Measuring trihalomethane concentrations in water using supported capillary membrane sampling-gas chromatography

被引:15
|
作者
Emmert, GL [1 ]
Cao, G
Duty, C
Wolcott, W
机构
[1] Univ Memphis, Dept Chem, Memphis, TN 38152 USA
[2] Arkansas State Univ, Dept Chem & Phys, State Univ, AR 72467 USA
[3] Global FIA, Fox Isl, WA USA
关键词
water analysis; trihalomethanes; disinfection by-products; gas chromatography;
D O I
10.1016/j.talanta.2003.12.008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The objective of this study was to develop and evaluate a supported capillary membrane sampling-gas chromatography method for the analysis of trihalomethanes (THMs) in drinking water. The effects of experimental parameters, such as flow rate of carrier gas, water temperature, ionic strengths of solutions and transfer line temperature on the system performance were investigated. The results of method detection limit and accuracy and precision studies are reported. (C) 2003 Elsevier B.V. All rights reserved.