Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas

被引:5
|
作者
Sabz, Mozhgan [1 ]
Tavernini, Scott [1 ]
Pillay, Kineshta [1 ]
Christianson, Cole [1 ]
Noga, Michelle [2 ]
Finlay, Warren H. [1 ]
Rouhani, Hossein [1 ]
Martin, Andrew R. [1 ]
机构
[1] Univ Alberta, Dept Mech Engn, Edmonton, AB, Canada
[2] Univ Alberta, Dept Radiol & Diagnost Imaging, Edmonton, AB, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Low-flow oxygen delivery; Nasal cannula; In vitro testing; Neonates; Infants; Airway replicas; Fraction of inspired oxygen; Oxygen concentration; Oxygen therapy; Supplemental oxygen; OBSTRUCTION; CHILDREN; HEALTHY;
D O I
10.1186/s12931-022-02260-x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: The nasal cannula is considered a trusted and effective means of administering low-flow oxygen and is widely used for neonates and infants requiring oxygen therapy, despite an understanding that oxygen concentrations delivered to patients are variable. Methods: In the present study, realistic nasal airway replicas derived from medical scans of children less than 3 months old were used to measure the fraction of oxygen inhaled (FiO(2)) through nasal cannulas during low-flow oxygen delivery. Parameters influencing variability in FiO(2) were evaluated, as was the hypothesis that measured FiO(2) values could be predicted using a simple, flow-weighted calculation that assumes ideal mixing of oxygen with entrained room air. Tidal breathing through neonatal and infant nasal airway replicas was controlled using a lung simulator. Parameters for nasal cannula oxygen flow rate, nasal airway geometry, tidal volume, respiratory rate, inhalation/exhalation, or I:E ratio (t(i)/t(e)), breath waveform, and cannula prong insertion position were varied to determine their effect on measured FiO(2). In total, FiO(2) was measured for 384 different parameter combinations, with each combination repeated in triplicate. Analysis of variance (ANOVA) was used to assess the influence of parameters on measured FiO(2). Results: Measured FiO(2) was not appreciably affected by the breath waveform shape, the replica geometry, or the cannula position but was significantly influenced by the tidal volume, the inhalation time, and the nasal cannula flow rate. Conclusions: The flow-weighted calculation overpredicted FiO(2) for measured values above 60%, but an empirical correction to the calculation provided good agreement with measured FiO(2) across the full range of experimental data.
引用
收藏
页数:10
相关论文
共 37 条
  • [21] Airway management with Hi-flow nasal cannula oxygen in children with severe laryngeal obstruction
    Xiong, Ling
    Liu, Jianxia
    Li, Haisu
    Tan, Yanzhe
    Tang, Linlin
    Du, Min
    Xu, Ying
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2024, 176
  • [22] Nebulizers and spacers for aerosol delivery through adult nasal cannula at low oxygen flow rate: An in-vitro study
    Madney, Yasmin M.
    Fathy, Maha
    Elberry, Ahmed A.
    Rabea, Hoda
    Abdelrahim, Mohamed E. A.
    JOURNAL OF DRUG DELIVERY SCIENCE AND TECHNOLOGY, 2017, 39 : 260 - 265
  • [23] In Vitro Comparison of Heliox and Oxygen in Aerosol Delivery Using Pediatric High Flow Nasal Cannula
    Ari, Arzu
    Harwood, Robert
    Sheard, Meryl
    Dailey, Patricia
    Fink, James B.
    PEDIATRIC PULMONOLOGY, 2011, 46 (08) : 795 - 801
  • [24] Airway Management of an Infant With Giant Neck Macro-Cystic Hygroma Utilizing a High-Flow Nasal Cannula
    Huang, Shiwei
    Wang, Zhihao
    Chan, Yauwai
    Jiang, Tao
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (10)
  • [25] Aerosol delivery through high flow nasal cannula compared to biphasic positive airway pressure, at two different pressure: an in-vitro study
    Marina E. Boules
    Nabila Ibrahim Laz
    Ahmed A. Elberry
    Raghda R. S. Hussein
    Mohamed E. A. Abdelrahim
    Beni-Suef University Journal of Basic and Applied Sciences, 10
  • [26] Aerosol delivery through high flow nasal cannula compared to biphasic positive airway pressure, at two different pressure: an in-vitro study
    Boules, Marina E.
    Laz, Nabila Ibrahim
    Elberry, Ahmed A.
    Hussein, Raghda R. S.
    Abdelrahim, Mohamed E. A.
    BENI-SUEF UNIVERSITY JOURNAL OF BASIC AND APPLIED SCIENCES, 2021, 10 (01)
  • [27] Apneic oxygenation with low-flow oxygen cannula for rapid sequence induction and intubation in pediatric patients: a randomized-controlled trial
    Aroonpruksakul, Naiyana
    Sangsungnern, Peerapong
    Kiatchai, Taniga
    TRANSLATIONAL PEDIATRICS, 2022, 11 (04) : 427 - 437
  • [28] Effectiveness and Safety of High-Flow Nasal Cannula Oxygen Delivery during Bronchoalveolar Lavage in Acute Respiratory Failure Patients
    Kim, Eun Jin
    Jung, Chi Young
    Kim, Kyung Chan
    TUBERCULOSIS AND RESPIRATORY DISEASES, 2018, 81 (04) : 319 - 329
  • [29] FIRST, DO NO HARM: A Response to "Oral Alimentation in Neonatal and Adult Populations Requiring High-Flow Oxygen via Nasal Cannula"
    Dodrill, Pamela
    Gosa, Memorie
    Thoyre, Suzanne
    Shaker, Catherine
    Pados, Britt
    Park, Jinhee
    DePalma, Nicole
    Hirst, Keith
    Larson, Kara
    Perez, Jennifer
    Hernandez, Kayla
    DYSPHAGIA, 2016, 31 (06) : 781 - 782
  • [30] FIRST, DO NO HARM: A Response to “Oral Alimentation in Neonatal and Adult Populations Requiring High-Flow Oxygen via Nasal Cannula”
    Pamela Dodrill
    Memorie Gosa
    Suzanne Thoyre
    Catherine Shaker
    Britt Pados
    Jinhee Park
    Nicole DePalma
    Keith Hirst
    Kara Larson
    Jennifer Perez
    Kayla Hernandez
    Dysphagia, 2016, 31 : 781 - 782