Interference by cellular phones with permanent implanted pacemakers:: an update

被引:42
|
作者
Hekmat, K
Salemink, B
Lauterbach, G
Schwinger, RHG
Südkamp, M
Weber, HJ
Mehlhorn, U
机构
[1] Univ Cologne, Dept Cardiothorac Surg, D-50924 Cologne, Germany
[2] Univ Aachen, Dept Cardiovasc Engn Julich, Aachen, Germany
[3] Univ Cologne, Dept Cardiol, D-5000 Cologne 41, Germany
来源
EUROPACE | 2004年 / 6卷 / 04期
关键词
cellular telephones; interference; feedthru filter; pacemaker;
D O I
10.1016/j.eupc.2004.03.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The aim of this study was to test pacemakers with feedthru fitters for interference with a digital cellular telephone. Methods The study comprised 100 patients having their first pacemaker implantation between January 2001 and May 2003. A GSM-standard cellular phone was tested in the standby, dialing and operating mode against 23 single-chamber and 77 dual-chamber pacemakers. Continuous surface electrocardiograms, intracardiac electrograms, and marker channels were recorded when calls were made by a land phone to the cellular phone. Results In 2 patients we observed pacing inhibition with the cellular phone positioned directly above the pacemaker pocket. The unipolar and bipolar ventricular sensitivity setting was 0.25 mV in one dual-chamber pacemaker, and when we changed the sensitivity to 0.50 mV and higher no interference could be detected. The second inhibition occurred in a single-chamber pacemaker with unipolar and bipolar ventricular sensitivity setting of 0.5 mV, where a sensitivity change to 1.0 mV eliminated the interference. Conclusion Anticipating a correct setting of ventricular sensitivity, currently available pacemakers equipped with feedthru fitters do not show any interference with cellular phones. Since interference was only observed with sensitivity settings below 0.50 mV, we recommend that permanent programming of ventricular sensitivity should be set at 2.0 mV and higher. (C) 2004 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:363 / 369
页数:7
相关论文
共 50 条
  • [41] Predictors of upper extremities veins obstruction in patients with permanent pacemakers
    Povarov, Vladislav O.
    Kalinin, Roman E.
    Mzhavanadze, Nina D.
    Zhurina, Olga N.
    Suchkov, Igor A.
    ACTA PHLEBOLOGICA, 2024, 25 (02): : 90 - 95
  • [42] How to test mode switching in pacemakers implanted in patients: The MOST study
    Padeletti, L
    Gasparini, M
    Porciani, MC
    Pieragnoli, P
    Colella, A
    Michelucci, A
    Proclemer, A
    Tognarini, S
    Mantica, M
    Coltorti, F
    Corbucci, G
    Sutton, R
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (02): : 156 - 160
  • [43] QUALITY OF LIFE OF PATIENTS WITH IMPLANTED CARDIAC PACEMAKERS IN NORTH WEST OF IRAN
    Kouchaksaraei, Fateme Ranjbar
    Akbarzadeh, Fariborz
    Hashemi, Mohammad
    ARYA ATHEROSCLEROSIS, 2007, 2 (04) : 197 - 203
  • [44] Left Ventricular Mass Index Predicts Pacemaker-Induced Cardiomyopathy in Patients with Dual Chamber Permanent Pacemakers Implanted Due to Complete Atrioventricular Block
    Hayiroglu, Mert Ilker
    Cinar, Tufan
    Cinier, Goeksel
    Yuksel, Gizem
    Ayan, Gokcem
    Pay, Levent
    Coskun, Cahit
    Keskin, Kivanc
    Cicek, Vedat
    Tekkesin, Ahmet Ilker
    ACTA CARDIOLOGICA SINICA, 2023, 39 (03) : 416 - 423
  • [45] The perioperative management of small animals with previously implanted pacemakers undergoing anaesthesia
    Carell, Alejandra Garcia de
    Casamian-Sorrosal, Domingo
    VETERINARY ANAESTHESIA AND ANALGESIA, 2022, 49 (01) : 7 - 17
  • [46] THE USEFULNESS OF A STRETCH-POLYESTER POUCH TO ENCASE IMPLANTED PACEMAKERS AND DEFIBRILLATORS
    PARSONNET, V
    BERNSTEIN, AD
    NEGLIA, D
    OMAR, A
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (12): : 2274 - 2278
  • [47] The Impact of the Internet of Things on Implanted Medical Devices including Pacemakers, and ICDs
    Stachel, Joshua R.
    Sejdic, Ervin
    Ogirala, Ajay
    Mickle, Marlin H.
    2013 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2013, : 839 - 844
  • [48] Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven
    Hirose, M
    Hida, M
    Sato, E
    Kokubo, K
    Nie, M
    Kobayashi, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (06): : 540 - 548
  • [49] REGISTRATION OF PACEMAKER-INDUCED SKIN POTENTIALS IN ROUTINE CONTROL OF IMPLANTED PACEMAKERS
    GRENDAHL, H
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1969, 24 (03) : 251 - &
  • [50] Atrial tachyarrhythmia recurrence temporal patterns in bradycardia patients implanted with antitachycardia pacemakers
    Ricci, R
    Santini, M
    Padeletti, L
    Boriani, G
    Capucci, A
    Botto, G
    Gulizia, M
    Inama, G
    Galati, A
    Solimene, F
    Pepe, M
    Grammatico, A
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (01) : 44 - 51