Quality assurance and system stability of a clinical MRI-guided focused ultrasound system: Four-year experience

被引:33
作者
McDannold, Nathan [1 ]
Hynynen, Kullervo [1 ]
机构
[1] Harvard Univ, Sch Med, Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
关键词
magnetic resonance; focused ultrasound; water proton chemical shift; temperature mapping;
D O I
10.1118/1.2352853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To retrospectively evaluate the four-year experience of a quality assurance method for a MRI-guided focused ultrasound system that uses temperature maps acquired during heating in an ultrasound/MRI phantom. This quality assurance method was performed before 148 clinical uterine fibroid thermal ablation treatments. The stability of the peak temperature rise, the targeting accuracy, the shape of the heated zone, and the noise level in the imaging was evaluated. The peak temperature rise was mostly stable for the first three years. An increase in heating was observed when the system was replaced after year three. Detection of this increase was taken into account in the subsequent clinical treatments. A small secondary hotspot was detected by the temperature maps and was seen to be resolved after system calibration. The average standard deviation in unheated regions of the phantom in the temperature maps was 0.5 +/- 0.2 degrees C; it was less than 1 degrees C in all but one procedure. The average initial targeting error was 2.8 +/- 1.8 and 2.8 +/- 2.1 mm in two radial directions and 7.7 +/- 2.9 mm along the ultrasound beam direction. The width of the heating profile was consistent over the four years. This simple method to evaluate the performance appeared to be sensitive to small changes in system performance, which was adequately stable over a four-year time period. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4307 / 4313
页数:7
相关论文
共 25 条
[1]   Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam [J].
Chung, AH ;
Hynynen, K ;
Colucci, V ;
Oshio, K ;
Cline, HE ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :745-752
[2]   THE RICIAN DISTRIBUTION OF NOISY MRI DATA [J].
GUDBJARTSSON, H ;
PATZ, S .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (06) :910-914
[3]  
GUDBJARTSSON H, 1996, MAGN RESON MED, V36, P332
[4]   MRI guidance of focused ultrasound therapy of uterine fibroids: Early results [J].
Hindley, J ;
Gedroyc, WM ;
Regan, L ;
Stewart, E ;
Tempany, C ;
Hynnen, K ;
Macdanold, N ;
Inbar, Y ;
Itzchak, Y ;
Rabinovici, J ;
Kim, K ;
Geschwind, JF ;
Hesley, G ;
Gostout, B ;
Ehrenstein, T ;
Hengst, S ;
Sklair-Levy, M ;
Shushan, A ;
Jolesz, F .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (06) :1713-1719
[5]   PROTON RESONANCE SHIFT OF WATER IN GAS AND LIQUID STATES [J].
HINDMAN, JC .
JOURNAL OF CHEMICAL PHYSICS, 1966, 44 (12) :4582-&
[6]   A clinical, noninvasive, MR imaging monitored ultrasound surgery method [J].
Hynynen, K ;
Freund, WR ;
Cline, HE ;
Chung, AH ;
Watkins, RD ;
Vetro, JP ;
Jolesz, FA .
RADIOGRAPHICS, 1996, 16 (01) :185-195
[7]  
*IEEE STD, 7901989 IEEE STD
[8]   CRST SYNDROME (CALCINOSIS CUTIS, RAYNAUDS-PHENOMENON, SCLERODACTYLY, AND TELANGIECTASIA) ASSOCIATED WITH AUTOIMMUNE HEPATITIS [J].
ISHIKAWA, M ;
OKADA, J ;
SHIBUYA, A ;
KONDO, H .
INTERNAL MEDICINE, 1995, 34 (01) :6-9
[9]   MR-guided focused-ultrasound ablation system: Determination of precision and accuracy in preparation for clinical trials [J].
Kendell, KR ;
Wu, T ;
Felmlee, JP ;
Lewis, BD ;
Ehman, RL .
RADIOLOGY, 1998, 209 (03) :856-861
[10]   AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS [J].
MARQUARDT, DW .
JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02) :431-441