Filtration Bleb Revision Guided by 3-Dimensional Anterior Segment Optical Coherence Tomography

被引:10
作者
Kojima, Sachi [1 ]
Inoue, Toshihiro [1 ]
Kawaji, Takahiro [1 ]
Tanihara, Hidenobu [1 ]
机构
[1] Kumamoto Univ, Dept Ophthalmol, Fac Life Sci, Kumamoto 8608556, Japan
关键词
anterior segment optical coherence tomography; bleb revision; trabeculectomy; TRABECULECTOMY;
D O I
10.1097/IJG.0b013e3182741ee6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To elucidate the usefulness of 3-dimensional anterior segment optical coherence tomography (3D AS-OCT) for bleb revision after trabeculectomy. Design: Case series. Participants: Two patients who underwent trabeculectomy and subsequent bleb revision. Methods: Three-dimensional AS-OCT and intraoperative observation. Main Outcome Measures: AS-OCT and intraoperative observation were analyzed to identify the filtration openings on the scleral flap margin and leaking points. Results: In 1 case with a leaking bleb, 3D AS-OCT examinations suggested the presence of numerous internal ostia (filtration openings). With 3D AS-OCT guidance, we carefully removed the leaking bleb with peripheral scar tissue. Histopathologic findings showed the presence of a thin bleb wall in the limbal conjunctiva corresponding to the point of aqueous leakage. In another case of an overhanging bleb, 3D AS-OCT examination revealed a thin bleb wall and multiloculated cystic structures. Histopathologic findings showed the presence of a thin region in the conjunctival epithelium in which connective tissue did not support the conjunctival epithelium, which seemed to correspond to the point at which oozing occurred. Conclusions: Three-dimensional AS-OCT enables evaluation of the internal morphology of filtration blebs and is useful for preoperative guidance for bleb revision.
引用
收藏
页码:312 / 315
页数:4
相关论文
共 5 条
[1]   Clinical characteristics and leakage of functioning blebs after trabeculectomy with mitomycin-C in primary glaucoma patients [J].
Hu, CY ;
Matsuo, H ;
Tomita, G ;
Suzuki, Y ;
Araie, M ;
Shirato, S ;
Tanaka, S .
OPHTHALMOLOGY, 2003, 110 (02) :345-352
[2]   Evaluation of Trabeculectomy Blebs Using 3-Dimensional Cornea and Anterior Segment Optical Coherence Tomography [J].
Kawana, Keisuke ;
Kiuchi, Takahiro ;
Yasuno, Yoshiaki ;
Oshika, Tetsuro .
OPHTHALMOLOGY, 2009, 116 (05) :848-855
[3]   Anterior segment optical coherence tomography imaging and histopathologic findings of an overhanging filtering bleb [J].
Kim, W. K. ;
Seong, G. J. ;
Lee, C. S. ;
Kim, Y. G. ;
Kim, C. Y. .
EYE, 2008, 22 (12) :1520-1521
[4]   Analysis of bleb morphology after trabeculectomy with Visante anterior segment optical coherence tomography [J].
Leung, Christopher Kai-shun ;
Yick, Doris Wai-fong ;
Kwong, Yolanda Yuen-ying ;
Li, Felix Chi-hong ;
Leung, Dexter Yu-lung ;
Mohamed, Shaheeda ;
Tham, Clement Chee-yung ;
Chung-chai, Chi ;
Lam, Dennis Shun-Chiu .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (03) :340-344
[5]   Imaging of trabeculectomy blebs using anterior segment optical coherence tomography [J].
Singh, Mandeep ;
Chew, Paul T. K. ;
Friedman, David S. ;
Nolan, Winifred P. ;
See, Jovina L. ;
Smith, Scott D. ;
Zheng, Ce ;
Foster, Paul J. ;
Aung, Tin .
OPHTHALMOLOGY, 2007, 114 (01) :47-53