Biomechanical analysis of a radial functionally graded dental implant-bone system under multi-directional dynamic loads

被引:12
作者
Jarrahi, A. [1 ]
Shirazi, H. Asgharzadeh [1 ,2 ]
Asnafi, A. [3 ]
Ayatollahi, M. R. [2 ]
机构
[1] Iran Univ Sci & Technol, Sch Mech Engn, Tissue Engn & Biol Syst Res Lab, Tehran 1684613114, Iran
[2] Iran Univ Sci & Technol, Sch Mech Engn, Ctr Excellence Expt Solid Mech & Dynam, Fatigue & Fracture Res Lab, Tehran, Iran
[3] Shiraz Univ, Hydroaeronaut Res Ctr, Shiraz 7134813668, Iran
关键词
Dental implant system; Functionally graded biomaterials; Dynamic behavior; Finite element analysis; 3-DIMENSIONAL FINITE-ELEMENT; NANO-HYDROXYAPATITE COATINGS; STRESS-DISTRIBUTION; QUALITY; BIOCOMPATIBILITY; TITANIUM; DESIGN;
D O I
10.1007/s40430-018-1166-9
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Development of new biomaterials such as FGBMs for dental clinical implant applications have been recently attracted a lot of attention due to their advantages and abilities in satisfying both mechanical and biocompatibility properties simultaneously. Drawing inspiration from the geometry of a real dental implant system, and using the properties of radial functionally graded biomaterials, a model was investigated to fabricate the artificial FGBM dental implant under periodic dynamic loads. The investigation is done using a finite element analyzer, and the problem is solved under the actual bone geometry, constraints, conditions and combined dynamic loads. The effects of radial FGBM dental implants on the stress and deformation values near the dental implant-bone interface under periodic dynamic loads are simulated. Response characteristics of FGBM implant models with different heterogeneous biomaterial parameters are determined and compared to the behavior of conventional titanium and titanium coated by a film of hydroxyapatite implants. The results show that using radial FGBM dental implants can improve the mechanical behaviors of dental implant systems regarding stress and deformation. It was also found that FGBM parameters can play an essential role in achieving an efficient biomechanical function in those systems.
引用
收藏
页数:13
相关论文
共 34 条
  • [1] Optimal design of thread height and width on an immediately loaded cylinder implant: A finite element analysis
    Ao, Jianhua
    Li, Tao
    Liu, Yanpu
    Ding, Yin
    Wu, Guofeng
    Hu, Kaijin
    Kong, Liang
    [J]. COMPUTERS IN BIOLOGY AND MEDICINE, 2010, 40 (08) : 681 - 686
  • [2] Mechanical and tribological properties of hydroxyapatite nanoparticles extracted from natural bovine bone and the bone cement developed by nano-sized bovine hydroxyapatite filler
    Ayatollahi, M. R.
    Yahya, Mohd Yazid
    Shirazi, H. Asgharzadeh
    Abu Hassan, Shukur
    [J]. CERAMICS INTERNATIONAL, 2015, 41 (09) : 10818 - 10827
  • [3] Finite element modelling of implant designs and cortical bone thickness on stress distribution in maxillary type IV bone
    Chou, I-Chiang
    Lee, Shyh-Yuan
    Wu, Ming-Chang
    Sun, Chia-Wei
    Jiang, Cho-Pei
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (05) : 516 - 526
  • [4] FE study of bone quality effect on load-carrying ability of dental implants
    Demenko, V.
    Linetsky, I.
    Nesvit, V.
    Linetska, L.
    Shevchenko, A.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (16) : 1751 - 1761
  • [5] Importance of diameter-to-length ratio in selecting dental implants: a methodological finite element study
    Demenko, V.
    Linetskiy, I.
    Nesvit, K.
    Hubalkova, H.
    Nesvit, V.
    Shevchenko, A.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (04) : 443 - 449
  • [6] Analysis of the effect of load direction on the stress distribution in dental implant
    Djebbar, N.
    Serier, B.
    Bouiadjra, B. Bachir
    Benbarek, S.
    Drai, A.
    [J]. MATERIALS & DESIGN, 2010, 31 (04) : 2097 - 2101
  • [7] Influence of implant shape, surface morphology, surgical technique and bone quality on the primary stability of dental implants
    Elias, Carlos Nelson
    Rocha, Felipe Assis
    Nascimento, Ana Lucia
    Coelho, Paulo Guilherme
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 16 : 169 - 180
  • [8] Load transfer along the bone-dental implant interface
    Faegh, Samira
    Mueftue, Sinan
    [J]. JOURNAL OF BIOMECHANICS, 2010, 43 (09) : 1761 - 1770
  • [9] Modification of electrophoretically deposited nano-hydroxyapatite coatings by wire brushing on Ti-6Al-4V substrates
    Farnoush, Hamidreza
    Mohandesi, Jamshid Aghazadeh
    Fatmehsari, Davoud Haghshenas
    Moztarzadeh, Fathollah
    [J]. CERAMICS INTERNATIONAL, 2012, 38 (06) : 4885 - 4893
  • [10] Investigation of the dental implant geometry effect on stress distribution at dental implant-bone interface
    Ghadiri, Majid
    Shafiei, Navvab
    Salekdeh, Sadegh Hemmati
    Mottaghi, Parya
    Mirzaie, Tahereh
    [J]. JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2016, 38 (02) : 335 - 343