clinical HF conditions. PUCA pump can maintain the hemodynamic status for 3 h in this acute HF model.
引用
收藏
页码:881 / 888
页数:9
相关论文
共 24 条
[1]   A MODEL TO PREDICT SURVIVAL AT TIME OF POSTCARDIOTOMY INTRAAORTIC BALLOON PUMP INSERTION [J].
BALDWIN, RT ;
SLOGOFF, S ;
NOON, GP ;
SEKELA, M ;
FRAZIER, OH ;
EDELMAN, SK ;
VAUGHN, WK .
ANNALS OF THORACIC SURGERY, 1993, 55 (04) :908-913
[2]   Preoperative intraaortic balloon pump enhances cardiac performance and improves the outcome of redo CABG [J].
Christenson, JT ;
Badel, P ;
Simonet, F ;
Schmuziger, M .
ANNALS OF THORACIC SURGERY, 1997, 64 (05) :1237-1244
[3]   Left ventricular function during support with an asynchronous pulsatile left ventricular assist device [J].
Dalby, MCD ;
Banner, NR ;
Tansley, P ;
Grieve, LA ;
Partridge, J ;
Yacoub, MH .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2003, 22 (03) :292-300
[4]  
Huang YF, 1997, ASAIO J, V43, pM408
[5]  
Kuroda H, 1995, ASAIO J, V41, P855, DOI 10.1097/00002480-199510000-00008
[6]  
Kusiak V M, 1985, Cardiovasc Clin, V15, P281
[7]   BETA ADRENOCEPTOR RECOVERY AFTER HEART-FAILURE IN THE DOG [J].
LAROSA, GL ;
ARMSTRONG, PW ;
SEEMAN, P ;
FORSTER, C .
CARDIOVASCULAR RESEARCH, 1993, 27 (03) :489-493
[8]  
LI ZC, 2005, J EMERGENCY MED, V14, P181
[9]  
Maccioli G A, 1988, J Cardiothorac Anesth, V2, P365, DOI 10.1016/0888-6296(88)90320-1
[10]   LARGE ANIMAL-MODEL OF LEFT-VENTRICULAR ANEURYSM [J].
MARKOVITZ, LJ ;
SAVAGE, EB ;
RATCLIFFE, MB ;
BAVARIA, JE ;
KREINER, G ;
IOZZO, RV ;
HARGROVE, WC ;
BOGEN, DK ;
EDMUNDS, LH .
ANNALS OF THORACIC SURGERY, 1989, 48 (06) :838-845