Evaluation of a pinhole collimator for I-131 SPECT head imaging

被引:13
作者
Gilland, DR
Johnson, EL
Turkington, TG
Coleman, RE
Jaszczak, RJ
机构
[1] Department of Radiology, Duke University, Medical Center, Durham, NC
关键词
D O I
10.1109/23.531887
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The performance of a pinhole collimator for I-131 SPECT of the head was evaluated. The evaluation included planar and SPECT spatial resolution, sensitivity in air and in water, septal penetration, reconstructed image quality, and activity quantitation within a simple phantom that models tumor uptake in the head. The pinhole collimator was compared to medium and high energy parallel hole collimators. The pinhole collimator demonstrated improved resolution/sensitivity trade-off compared with the parallel hole collimators over the range of distances relevant to head imaging. In planar point source images the pinhole collimator showed reduced penetration effects although in reconstructed images penetration effects were not apparent for either pinhole or high energy parallel hole collimators. Comparable activity quantitation accuracy was observed with all collimators. The accuracy was dependent on the segmentation threshold and calibration procedure. These results indicate that the pinhole collimator can provide improved performance over conventional parallel hole collimators for I-131 imaging in the head.
引用
收藏
页码:2230 / 2238
页数:9
相关论文
共 24 条
[1]  
BICE A N, 1991, Journal of Nuclear Medicine, V32, P1058
[3]  
CULLOM S J, 1989, Journal of Nuclear Medicine, V30, P876
[4]  
DILLMAN LT, 1975, 10 MIRD SOC NUCL MED
[5]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[6]   MEASURED ACCURACY AND PRECISION IN QUANTITATIVE SPECT IMAGING WITH I-123 [J].
GILLAND, DR ;
JASZCZAK, RJ ;
BOWSHER, JE ;
GREER, KL ;
COLEMAN, RE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1120-1125
[7]   ACCURATE QUANTIFICATION OF I-131 DISTRIBUTION BY GAMMA-CAMERA IMAGING [J].
GREEN, AJ ;
DEWHURST, SE ;
BEGENT, RHJ ;
BAGSHAWE, KD ;
RIGGS, SJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1990, 16 (4-6) :361-365
[8]  
Hamming R.W., 1977, DIGITAL FILTERS
[9]   EXTERNAL IMAGING TECHNIQUES FOR QUANTITATION OF DISTRIBUTION OF I-131 F(AB)2 FRAGMENTS OF MONOCLONAL-ANTIBODY IN HUMANS [J].
HAMMOND, ND ;
MOLDOFSKY, PJ ;
BEARDSLEY, MR ;
MULHERN, CB .
MEDICAL PHYSICS, 1984, 11 (06) :778-783
[10]  
ISRAEL O, 1990, J NUCL MED, V31, P1945