Energy efficiency of a new trifocal intraocular lens

被引:19
作者
Vega, F. [1 ]
Alba-Bueno, F. [1 ]
Millan, M. S. [1 ]
机构
[1] Univ Politecn Cataluna, Dept Opt & Optometria, Terrassa 08222, Spain
关键词
Trifocal intraocular lens; diffraction efficiency; cataract; eye model; OPTICAL ABERRATIONS; LIGHT-DISTRIBUTION; CATARACT-SURGERY; IMPLANTATION; CORNEA; DISTANCE; OUTCOMES; QUALITY; DESIGNS; BENCH;
D O I
10.2971/jeos.2014.14002
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The light distribution among the far, intermediate and near foci of a new trifocal intraocular lens (IOL) is experimentally determined, as a function of the pupil size, from image analysis. The concept of focus energy efficiency is introduced because, in addition to the theoretical diffraction efficiency of the focus, it accounts for other factors that are naturally presented in the human eye such as the level of spherical aberration (SA) upon the IOL, light scattering at the diffractive steps or the depth of focus. The trifocal IOL is tested in-vitro in two eye models: the aberration-free ISO model, and a so called modified-ISO one that uses an artificial cornea with positive spherical SA in instead. The SA upon the IOL is measured with a Hartmann-Shack sensor and compared to the values of theoretical eye models. The results show, for large pupils, a notorious reduction of the energy efficiency of the far and near foci of the trifocal IOL due to two facts: the level of SA upon the IOL is larger than the value the lens is able to compensate for and there is significant light scattering at the diffractive steps. On the other hand, the energy efficiency of the intermediate focus for small pupils is enhanced by the contribution of the extended depth of focus of the near and far foci. Thus, while IOLs manufacturers tend to provide just the theoretical diffraction efficiency of the foci to show which would be the performance of the lens in terms of light distribution among the foci, our results put into evidence that this is better described by using the energy efficiency of the foci.
引用
收藏
页数:8
相关论文
共 33 条
[11]   Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench [J].
Gatinel, Damien ;
Houbrechts, Yvette .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (07) :1093-1099
[12]   Design and qualification of a diffractive trifocal optical profile for intraocular lenses [J].
Gatinel, Damien ;
Pagnoulle, Christophe ;
Houbrechts, Yvette ;
Gobin, Laure .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (11) :2060-2067
[13]   Optical aberrations of the human cornea as a function of age [J].
Guirao, A ;
Redondo, M ;
Artal, P .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2000, 17 (10) :1697-1702
[14]  
International Organization for Standardization (ISO), 1999, 119792 ISO
[15]   Effect of Intraocular Lens Asphericity on Quality of Vision after Cataract Removal An Intraindividual Comparison [J].
Kohnen, Thomas ;
Klaproth, Oliver K. ;
Buehren, Jens .
OPHTHALMOLOGY, 2009, 116 (09) :1697-1706
[16]   Change in corneal aberrations after cataract surgery with 2 types of aspherical intraocular lenses [J].
Marcos, Susana ;
Rosales, Patricia ;
Llorente, Lourdes ;
Jimenez-Alfaro, Ignacio .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (02) :217-226
[17]   Performance of presbyopia-correcting intraocular lenses in distance optical bench tests [J].
Maxwell, W. Andrew ;
Lane, Stephen S. ;
Zhou, Fan .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) :166-171
[18]  
Meyers MM, 1998, U.S. Patent, Patent No. [5,801,889, 5801889]
[19]  
Montin JF, 2012, CATARACT REFRACTIV S, V3, P16
[20]   ISO eye model not valid for assessing aspherical lenses [J].
Norrby, Sverker .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (07) :1056-1057