Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas

被引:5
|
作者
Sabz, Mozhgan [1 ]
Tavernini, Scott [1 ]
Pillay, Kineshta [1 ]
Christianson, Cole [1 ]
Noga, Michelle [2 ]
Finlay, Warren H. [1 ]
Rouhani, Hossein [1 ]
Martin, Andrew R. [1 ]
机构
[1] Univ Alberta, Dept Mech Engn, Edmonton, AB, Canada
[2] Univ Alberta, Dept Radiol & Diagnost Imaging, Edmonton, AB, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Low-flow oxygen delivery; Nasal cannula; In vitro testing; Neonates; Infants; Airway replicas; Fraction of inspired oxygen; Oxygen concentration; Oxygen therapy; Supplemental oxygen; OBSTRUCTION; CHILDREN; HEALTHY;
D O I
10.1186/s12931-022-02260-x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: The nasal cannula is considered a trusted and effective means of administering low-flow oxygen and is widely used for neonates and infants requiring oxygen therapy, despite an understanding that oxygen concentrations delivered to patients are variable. Methods: In the present study, realistic nasal airway replicas derived from medical scans of children less than 3 months old were used to measure the fraction of oxygen inhaled (FiO(2)) through nasal cannulas during low-flow oxygen delivery. Parameters influencing variability in FiO(2) were evaluated, as was the hypothesis that measured FiO(2) values could be predicted using a simple, flow-weighted calculation that assumes ideal mixing of oxygen with entrained room air. Tidal breathing through neonatal and infant nasal airway replicas was controlled using a lung simulator. Parameters for nasal cannula oxygen flow rate, nasal airway geometry, tidal volume, respiratory rate, inhalation/exhalation, or I:E ratio (t(i)/t(e)), breath waveform, and cannula prong insertion position were varied to determine their effect on measured FiO(2). In total, FiO(2) was measured for 384 different parameter combinations, with each combination repeated in triplicate. Analysis of variance (ANOVA) was used to assess the influence of parameters on measured FiO(2). Results: Measured FiO(2) was not appreciably affected by the breath waveform shape, the replica geometry, or the cannula position but was significantly influenced by the tidal volume, the inhalation time, and the nasal cannula flow rate. Conclusions: The flow-weighted calculation overpredicted FiO(2) for measured values above 60%, but an empirical correction to the calculation provided good agreement with measured FiO(2) across the full range of experimental data.
引用
收藏
页数:10
相关论文
共 37 条
  • [31] Assessment of the application of double-modified nasopharyngeal airways versus the use of low-flow nasal cannula during pediatric upper gastrointestinal endoscopy: A prospective, randomized, noninferiority, controlled trial
    Hussein, Mostafa M.
    Amer, Akram M.
    Maarouf, Mohammed M.
    JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2024, 40 (03) : 403 - 409
  • [32] High-flow oxygen therapy through nasal cannulae versus low-flow oxygen therapy via Venturi mask after extubation in adult, critically ill patients
    F Antonicelli
    A Cataldo
    R Festa
    F Idone
    A Moccaldo
    M Antonelli
    SM Maggiore
    Critical Care, 15 (Suppl 1):
  • [33] Nasal High-Flow Therapy Versus Low-Flow Oxygen Therapy Post-Discharge in Infants With Bronchopulmonary Dysplasia: A Retrospective Observational Study
    Mariani, Alessia
    Ghezzi, Elisabetta
    Pesenti, Nicola
    Ophorst, Marijike
    Polimeni, Valentina
    Gangi, Silvana
    Colnaghi, Mariarosa
    Mosca, Fabio
    Lavizzari, Anna
    PEDIATRIC PULMONOLOGY, 2025, 60 (01)
  • [34] Preliminary evaluation of the use of high-flow nasal cannula oxygen therapy during recovery from general anesthesia in dogs with obstructive upper airway breathing
    Jagodich, Tiffany A.
    Bersenas, Alexa M. E.
    Bateman, Shane W.
    Kerr, Carolyn L.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (04) : 487 - 492
  • [35] High flow nasal cannula and low level continuous positive airway pressure have different physiological effects during de novo acute hypoxemic respiratory failure
    Tuffet, Samuel
    Boujelben, Mohamed Ahmed
    Haudebourg, Anne-Fleur
    Maraffi, Tommaso
    Perier, Francois
    Labedade, Pascale
    Moncomble, Elsa
    Gendreau, Segolene
    Lacheny, Matthieu
    Vivier, Emmanuel
    Mekontso-Dessap, Armand
    Carteaux, Guillaume
    ANNALS OF INTENSIVE CARE, 2024, 14 (01):
  • [36] Respiratory support with standard low-flow oxygen therapy, high-flow oxygen therapy or continuous positive airway pressure in adults with acute hypoxaemic respiratory failure in a resource-limited setting: protocol for a randomised, open-label, clinical trial - the Acute Respiratory Intervention StudiEs in Africa (ARISE-AFRICA) study
    Kwizera, Arthur
    Kabatoro, Daphne
    Owachi, Darius
    Kansiime, Jackson
    Kateregga, George
    Nanyunja, Doreen
    Sendagire, Cornelius
    Nyakato, Doreen
    Olaro, Charles
    Audureau, Etienne
    Dessap, Armand Mekontso
    BMJ OPEN, 2024, 14 (06):
  • [37] High-flow nasal cannula heated-humidified (HFNC) oxygen (O2) at 2 litres (lt)/kg/min versus low-flow O2 (up to 2lt/min) in pediatric emergency department (PED) and pediatric ward (PW) in infants with bronchiolitis who need O2 administration: A combined prospective randomized controlled and retrospective study: Preliminary results
    Kefala, Kalomoira
    Nyamugabo, Kindja
    Farhat, Nesrine
    Janssen, Aurelie
    Seghaye, Marie-Christine
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46