Efficacy of eluted antibiotics through 3D printed femoral implants

被引:26
作者
Benmassaoud, Mohammed Mehdi [1 ]
Kohama, Christopher [2 ]
Kim, Tae Won B. [3 ]
Kadlowec, Jennifer A. [1 ,2 ]
Foltiny, Brandon [1 ]
Mercurio, Theo [1 ]
Ranganathan, Shivakumar I. [4 ]
机构
[1] Rowan Univ, Dept Mech Engn, 201 Mullica Hill Rd, Glassboro, NJ 08028 USA
[2] Rowan Univ, Dept Biomed Engn, 201 Mullica Hill Rd, Glassboro, NJ 08028 USA
[3] Cooper Univ, Dept Orthopaed Surg, Hlth Care, 3 Cooper Plaza, Camden, NJ 08103 USA
[4] Virginia Tech, Dept Mech Engn, 7054 Haycock Rd, Falls Church, VA 22043 USA
关键词
3D printing; PLA; PCL; Ti6Al4V; Doxycycline; Drug elution; GENTAMICIN-PMMA BEADS; 2-STAGE REIMPLANTATION; ANTIMICROBIAL EFFICACY; MECHANICAL-PROPERTIES; CONTROLLED-RELEASE; BONE-CEMENT; STAPHYLOCOCCUS; OSTEOMYELITIS; VANCOMYCIN; RESISTANT;
D O I
10.1007/s10544-019-0395-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Costs associated with musculoskeletal diseases in the United States account for 5.7% of the Gross Domestic Product (GDP) (Weinstein et al. 2018). As such, there is a need to pursue new ideas in orthopaedic implants that can decrease cost and improve patient care. In the recent years, 3D printing of polymers using Fused Deposition Modeling (FDM) and metals using Direct Metal Laser Sintering (DMLS) has opened several exciting possibilities to create customized orthopaedic implants. Such implants can be engineered to release antibiotics in a controlled manner by infusing the drug into the material during manufacturing stage. However, the prevalence of high temperature could impact the anti-bacterial effectiveness of the eluted antibiotics in such implants. An alternative approach to circumvent this issue would be to modify the implant geometry to incorporate built-in design features such as micro-channels and reservoirs in which antibiotics can be introduced prior to the surgical procedure. Irrespective of the approach used, the ability of 3D printed orthopaedic implants to elute antibiotics, and the rate of elution are not well understood. The purpose of this article is to study the elution of doxycycline through 3D printed femoral implants using three different materials: Poly-Lactic Acid (PLA), Poly-Caprolactone (PCL) and Titanium grade Ti-6Al-4V. The PLA and Ti-6Al-4V implants were designed with built-in reservoirs and micro-channels in which doxycycline was introduced post the manufacturing stage. However, the PCL implants were printed from a PCL spool that was infused with doxycycline using an extruder. The PLA and Ti-6Al-4V experiments were run for a period of 31 days and the PCL experiment for one day. The antibacterial ability of eluted doxycycline from all implants were examined using Kirby-Bauer test on the bacteria E.coli k-12. The results show that most of doxycycline eluted through the three materials in the first 24 hours. After the initial spike, a steady release was achieved for the PLA and Ti-6Al-4V implants for 30 days. During this timeframe, Ti-6Al-4V implants released more doxycycline than the PLA implant. The eluted antibiotics through all the implants demonstrated the ability to kill bacteria in the subsequent Kirby-Bauer test. These outcomes show that irrespective of how the antibiotics were introduced, 3D printed polymeric and metallic implants have great potential in orthopaedic applications.
引用
收藏
页数:10
相关论文
共 44 条
[1]   Local delivery of a zoledronate solution improves osseointegration of titanium implants in a rat distal femur model [J].
AbuMoussa, Samuel ;
Ruppert, David S. ;
Lindsay, Christopher ;
Dahners, Laurence ;
Weinhold, Paul .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2018, 36 (12) :3294-3298
[2]   Controlled Release of Vancomycin from Thin Sol-Gel Films on Implant Surfaces Successfully Controls Osteomyelitis [J].
Adams, Christopher S. ;
Antoci, Valentin, Jr. ;
Harrison, Gerald ;
Patal, Payal ;
Freeman, Terry A. ;
Shapiro, Irving M. ;
Parvizi, Javad ;
Hickok, Noreen J. ;
Radin, Shula ;
Ducheyne, Paul .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2009, 27 (06) :701-709
[3]   Persistence of bacterial growth on antibiotic-loaded beads - Is it actually a problem? [J].
Anagnostakos, Konstantinos ;
Hitzler, Philip ;
Pape, Dietrich ;
Kohn, Dieter ;
Kelm, Jens .
ACTA ORTHOPAEDICA, 2008, 79 (02) :302-307
[4]   Enhancement of Antibiotic Elution From Acrylic Bone Cement [J].
Anagnostakos, Konstantinos ;
Kelm, Jens .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 90B (01) :467-475
[5]   Determination of minimum inhibitory concentrations [J].
Andrews, JM .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2001, 48 :5-16
[6]  
BLAHA JD, 1993, CLIN ORTHOP RELAT R, P8
[7]   Staphylococcus aureus prosthetic joint infection treated with prosthesis removal and delayed reimplantation arthroplasty [J].
Brandt, CM ;
Duffy, MCT ;
Berbari, EF ;
Hanssen, AD ;
Steckelberg, JM ;
Osmon, DR .
MAYO CLINIC PROCEEDINGS, 1999, 74 (06) :553-558
[8]   Nanobiomaterial applications in orthopedics [J].
Christenson, Elizabeth M. ;
Anseth, Kristi S. ;
van den Beucken, Leroen J. J. P. ;
Chan, Casey K. ;
Ercan, Batur ;
Jansen, John A. ;
Laurencin, Cato T. ;
Li, Wan-Ju ;
Murugan, Ramalingam ;
Nair, Lakshmi S. ;
Ramakrishna, Seeram ;
Tuan, Rocky S. ;
Webster, Thomas J. ;
Mikos, Antonios G. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2007, 25 (01) :11-22
[9]   VANCOMYCIN PENETRATION INTO BIOFILM COVERING INFECTED PROSTHESES AND EFFECT ON BACTERIA [J].
DAROUICHE, RO ;
DHIR, A ;
MILLER, AJ ;
LANDON, GC ;
RAAD, II ;
MUSHER, DM .
JOURNAL OF INFECTIOUS DISEASES, 1994, 170 (03) :720-723
[10]   Mechanical and microstructural properties of polycaprolactone scaffolds with one-dimensional, two-dimensional, and three-dimensional orthogonally oriented porous architectures produced by selective laser sintering [J].
Eshraghi, Shaun ;
Das, Suman .
ACTA BIOMATERIALIA, 2010, 6 (07) :2467-2476