Practical design of a 4 Tesla double-tuned RF surface coil for interleaved 1H and 23Na MRI of rat brain

被引:75
作者
Alecci, M. [1 ]
Romanzetti, S.
Kaffanke, J.
Celik, A.
Wegener, H. P.
Shah, N. J.
机构
[1] Univ Aquila, Dipartimento Sci & Tecnol Biomed, I-67100 Laquila, Italy
[2] Univ Aquila, CNR, INFM, I-67100 Laquila, Italy
[3] Res Ctr Juelich, Inst Med, D-52425 Julich, Germany
[4] Res Ctr Juelich, Cent Elect Inst, D-52425 Julich, Germany
[5] Univ Dortmund, Inst Phys, D-44221 Dortmund, Germany
关键词
MRI; radio frequency surface coil; double tuned; proton; sodium;
D O I
10.1016/j.jmr.2006.04.011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
MRI is proving to be a very useful tool for sodium quantification in animal models of stroke, ischemia, and cancer. In this work, we present the practical design of a dual-frequency RF surface coil that provides H-1 and Na-23 images of the rat head at 4 T. The dual-frequency RF surface coil comprised of a large loop tuned to the H-1 frequency and a smaller co-planar loop tuned to the Na-23 frequency. The mutual coupling between the two loops was eliminated by the use of a trap circuit inserted in the smaller coil. This independent-loop design was versatile since it enabled a separate optimisation of the sensitivity and RF field distributions of the two coils. To allow for an easy extension of this simple double-tuned coil design to other frequencies (nuclei) and dimensions, we describe in detail the practical aspects of the workbench design and MRI testing using a phantom that mimics in vivo conditions. A comparison between our independent-loop, double-tuned coil and a single-tuned Na-23 coil of equal size obtained with a phantom matching in vivo conditions, showed a reduction of the Na-23 sensitivity (about 28%) because of signal losses in the trap inductance. Typical congruent H-1 and Na-23 rat brain images showing good SNR (Na-23: brain 7, ventricular cerebrospinal fluid 11) and spatial resolution (Na-23: 1.25 x 1.25 x 5 mm(3)) are also reported. The in vivo SNR values obtained with this coil were comparable to, if not better than, other contemporary designs in the literature. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:203 / 211
页数:9
相关论文
共 35 条
  • [1] Sodium T2*-weighted MR imaging of acute focal cerebral ischemia in rabbits
    Bartha, R
    Lee, TY
    Hogan, MJ
    Hughes, S
    Barberi, E
    Rajakumar, N
    Menon, RS
    [J]. MAGNETIC RESONANCE IMAGING, 2004, 22 (07) : 983 - 991
  • [2] Chen CN, 1989, BIOMEDICAL MAGNETIC
  • [3] A NOVEL DOUBLE-SURFACE COIL APPROACH TO P-31 SPECTROSCOPY - A STUDY OF HEMISPHERIC BRAIN INJURY IN THE RAT
    CHEW, WM
    MOSELEY, ME
    NISHIMURA, MC
    HASHIMOTO, T
    PITTS, LH
    JAMES, TL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (06) : 567 - 575
  • [4] A SIGNAL-TO-NOISE CALIBRATION PROCEDURE FOR NMR IMAGING-SYSTEMS
    EDELSTEIN, WA
    BOTTOMLEY, PA
    PFEIFER, LM
    [J]. MEDICAL PHYSICS, 1984, 11 (02) : 180 - 185
  • [5] DOUBLE RESONANT QUADRATURE BIRDCAGE
    FITZSIMMONS, JR
    BECK, BL
    BROOKER, HR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (01) : 107 - 114
  • [6] A TRANSFORMER-COUPLED DOUBLE-RESONANT PROBE FOR NMR IMAGING AND SPECTROSCOPY
    FITZSIMMONS, JR
    BROOKER, HR
    BECK, B
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (05) : 471 - 477
  • [7] A COMPARISON OF DOUBLE-TUNED SURFACE COILS
    FITZSIMMONS, JR
    BROOKER, HR
    BECK, B
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 10 (03) : 302 - 309
  • [8] GROVER FW, 1962, INDUCTANCE CALCULATI, P66
  • [9] THE DESIGN AND USE OF A DUAL-FREQUENCY SURFACE COIL PROVIDING PROTON IMAGES FOR IMPROVED LOCALIZATION IN P-31 SPECTROSCOPY OF SMALL LESIONS
    LEACH, MO
    HIND, A
    SAUTER, R
    REQUARDT, H
    WEBER, H
    [J]. MEDICAL PHYSICS, 1986, 13 (04) : 510 - 513
  • [10] A TECHNIQUE OF DOUBLE-RESONANT OPERATION OF F-19 AND H-1 QUADRATURE BIRDCAGE COILS
    LU, DF
    JOSEPH, PM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 19 (01) : 180 - 185