Material optimization of a cemented tibia tray using functionally graded material

被引:6
作者
Hedia, Hassan S. [1 ]
Aldousari, Saad M. [2 ]
Fouda, Noha [3 ]
机构
[1] King Abdulaziz Univ, Mat & Solid Mech, Jeddah 21589, Saudi Arabia
[2] King Abdulaziz Univ, Jeddah 21589, Saudi Arabia
[3] Mansoura Univ, Fac Engn, Mansoura, Egypt
关键词
Functionally graded material; titanium; bioglass; finite element analysis; TOTAL KNEE ARTHROPLASTY; STRESS-DISTRIBUTION; DENTAL IMPLANT; DESIGN; COMPONENT; FIXATION; DENSITY; REDUCE; SHAPE; PAIN;
D O I
10.3139/120.110842
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Joint replacement surgeries are doubtlessly the gift of science and technology for human welfare. In the cemented knee replacement joint, the stiffer implant carries the majority of the load, which is actually carried by the bone itself before implantation. The resulting implant induced stress-shielding and subsequent bone remodeling causes bone resorption at the proximal bone under the tibia tray. The aim of this study is to optimize the material of the cemented tibia tray using functionally graded material FGM. The target is to find the optimal material and the optimal gradation direction for the cemented tibia tray. The results showed that the optimal vertical FGM changed from titanium at the lower tibia stem to bioglass at the upper layers of the tibia tray. The vertical FGM reduced the stress shielding by 56 and 80 % in cancellous epiphyseal and diaphyseal bone, respectively, compared to cemented titanium tibia tray. However, the optimal horizontal FGM changes from titanium at the stem core to bioglass at the rim of tibia tray layers. This horizontal gradation reduced the stress shielding in cancellous epiphyseal bone by 62 %. However, the stress shielding in cancellous diaphyseal bone did not change compared to cemented titanium tibia tray.
引用
收藏
页码:260 / 268
页数:9
相关论文
共 34 条
[1]   Peri-prosthetic bone mineral density after total knee arthroplasty -: Cemented versus cementless fixation [J].
Abu-Rajab, R. B. ;
Watson, W. S. ;
Walker, B. ;
Roberts, J. ;
Gallacher, S. J. ;
Meek, R. M. D. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (05) :606-613
[2]   A parametric analysis of fixation post shape in tibial knee prostheses [J].
Au, AG ;
Liggins, AB ;
Raso, VJ ;
Amirfazli, A .
MEDICAL ENGINEERING & PHYSICS, 2005, 27 (02) :123-134
[3]   Material tailoring of the femoral component in a total knee replacement to reduce the problem of aseptic loosening [J].
Bahraminasab, Marjan ;
Sahari, B. B. ;
Edwards, K. L. ;
Farahmand, Farzam ;
Hong, Tang Sai ;
Naghibi, Hamid .
MATERIALS & DESIGN, 2013, 52 :441-451
[4]   Tibial component fixation in total knee arthroplasty - A comparison of pegged and stemmed designs [J].
Bertin, Kim C. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (05) :670-678
[5]   Full and surface tibial cementation in total knee arthroplasty: A biomechanical investigation of stress distribution and remodeling in the tibia [J].
Cawley, Derek T. ;
Kelly, Nicola ;
Simpkin, Andrew ;
Shannon, Fintan J. ;
McGarry, J. Patrick .
CLINICAL BIOMECHANICS, 2012, 27 (04) :390-397
[6]   THE INFLUENCE OF TIBIAL PROSTHESIS DESIGN FEATURES ON STRESSES RELATED TO ASEPTIC LOOSENING AND STRESS SHIELDING [J].
Chong, Desmond Y. R. ;
Hansen, Ulrich N. ;
Amis, Andrew A. .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2011, 11 (01) :55-72
[7]   A new press-fit stem concept to reduce the risk of end-of-stem pain at revision TKA: A pre-clinical study [J].
Completo, A. ;
Fonseca, F. ;
Simoes, J. A. ;
Ramos, A. ;
Relvas, C. .
KNEE, 2012, 19 (05) :537-542
[8]  
Dargahi J, 2003, BIO-MED MATER ENG, V13, P439
[9]   The effect of cement stiffness and tibia tray material on the stresses developed in artificial knee [J].
Darwish, S. M. ;
Al-Samhan, A. .
INTERNATIONAL JOURNAL OF ADHESION AND ADHESIVES, 2008, 28 (03) :120-125
[10]  
Gatti A. M., 1998, BIOCERAMICS, V11, P349