Effect of Spectral Estimation on Ultrasonic Backscatter Parameters in Measurements of Cancellous Bones

被引:5
|
作者
Li, Boyi [1 ]
Xu, Feng [1 ]
Liu, Chengcheng [2 ]
Li, Dan [1 ]
Le, Lawrence H. [3 ]
Ta, Dean [4 ]
Wang, Weiqi [1 ]
机构
[1] Fudan Univ, Dept Elect Engn, Shanghai 200433, Peoples R China
[2] Tongji Univ, Inst Acoust, Shanghai 200092, Peoples R China
[3] Univ Alberta, Dept Radiol & Diagnost Imaging, Edmonton, AB T6G 2B7, Canada
[4] Fudan Univ, State Key Lab ASIC & Syst, Shanghai 200433, Peoples R China
来源
IEEE ACCESS | 2019年 / 7卷
基金
中国国家自然科学基金;
关键词
Ultrasonic backscatter; osteoporosis; cancellous bone; autoregressive spectral estimation; TRABECULAR-BONE; QUANTITATIVE ULTRASOUND; DIFFERENCE MEASUREMENTS; MECHANICAL-PROPERTIES; FREQUENCY-DEPENDENCE; OSTEOPOROSIS; COEFFICIENT; ATTENUATION; FRACTURES; CALCANEUS;
D O I
10.1109/ACCESS.2019.2924041
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Objective: the purpose of this paper was to investigate the effect of spectral computation methods on the estimation of the four ultrasonic backscatter parameters, namely, the apparent integrated backscatter (AIB), the zero frequency intercept of apparent backscatter (FIAB), the frequency slope of apparent backscatter (FSAB), and the backscatter spectral centroid shift (SCS) from the backscattered signal of interest (SOI), and their subsequent correlations with cancellous bone parameter [bone volume/total volume (BV/TV)]. Methods: ultrasonic backscatter measurements were performed on 26 bovine cancellous bone specimens using a 1.0-MHz focused transducer. Four spectral estimation algorithms, including the classical periodogram, the autoregressive (AR) Burg algorithm, the AR covariance algorithm, and the AR modified covariance algorithm, were used to calculate the ultrasonic parameters. Influence of the signal's delay time (T1) and its length (T2) on the strength of correlation between BV/TV and backscatter parameters was also studied. Results: the results have demonstrated that the AR-based estimators provide much more reliable and stronger correlations between the BV/TV and backscatter parameters than the classical periodogram. Recommendations for choosing SOI were also suggested. Conclusion: these results indicate that the AR-based method has a promising potential to enhance the performance of evaluation and diagnosis of the cancellous bone using the ultrasonic backscatter method. Significance: the enhancement of the correlation may provide a positive impact on the ultrasonic backscatter method to diagnose and monitor the bone quantity.
引用
收藏
页码:83034 / 83045
页数:12
相关论文
共 50 条
  • [1] An Amplitude Modulation Ultrasonic Backscatter Method for Estimation Characterization of Cancellous Bones
    Li, Boyi
    Liu, Chengcheng
    Li, Ying
    Tran, Tho N. H. T.
    Ta, Dean
    INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021), 2021,
  • [2] Effect of the cortex on ultrasonic backscatter measurements of cancellous bone
    Hoffmeister, Brent K.
    Holt, Andrew P.
    Kaste, Sue C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (19): : 6243 - 6255
  • [3] Effect of transducer position on ultrasonic backscatter measurements of cancellous bone
    Hoffmeister, Brent K.
    Lawler, Blake C.
    Viano, Ann M.
    Mobley, Joel
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2023, 154 (05): : 2858 - 2868
  • [4] Influence of the precision of spectral backscatter measurements on the estimation of scatterers size in cancellous bone
    Padilla, F.
    Jenson, F.
    Laugier, P.
    ULTRASONICS, 2006, 44 : E57 - E60
  • [5] Measurement of spectral maximum shift of ultrasonic backscatter signals in cancellous bone
    Ta, De-an
    Zhou, Guo-Hui
    Wang, Wei-qi
    Yu, Jian-guo
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 2703 - 2706
  • [6] Ultrasonic backscatter and attenuation of cancellous bone
    Deligianni, D
    Apostolopoulos, K
    ADVANCES IN SCATTERING AND BIOMEDICAL ENGINEERING, PROCEEDINGS, 2004, : 504 - 511
  • [7] Precision of the estimation of scatterers size in cancellous bone using spectral backscatter
    Padilla, F
    Jenson, F
    Laugier, P
    2005 IEEE Ultrasonics Symposium, Vols 1-4, 2005, : 2026 - 2029
  • [8] The spectrum entropy estimation method for ultrasonic backscattering signal in cancellous bones
    XU Feng
    XU Yiwen
    LIU Chengcheng
    TA De’an
    GUO Jianzhong
    ChineseJournalofAcoustics, 2017, 36 (03) : 337 - 346
  • [9] THE EFFECT OF FOCUSING ON ULTRASONIC BACKSCATTER MEASUREMENTS
    YUAN, YW
    SHUNG, KK
    ULTRASONIC IMAGING, 1986, 8 (02) : 121 - 130
  • [10] Backscatter-difference Measurements of Cancellous Bone Using an Ultrasonic Imaging System
    Hoffmeister, Brent K.
    Smathers, Morgan R.
    Miller, Catherine J.
    McPherson, Joseph A.
    Thurston, Cameron R.
    Spinolo, P. Luke
    Lee, Sang-Rok
    ULTRASONIC IMAGING, 2016, 38 (04) : 285 - 297