Alumina-alumina artificial hip joints. Part II: Characterisation of the wear debris from in vitro hip joint simulations

被引:79
作者
Tipper, JL
Hatton, A
Nevelos, JE
Ingham, E
Doyle, C
Streicher, R
Nevelos, AB
Fisher, J [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Div Microbiol, Sch Mol Biol & Biochem, Leeds LS2 9JT, W Yorkshire, England
[3] Stryker Howmedica Osteon, Newbury, Berks, England
[4] Stryker Howmedica Osteon, Kilchberg, Switzerland
[5] Bradford Royal Infirm, Dept Orthopaed Surg, Bradford BD9 6RJ, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
alumina ceramic; total hip arthroplasty; wear debris; in vitro simulation;
D O I
10.1016/S0142-9612(02)00048-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Until recently it was not possible to reproduce clinically relevant wear rates and wear patterns in in Vitro hip joint simulators for alumina ceramic-on-ceramic hip prostheses. The introduction of microseparation of the prosthesis components into in vitro wear simulations produced clinically relevant wear rates and wear patterns for the first time. The aim of this study was to characterise the wear particles generated from standard simulator testing and microseparation simulator testing of hot isostatically pressed (HIPed) and non-HIPed alumina ceramic-on-ceramic hip prostheses and compare these particles to those generated in vivo. Standard simulation conditions produced wear rates of approximate to0.1 mm(3) per million cycles for both material types. No change in surface roughness was detected and very few wear features were observed. In contrast, when microseparation was introduced into the wear simulation. wear rates of between 1.24 (HIPed) and 1.74mm(3) per million cycles (non-HIPed) were produced. Surface roughness increased and a wear stripe often observed clinically on retrieved femoral heads was also reproduced. Under standard simulation conditions only nanometre-sized wear particles (2-27.5 nm) were observed by TEM mid it was thought likely that these particles resulted from relief polishing of the alumina ceramic. However, when microseparation of the prosthesis components was introduced into the simulation, a bi-modal distribution of particle sizes was observed. The nanometre-sized particles produced by relief polishing were present (1-35nm), however, larger micrometre-sized particles were also observed by both transmission electron microscopy (TEM) (0.02-1 mum) and scanning, electron microscopy (SEM) (0.05- > 10mum). These larger particles were thought to originate from the wear stripe and were produced by trans-granular fracture of the alumina ceramic. In Part I of this study, aluirlina ceramic wear particles were isolated from the periprosthetic tissues from around Mittelmeier ceramic-on-ceramic hip prostheses. Characterisation of the particles by TEM and SEM revealed a bi-modal size distribution. SEM analysis revealed particles in the 0.05-3.2mum size range, and TEM revealed particles in the 5-90nm size range, indicating that microseparation of the prosthesis components may be a common event in vivo. This study (Part II) has revealed that the introduction of microseparation of the prosthesis components during the swing phase of the wear simulation reproduced clinically relevant wear rates, wear patterns and wear particles in in vitro hip joint simulators.(C) 2002 Elsevier Science Ltd. All rights reserved..
引用
收藏
页码:3441 / 3448
页数:8
相关论文
共 32 条
[1]  
[Anonymous], 1992, SOCIOLOGICA, V7
[2]   A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) :455-467
[3]  
Campbell P, 1996, Proc Inst Mech Eng H, V210, P167, DOI 10.1243/PIME_PROC_1996_210_409_02
[4]  
Firkins PJ, 2001, BIO-MED MATER ENG, V11, P143
[5]   TISSUE INGROWTH AND DIFFERENTIATION IN THE BONE-HARVEST CHAMBER IN THE PRESENCE OF COBALT-CHROMIUM-ALLOY AND HIGH-DENSITY-POLYETHYLENE PARTICLES [J].
GOODMAN, S ;
ASPENBERG, P ;
SONG, Y ;
KNOBLICH, G ;
HUIE, P ;
REGULA, D ;
LIDGREN, L .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1025-1035
[6]   Polyethylene particles of a 'critical size' are necessary for the induction of cytokines by macrophages in vitro [J].
Green, TR ;
Fisher, J ;
Stone, M ;
Wroblewski, BM ;
Ingham, E .
BIOMATERIALS, 1998, 19 (24) :2297-2302
[7]   Alumina-alumina artificial hip joints. Part I: a histological analysis and characterisation of wear debris by laser capture microdissection of tissues retrieved at revision [J].
Hatton, A ;
Nevelos, JE ;
Nevelos, AA ;
Banks, RE ;
Fisher, J ;
Ingham, E .
BIOMATERIALS, 2002, 23 (16) :3429-3440
[8]   AL2O3 AGAINST AL2O3 COMBINATION IN HIP ENDOPROSTHESES - HISTOLOGIC INVESTIGATIONS WITH SEMIQUANTITATIVE GRADING OF REVISION AND AUTOPSY CASES AND ABRASION MEASURES [J].
HENSSGE, EJ ;
BOS, I ;
WILLMANN, G .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (9-10) :657-661
[9]   A RAT MODEL OF RESORPTION OF BONE AT THE CEMENT-BONE INTERFACE IN THE PRESENCE OF POLYETHYLENE WEAR PARTICLES [J].
HOWIE, DW ;
VERNONROBERTS, B ;
OAKESHOTT, R ;
MANTHEY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (02) :257-263
[10]   PRODUCTION OF CYTOKINES AROUND LOOSENED CEMENTED ACETABULAR COMPONENTS - ANALYSIS WITH IMMUNOHISTOCHEMICAL TECHNIQUES AND IN-SITU HYBRIDIZATION [J].
JIRANEK, WA ;
MACHADO, M ;
JASTY, M ;
JEVSEVAR, D ;
WOLFE, HJ ;
GOLDRING, SR ;
GOLDBERG, MJ ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (06) :863-879