Long-term outcome results after endovascular aortoiliac aneurysm repair with the bifurcated EXCLUDER Endoprosthesis

被引:5
作者
Van Gool, Frederick [1 ]
Houthoofd, Sabrina [2 ]
Mufty, Hozan [2 ]
Bonne, Lawrence [1 ]
Fourneau, Inge [2 ]
Maleux, Geert [1 ]
机构
[1] Univ Hosp Leuven, Dept Radiol, Herestr 49, B-3000 Leuven, Belgium
[2] Univ Hosp Leuven, Dept Vasc Surg, Leuven, Belgium
关键词
Endovascular aortic repair; Excluder stent graft; Outcome; Long-term; ABDOMINAL AORTIC-ANEURYSM; SINGLE-CENTER EXPERIENCE; STENT-GRAFT; REGISTRY; ENDOGRAFT; BEHAVIOR; TRIALS; SAC;
D O I
10.1016/j.jvs.2021.09.042
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: To report the long-term outcome of patients presenting with an aortic, aortoiliac, or isolated common iliac aneurysm treated with the bifurcated EXCLUDER Endoprosthesis. Furthermore, potential differences in late outcome results between the original- and low-permeability endoprosthesis were analyzed. Methods: A retrospective analysis of prospectively collected data of 182 patients who underwent endovascular aneurysm repair with the EXCLUDER Endoprosthesis between June 1998 and October 2015 in an academic, tertiary care center for aortic disease was performed. Patient follow-up was from 3 to 20 years (mean follow-up of 6.9 years). Primary end points were overall survival and reintervention-free survival. Secondary end points were device-related complications, endoleaks, and reinterventions. Results: Overall survival at 5, 10, and 15 years was 72.8%, 42.1%, and 12.2%, respectively, with no aneurysm-related mortality and no difference in overall survival between the original- vs low-permeability endoprosthesis group (P = .617). Freedom from type I endoleak at 5 years was 94.8%. No new type I endoleak was detected beyond the 5-year follow-up mark. No type III endoleak was identified. Reintervention-free survival was 83.6%, 66.7%, and 66.7% at 5-, 10-, and 15-year follow-up, respectively. There was a significant difference in intervention-free survival between the original- vs low-permeability endoprosthesis group (P = .029) and after the 5-year follow-up mark. In addition, patients with the low-permeability endoprosthesis showed significantly fewer device-related complications (P = .002) and endoleaks (P = .005). Conclusions: Endovascular aneurysm repair using the EXCLUDER Endoprosthesis is effective and durable on long-term follow-up, with acceptably low device-related complications and reinterventions. The low-permeability endoprosthesis was associated with significantly fewer new device-related complications and endoleaks after 5 years of follow-up.
引用
收藏
页码:1882 / +
页数:10
相关论文
共 27 条
  • [1] Bos WTGJ, 2009, J CARDIOVASC SURG, V50, P159
  • [2] The Society for Vascular Surgery practice guidelines on the care of patients with an abdominal aortic aneurysm
    Chaikof, Elliot L.
    Dalman, Ronald L.
    Eskandari, Mark K.
    Jackson, Benjamin M.
    Lee, W. Anthony
    Mansour, M. Ashraf
    Mastracci, Tara M.
    Mell, Matthew
    Murad, M. Hassan
    Nguyen, Louis L.
    Oderich, Gustavo S.
    Patel, Madhukar S.
    Schermerhorn, Marc L.
    Starnes, Benjamin W.
    [J]. JOURNAL OF VASCULAR SURGERY, 2018, 67 (01) : 2 - +
  • [3] Late abdominal aortic aneurysm enlargement after endovascular repair with the excluder device
    Cho, JS
    Dillavou, ED
    Rhee, RY
    Makaroun, AS
    [J]. JOURNAL OF VASCULAR SURGERY, 2004, 39 (06) : 1236 - 1241
  • [4] Long-Term Outcome of Open or Endovascular Repair of Abdominal Aortic Aneurysm
    De Bruin, Jorg L.
    Baas, Annette F.
    Buth, Jaap
    Prinssen, Monique
    Verhoeven, Eric L. G.
    Cuypers, Philippe W. M.
    van Sambeek, Marc R. H. M.
    Balm, Ron
    Grobbee, Diederick E.
    Blankensteijn, Jan D.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (20) : 1881 - 1889
  • [5] Three-dimensional analysis of enlarging aneurysms after endovascular abdominal aortic aneurysm repair in the Gore Excluder Pivotal clinical trial
    Fillinger, M
    [J]. JOURNAL OF VASCULAR SURGERY, 2006, 43 (05) : 888 - 895
  • [6] Ghotbi R, 2010, J CARDIOVASC SURG, V51, P475
  • [7] Decrease in total aneurysm-related deaths in the era of endovascular aneurysm repair
    Giles, Kristina A.
    Pomposelli, Frank
    Hamdan, Allen
    Wyers, Mark
    Jhaveri, Ami
    Schermerhorn, Marc L.
    [J]. JOURNAL OF VASCULAR SURGERY, 2009, 49 (03) : 543 - 549
  • [8] Adequate seal and no endoleak on the first postoperative computed tomography angiography as criteria for no additional imaging up to 5 years after endovascular aneurysm repair
    Goncalves, Frederico Bastos
    van de Luijtgaarden, Koen M.
    Hoeks, Sanne E.
    Hendriks, Johanna M.
    ten Raa, Sander
    Rouwet, Ellen V.
    Stolker, Robert J.
    Verhagen, Hence J. M.
    [J]. JOURNAL OF VASCULAR SURGERY, 2013, 57 (06) : 1503 - 1511
  • [9] Clinical outcome and morphologic analysis after endovascular aneurysm repair using the Excluder endograft
    Goncalves, Frederico Bastos
    Jairam, An
    Voute, Michiel T.
    Moelker, Adriaan D.
    Rouwet, Ellen V.
    ten Raa, Sander
    Hendriks, Johanna M.
    Verhagen, Hence J. M.
    [J]. JOURNAL OF VASCULAR SURGERY, 2012, 56 (04) : 920 - 928
  • [10] Sac behavior after aneurysm treatment with the Gore Excluder low-permeability aortic endoprosthesis: 12-month comparison to the original Excluder device
    Haider, Shan-e-ali
    Najjar, Samer F.
    Cho, Jae-Sung
    Rhee, Robert Y.
    Eskandari, Mark K.
    Matsumura, Jon S.
    Makaroun, Michel S.
    Morasch, Mark D.
    [J]. JOURNAL OF VASCULAR SURGERY, 2006, 44 (04) : 694 - 700