THE EFFECT OF ASPHERICITY OF ACETABULAR BEARING SURFACE ON CONTACT MECHANICS OF UHMWPE TOTAL HIP IMPLANTS BY FINITE ELEMENT ANALYSIS

被引:1
作者
Zhang, Xuan [1 ]
Wang, Ling [1 ]
Peng, Xifeng [1 ]
Li, Dichen [1 ]
He, Jiankang [1 ]
Liu, Yaxiong [1 ]
Lian, Qin [1 ]
Jin, Zhongmin [1 ,2 ]
Liu, Chaozong [3 ]
机构
[1] Xi An Jiao Tong Univ, State Key Lab Mfg Syst Engn, Xian 710049, Peoples R China
[2] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[3] UCL, Royal Natl Orthopaed Hosp, John Scales Ctr Biomed Engn, London HA7 4LP, England
基金
中国国家自然科学基金;
关键词
Total hip arthroplasty; asphericity; contact mechanics; finite element analysis; edge loading; MOLECULAR-WEIGHT POLYETHYLENE; METAL-ON-METAL; WEAR DEBRIS; REPLACEMENTS; ARTHROPLASTY; CUPS; PROSTHESES; RATES;
D O I
10.1142/S0219519417500117
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Asphericity and out-of-roundness are generally used to evaluate the manufacturing quality of ultra-high molecular weight polyethylene (UHMWPE) cup inner surfaces, which can potentially affect initial clinical wear and contribute to osteolysis of total hip arthroplasty. This study measured the location and magnitude of asphericity and the out-of-roundness value for four UHMWPE cups in a single set, and then investigated the effects of the asphericity on the contact mechanics of UHMWPE cups. A co-ordinate measuring machine (CMM) was used for the surface measurement and finite element analysis (FEA) was adopted for contact mechanics study. The results demonstrated that the asphericity varied between cups with the maximum value as 0.088 +/- 0.004 mm. Although such a value met the ISO specification, large difference of volume appeared for the asphericity above 0.060 mm. Actual surface profile accounting for the asphericity was found to affect the value of contact pressure and contact area by around 12%. The inferior asphericity resulted in a nonsmoothly distributed contact pressure, which had a negative effect on the contact mechanics of UHMWPE cups and the edge loading was predicted to occur for the sample with a large asphericity. In conclusion, the asphericity of UHMWPE cup could affect the contact mechanics of the articular bearings and may subsequently contribute to initial wear during bedding-in phase.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Analysis of contact mechanics in McKee-Farrar metal-on-metal hip implants
    Yew, A
    Jagatia, M
    Ensaff, H
    Jin, ZM
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H5) : 333 - 340
  • [22] Long-term wear failure analysis of uhmwpe acetabular cup in total hip replacement
    Shahemi, N.
    Liza, S.
    Abbas, A. A.
    Merican, A. M.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 87 : 1 - 9
  • [23] Patient-specific finite element analysis of femurs with cemented hip implants
    Katz, Yekutiel
    Lubovsky, Omri
    Yosibash, Zohar
    [J]. CLINICAL BIOMECHANICS, 2018, 58 : 74 - 89
  • [24] The influence of contact ratio and its location on the primary stability of cementless total hip arthroplasty: A finite element analysis
    Reimeringer, M.
    Nuno, N.
    [J]. JOURNAL OF BIOMECHANICS, 2016, 49 (07) : 1064 - 1070
  • [25] Study the effect of wear rate on impingement failure of an acetabular liner surface based on finite element analysis
    Saputra, Eko
    Anwar, Iwan Budiwan
    van der Heide, Emile
    Ismail, Rifky
    Jamari, J.
    [J]. INTERNATIONAL JOURNAL OF MATERIALS & PRODUCT TECHNOLOGY, 2017, 55 (04) : 340 - 353
  • [26] Finite Element Analysis of Proximal Cement Fixation in Total Hip Arthroplasty
    Ahmad, Maizatul Afirah
    Zulkifli, Nurul Nadhirah Mohamed Elias
    Shuib, Solehuddin
    Sulaiman, Shahrul Hisham
    Abdullah, Abdul Halim
    [J]. INTERNATIONAL JOURNAL OF TECHNOLOGY, 2020, 11 (05) : 1046 - 1055
  • [27] Comparative Finite Element Analysis of the Debonding Process in Different Concepts of Cemented Hip Implants
    Perez, M. A.
    Palacios, J.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (06) : 2093 - 2106
  • [28] Finite element submodeling technique to analyze the contact pressure and wear of hard bearing couples in hip prosthesis
    Shankar, S.
    Nithyaprakash, R.
    Santhosh, B. R.
    Uddin, M. S.
    Pramanik, Alokesh
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2020, 23 (08) : 422 - 431
  • [29] Segmental acetabular rim defects, bone loss, oversizing, and press fit cup in total hip arthroplasty evaluated with a probabilistic finite element analysis
    Amirouche, Farid
    Solitro, Giovanni F.
    Walia, Amit
    Gonzalez, Mark
    Bobko, Aimee
    [J]. INTERNATIONAL ORTHOPAEDICS, 2017, 41 (08) : 1527 - 1533
  • [30] Contact mechanics and elastohydrodynamic lubrication in a novel metal-on-metal hip implant with an aspherical bearing surface
    Meng, Qingen
    Gao, Leiming
    Liu, Feng
    Yang, Peiran
    Fisher, John
    Jin, Zhongmin
    [J]. JOURNAL OF BIOMECHANICS, 2010, 43 (05) : 849 - 857