Interference by cellular phones with permanent implanted pacemakers:: an update

被引:42
|
作者
Hekmat, K
Salemink, B
Lauterbach, G
Schwinger, RHG
Südkamp, M
Weber, HJ
Mehlhorn, U
机构
[1] Univ Cologne, Dept Cardiothorac Surg, D-50924 Cologne, Germany
[2] Univ Aachen, Dept Cardiovasc Engn Julich, Aachen, Germany
[3] Univ Cologne, Dept Cardiol, D-5000 Cologne 41, Germany
来源
EUROPACE | 2004年 / 6卷 / 04期
关键词
cellular telephones; interference; feedthru filter; pacemaker;
D O I
10.1016/j.eupc.2004.03.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The aim of this study was to test pacemakers with feedthru fitters for interference with a digital cellular telephone. Methods The study comprised 100 patients having their first pacemaker implantation between January 2001 and May 2003. A GSM-standard cellular phone was tested in the standby, dialing and operating mode against 23 single-chamber and 77 dual-chamber pacemakers. Continuous surface electrocardiograms, intracardiac electrograms, and marker channels were recorded when calls were made by a land phone to the cellular phone. Results In 2 patients we observed pacing inhibition with the cellular phone positioned directly above the pacemaker pocket. The unipolar and bipolar ventricular sensitivity setting was 0.25 mV in one dual-chamber pacemaker, and when we changed the sensitivity to 0.50 mV and higher no interference could be detected. The second inhibition occurred in a single-chamber pacemaker with unipolar and bipolar ventricular sensitivity setting of 0.5 mV, where a sensitivity change to 1.0 mV eliminated the interference. Conclusion Anticipating a correct setting of ventricular sensitivity, currently available pacemakers equipped with feedthru fitters do not show any interference with cellular phones. Since interference was only observed with sensitivity settings below 0.50 mV, we recommend that permanent programming of ventricular sensitivity should be set at 2.0 mV and higher. (C) 2004 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:363 / 369
页数:7
相关论文
共 50 条
  • [31] Programmable Hypertension Control: Another Possible Indication for Implanted Pacemakers
    Do, Duc H.
    Shivkumar, Kalyanam
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2017, 6 (12):
  • [32] Clinical Benefits of Remote Versus Transtelephonic Monitoring of Implanted Pacemakers
    Crossley, George H.
    Chen, Jane
    Choucair, Wassim
    Cohen, Todd J.
    Gohn, Douglas C.
    Ben Johnson, W.
    Kennedy, Eleanor E.
    Mongeon, Luc R.
    Serwer, Gerald A.
    Qiao, Hongyan
    Wilkoff, Bruce L.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (22) : 2012 - 2019
  • [33] The combined transvenous implantation of cardioverter defibrillators and permanent pacemakers
    Mattke, S
    Markewitz, A
    Muller, D
    Grunewald, A
    Fiek, M
    Schmockel, M
    Hoffmann, E
    Steinbeck, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (11): : 2775 - 2782
  • [34] Effects of Radiofrequency Catheter Ablation on Patients with Permanent Pacemakers
    Sadoul N.
    Blankoff I.
    De Chillou C.
    Beurrier D.
    Messier M.
    Bizeau O.
    Magnin I.
    Dodinot B.
    Aliot E.
    Journal of Interventional Cardiac Electrophysiology, 1997, 1 (3) : 227 - 233
  • [35] Electromagnetic interference with pacemakers caused by portable media players
    Thaker, Jay P.
    Patel, Mehul B.
    Jongnarangsin, Krit
    Liepa, Valdis V.
    Thakur, Ranjan K.
    HEART RHYTHM, 2008, 5 (04) : 538 - 544
  • [36] Interference by Modern Smartphones and Accessories with Cardiac Pacemakers and Defibrillators
    Fahd Nadeem
    Cao Thach Tran
    Estelle Torbey
    Daniel Philbin
    Carlos Morales
    Michael Wu
    Current Cardiology Reports, 2022, 24 : 347 - 353
  • [37] Concerns about sources of electromagnetic interference in patients with pacemakers
    Sakakibara, Y
    Mitsui, T
    JAPANESE HEART JOURNAL, 1999, 40 (06): : 737 - 743
  • [38] Interference by Modern Smartphones and Accessories with Cardiac Pacemakers and Defibrillators
    Nadeem, Fahd
    Tran, Cao Thach
    Torbey, Estelle
    Philbin, Daniel
    Morales, Carlos
    Wu, Michael
    CURRENT CARDIOLOGY REPORTS, 2022, 24 (04) : 347 - 353
  • [39] Do cellular phones really interfere with monitoring?
    Mellor A.
    Critical Care, 5 (1):
  • [40] Left atrial spontaneous echo contrast in patients with permanent pacemakers
    Lee, TM
    Su, SF
    Chien, KL
    Lin, YJ
    Chen, WJ
    Chen, MF
    Liau, CS
    Lee, YT
    HEART, 1997, 78 (03) : 262 - 267