摘要
Researchers have hypothesized that the increased revision rate of LHMoM (Large Head Metal-on-Metal) hips compared to MoM hip resurfacings may be attributed to corrosion and wear at the head neck taper junction. Studies have reported visual evidence of fretting and corrosion at the taper junction, but no method has been described in the literature to quantify the amount of material lost from the taper junction. This paper describes a measurement protocol using a Taylor Hobson Talyrond Roundness instrument that allows the simultaneous measurement of surface form (wear) and surface topography (roughness). The methodology allows the measurement of the taper angle, geometry of worn region (depth, length), 3D surface maps and surface topography of the head conical taper. The accurate quantification of the taper geometry, wear and topography is essential to the understanding of the in vivo wear and corrosion mechanisms of taper junctions in LHMoM hips.
引用
收藏
页码:209 / 234
页数:26
相关论文
共 14 条
[1]  
[Anonymous], 2010, 7 ENGL WAL NAT JOINT
[2]   IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[3]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[4]  
Hart A., 2012, ANN M AM AC ORTH SUR
[5]  
Higgs G. B., 2012, ASTM S MET ON MET TO
[6]   Metal-on-metal bearings surfaces: materials, manufacture, design, optimization, and alternatives [J].
Isaac, GH ;
Thompson, J ;
Williams, S ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :119-133
[7]   Taper junction failure in large-diameter metal-on-metal bearings [J].
Langton, D. J. ;
Sidaginamale, R. ;
Lord, J. K. ;
Nargol, A. V. F. ;
Joyce, T. J. .
BONE & JOINT RESEARCH, 2012, 1 (04) :56-63
[8]   Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses [J].
Lord, J. K. ;
Langton, D. J. ;
Nargol, A. V. F. ;
Joyce, T. J. .
WEAR, 2011, 272 (01) :79-87
[9]   Impingement with total hip replacement [J].
Malik, Aamer ;
Maheshwari, Aditya ;
Dorr, Lawrence D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (08) :1832-1842
[10]   Retrieval analysis of 240 metal-on-metal hip components, comparing modular total hip replacement with hip resurfacing [J].
Matthies, A. ;
Underwood, R. ;
Cann, P. ;
Ilo, K. ;
Nawaz, Z. ;
Skinner, J. ;
Hart, A. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (03) :307-314