Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning

被引:208
作者
Yoo, Sua [1 ]
Yin, Fang-Fang [1 ]
机构
[1] Duke Univ, Ctr Med, Dept Radiat Oncol, Durham, NC 27710 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2006年 / 66卷 / 05期
关键词
cone-beam CT; HU values; treatment planning;
D O I
10.1016/j.ijrobp.2006.08.031
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Cone-beam computed tomography (CBCT) images are currently used for positioning verification. However, it is yet unknown whether CBCT could be used in dose calculation for replanning in adaptive radiation therapy. This study investigates the dosimetric feasibility of CBCT-based treatment planning. Methods and Materials: Hounsfield unit (HU) values and profiles of Catphan, homogeneous/inhomogeneous phantoms, and various tissue regions of patients in CBCT images were compared to those in CT. The dosimetric consequence of the HU variation was investigated by comparing CBCT-based treatment plans to conventional CT-based plans for both phantoms and patients. Results: The maximum HU difference between CBCT and CT of Catphan was 34 HU in the Teflon. The differences in other materials were less than 10 HU. The profiles for the homogeneous phantoms in CBCT displayed reduced HU values up to 150 HU in the peripheral regions compared to those in CT. The scatter and artifacts in CBCT became severe surrounding inhomogeneous tissues with reduced HU values up to 200 HU. The MU/cGy differences were less than 1% for most phantom cases. The isodose distributions between CBCT-based and CT-based plans agreed very well. However, the discrepancy was larger when CBCT was scanned without a bowtie filter than with bowtie filter. Also, up to 3% dosimetric error was observed in the plans for the inhomogeneous phantom. In the patient studies, the discrepancies of isodose lines between CT-based and CBCT-based plans, both 3D and IMRT, were less than 2 mm. Again, larger discrepancy occurred for the lung cancer patients. Conclusion: This study demonstrated the feasibility of CBCT-based treatment planning. CBCT-based treatment plans were dosimetrically comparable to CT-based treatment plans. Dosimetric data in the inhomogeneous tissue regions should be carefully validated. (c) 2006 Elsevier Inc.
引用
收藏
页码:1553 / 1561
页数:9
相关论文
共 28 条
[1]  
*AAPM, 2004, 85 AAPM
[2]  
BARRETT J, 2004, RADIO GRAPHICS, P24
[3]   COMPUTED-TOMOGRAPHY FOR RADIOTHERAPY PLANNING [J].
BATTISTA, JJ ;
RIDER, WD ;
VANDYK, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1980, 6 (01) :99-107
[4]  
Bushberg JT., ESSENTIAL PHYS MED I, P359
[5]   AN ELECTRON-DENSITY CALIBRATION PHANTOM FOR CT-BASED TREATMENT PLANNING COMPUTERS [J].
CONSTANTINOU, C ;
HARRINGTON, JC ;
DEWERD, LA .
MEDICAL PHYSICS, 1992, 19 (02) :325-327
[6]   Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J].
Cozzi, L ;
Fogliata, A ;
Buffa, F ;
Bieri, S .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :335-338
[7]   IMPROVED LUNG DOSE CALCULATION USING TISSUE MAXIMUM RATIOS IN THE BATHO CORRECTION [J].
ELKHATIB, E ;
BATTISTA, JJ .
MEDICAL PHYSICS, 1984, 11 (03) :279-286
[8]   Effect of scatter radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :514-521
[9]   NON-LINEAR PARTIAL VOLUME ARTIFACTS IN X-RAY COMPUTED-TOMOGRAPHY [J].
GLOVER, GH ;
PELC, NJ .
MEDICAL PHYSICS, 1980, 7 (03) :238-248
[10]   Organ and tumor motion: An overview [J].
Goitein, M .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :2-9