Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients - A prospective study with 115 examinations

被引:199
作者
Sommer, Torsten
Naehle, Claas P.
Yang, Alexander
Zeijlemaker, Volkert
Hackenbroch, Matthias
Schmiedel, Alexandra
Meyer, Carsten
Strach, Katharina
Skowasch, Dirk
Vahlhaus, Christian
Litt, Harold
Schild, Hans
机构
[1] Univ Bonn, Dept Radiol, Cardiovasc Imaging Sect, D-53127 Bonn, Germany
[2] Univ Bonn, Dept Cardiol, D-53127 Bonn, Germany
[3] Medtron Bakker Res Ctr, Maastricht, Netherlands
[4] Hosp Univ Muenster, Dept Cardiol & Angiol, Munster, Germany
[5] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
关键词
magnetic resonance imaging; pacemakers; safety; imaging;
D O I
10.1161/CIRCULATIONAHA.105.597013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers. Methods and Results-Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was < 60 bpm, the asynchronous mode was programmed to avoid magnetic resonance (MR)-induced inhibition; if heart rate was > 60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI (P = 0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT. Conclusions-Extrathoracic MRI of non-pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR-and pacemaker-related precautions.
引用
收藏
页码:1285 / 1292
页数:8
相关论文
共 26 条
[1]   Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]  
ARMOUR EP, 1993, CANCER RES, V53, P2740
[3]   ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING [J].
BOTTOMLEY, PA ;
REDINGTON, RW ;
EDELSTEIN, WA ;
SCHENCK, JF .
MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) :336-349
[4]   Markers of myocardial damage, tissue healing, and inflammation after radiofrequency catheter ablation of atrial tachyarrhythmias [J].
Brueckmann, M ;
Wolpert, C ;
Bertsch, T ;
Sueselbeck, T ;
Liebetrau, C ;
Kaden, JJ ;
Huhle, G ;
Neumaier, M ;
Borggrefe, M ;
Haase, KK .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (06) :686-691
[5]  
Chou CK, 1996, BIOELECTROMAGNETICS, V17, P195, DOI 10.1002/(SICI)1521-186X(1996)17:3<195::AID-BEM5>3.0.CO
[6]  
2-Z
[7]   Government viewpoint: US food & drug administration: Pacemakers, ICDs and MRI [J].
Faris, OP ;
Shein, MJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04) :268-269
[8]   Can patients with implantable pacemakers safely undergo magnetic resonance imaging? [J].
Gimbel, JR ;
Kanal, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) :1325-1327
[9]  
HENRIQUES FC, 1947, ARCH PATHOL, V43, P489
[10]   Transesophageal cardiac pacing during magnetic resonance imaging: Feasibility and safety considerations [J].
Hofman, MBM ;
deCock, CC ;
vanderLinden, JC ;
vanRossum, AC ;
Visser, FC ;
Sprenger, M ;
Westerhof, N .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :413-422