Technical Note: Patient-weight dependence of the effective dose conversion coefficients for diagnostic x-ray imaging procedures

被引:0
|
作者
Golikov, Vladislav [1 ]
Druzhinina, Polina [1 ]
机构
[1] St Petersburg Res Inst Radiat Hyg, Mira Str 8, St Petersburg 197101, Russia
关键词
conversion coefficients; effective dose; patients; x-ray examination; PEDIATRIC BARIUM MEAL; COMPUTATIONAL PHANTOMS; ORGAN; ADULT; AGE;
D O I
10.1002/mp.14446
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this study was to provide a method and reference data for the evaluation of the weight dependence of the conversion coefficients from air collision kerma integrated over beam area to the effective dose for patients undergoing diagnostic x-ray examinations. Method A simple physics-based analytical model was developed in order to calculate the weight dependence of the effective dose conversion coefficients based on existing values of the conversion coefficients between patient effective dose and air collision kerma integrated over beam area calculated by the Monte Carlo method on the base of computational MIRD-5-type stylized models. Results The analytical expressions that make possible taking into account a patient's weight when assessing the effective dose were presented. The following categories of diagnostic x-ray examinations are considered: CT; plain radiography; cardiac angiography; fluoroscopy. The results indicate that the weight dependence of the effective dose conversion coefficients for different categories of diagnostic x-ray examinations has the same form. Correction factor for evaluation of the new value of the effective dose conversion coefficient for a patient with weightW(pa.)when we know its value for a standard patient with weightW(st.)depends on the ratios of their weights (W-st./W-pat.) and fractions of the incident energy which is imparted to a patient. Conclusions The paper presents an approach and necessary data for the estimation of the weight dependence of the effective dose conversion coefficients for patients undergoing different diagnostic x-ray examinations.
引用
收藏
页码:5366 / 5372
页数:7
相关论文
共 13 条
  • [1] Technical Note: Patient-weight dependence of the effective dose conversion coefficients for diagnostic x-ray imaging procedures
    Golikov, Vladislav
    Druzhinina, Polina
    MEDICAL PHYSICS, 2020, 47 (10) : 5366 - 5372
  • [2] Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy X-ray examinations
    Gosch, D.
    Gosch, K.
    Kahn, T.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2007, 179 (10): : 1035 - 1042
  • [3] Higher patient doses through X-ray imaging procedures
    Rehani, Madan M.
    Nacouzi, David
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 : 80 - 86
  • [4] Measurement of conversion coefficients between free in air kerma and personal dose equivalent for diagnostic X-ray beams
    Rosado, P. H. G.
    Nogueira, M. S.
    Genezini, F.
    Vilela, E. C.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 968 - 971
  • [5] Ambient Dose Equivalent, Effective Dose and Mean Sv/Gy Conversion Coefficients from Scattered X-Ray Spectra in Mammography
    Kunzel, R.
    Herdade, S. B.
    Costa, P. R.
    Terini, R. A.
    Levenhagen, R. S.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 2092 - +
  • [6] Technique and Gender Specific Conversion Coefficients for Estimation of Effective Dose from Kerma Area Product During X-Ray Radiography of Chest
    Dimov, Asen
    Tsanev, Ivan
    Penev, Dimitar
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1, 2019, 68 (01): : 719 - 723
  • [7] Effective Dose Calculations in Conventional Diagnostic X-Ray Examinations for Patients in Major Sudanese Hospitals
    Suliman, I. I.
    Habbani, F. I.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 2087 - +
  • [8] Patient effective dose evaluation for chest X-ray examination in three digital radiography centers
    Paydar, R.
    Takavar, A.
    Kardan, M. R.
    Babakhani, A.
    Deevband, M. R.
    Saber, S.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2012, 10 (3-4): : 139 - 143
  • [9] Effective dose in diagnostic radiology as a function of x-ray beam filtration for a constant exit dose and constant film density
    Behrman, RH
    Yasuda, G
    MEDICAL PHYSICS, 1998, 25 (05) : 780 - 790
  • [10] A conversion method of air kerma from the primary, scatter, and leakage radiations to effective dose for calculating x-ray shielding barriers in mammography
    Kharrati, H