Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients - A prospective study with 115 examinations

被引:200
作者
Sommer, Torsten
Naehle, Claas P.
Yang, Alexander
Zeijlemaker, Volkert
Hackenbroch, Matthias
Schmiedel, Alexandra
Meyer, Carsten
Strach, Katharina
Skowasch, Dirk
Vahlhaus, Christian
Litt, Harold
Schild, Hans
机构
[1] Univ Bonn, Dept Radiol, Cardiovasc Imaging Sect, D-53127 Bonn, Germany
[2] Univ Bonn, Dept Cardiol, D-53127 Bonn, Germany
[3] Medtron Bakker Res Ctr, Maastricht, Netherlands
[4] Hosp Univ Muenster, Dept Cardiol & Angiol, Munster, Germany
[5] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
关键词
magnetic resonance imaging; pacemakers; safety; imaging;
D O I
10.1161/CIRCULATIONAHA.105.597013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers. Methods and Results-Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was < 60 bpm, the asynchronous mode was programmed to avoid magnetic resonance (MR)-induced inhibition; if heart rate was > 60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI (P = 0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT. Conclusions-Extrathoracic MRI of non-pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR-and pacemaker-related precautions.
引用
收藏
页码:1285 / 1292
页数:8
相关论文
共 26 条
  • [1] Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes
    Achenbach, S
    Moshage, W
    Diem, B
    Bieberle, T
    Schibgilla, V
    Bachmann, K
    [J]. AMERICAN HEART JOURNAL, 1997, 134 (03) : 467 - 473
  • [2] ARMOUR EP, 1993, CANCER RES, V53, P2740
  • [3] ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING
    BOTTOMLEY, PA
    REDINGTON, RW
    EDELSTEIN, WA
    SCHENCK, JF
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) : 336 - 349
  • [4] Markers of myocardial damage, tissue healing, and inflammation after radiofrequency catheter ablation of atrial tachyarrhythmias
    Brueckmann, M
    Wolpert, C
    Bertsch, T
    Sueselbeck, T
    Liebetrau, C
    Kaden, JJ
    Huhle, G
    Neumaier, M
    Borggrefe, M
    Haase, KK
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (06) : 686 - 691
  • [5] Chou CK, 1996, BIOELECTROMAGNETICS, V17, P195, DOI 10.1002/(SICI)1521-186X(1996)17:3<195::AID-BEM5>3.0.CO
  • [6] 2-Z
  • [7] Government viewpoint: US food & drug administration: Pacemakers, ICDs and MRI
    Faris, OP
    Shein, MJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04): : 268 - 269
  • [8] Can patients with implantable pacemakers safely undergo magnetic resonance imaging?
    Gimbel, JR
    Kanal, E
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) : 1325 - 1327
  • [9] HENRIQUES FC, 1947, ARCH PATHOL, V43, P489
  • [10] Transesophageal cardiac pacing during magnetic resonance imaging: Feasibility and safety considerations
    Hofman, MBM
    deCock, CC
    vanderLinden, JC
    vanRossum, AC
    Visser, FC
    Sprenger, M
    Westerhof, N
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) : 413 - 422