Quantitative Ultrasound Imaging for Monitoring In Situ High-Intensity Focused Ultrasound Exposure

被引:17
作者
Ghoshal, Goutam [1 ,2 ]
Kemmerer, Jeremy P. [2 ]
Karunakaran, Chandra [2 ]
Abuhabsah, Rami [2 ]
Miller, Rita J. [2 ]
Sarwate, Sandhya [2 ]
Oelze, Michael L. [2 ]
机构
[1] Acoust MedSyst Inc, Savoy, IL USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
关键词
quantitative ultrasound imaging; HIFU; effective scatterer diameter; effective scatterer concentration; parametric imaging; NONINVASIVE TEMPERATURE ESTIMATION; TISSUE TEMPERATURE; PROPAGATION SPEED; VIVO; ATTENUATION; BACKSCATTER; DEPENDENCE; FEASIBILITY; THERAPY; PHANTOM;
D O I
10.1177/0161734614524179
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Quantitative ultrasound (QUS) imaging is hypothesized to map temperature elevations induced in tissue with high spatial and temporal resolution. To test this hypothesis, QUS techniques were examined to monitor high-intensity focused ultrasound (HIFU) exposure of tissue. In situ experiments were conducted on mammary adenocarcinoma tumors grown in rats and lesions were formed using a HIFU system. A thermocouple was inserted into the tumor to provide estimates of temperature at one location. Backscattered time-domain waveforms from the tissue during exposure were recorded using a clinical ultrasonic imaging system. Backscatter coefficients were estimated using a reference phantom technique. Two parameters were estimated from the backscatter coefficient (effective scatterer diameter (ESD) and effective acoustic concentration (EAC). The changes in the average parameters in the regions corresponding to the HIFU focus over time were correlated to the temperature readings from the thermocouple. The changes in the EAC parameter were consistently correlated to temperature during both heating and cooling of the tumors. The changes in the ESD did not have a consistent trend with temperature. The mean ESD and EAC before exposure were 120 +/- 16 mm and 32 +/- 3 db/cm(3), respectively, and changed to 144 +/- 9 mm and 51 +/- 7 db/cm(3), respectively, just before the last HIFU pulse was delivered to the tissue. After the tissue cooled down to 37 degrees C, the mean ESD and EAC were 126 +/- 8 mm and 35 +/- 4 db/cm(3), respectively. Peak temperature in the range of 50-60 degrees C was recorded by a thermocouple placed just behind the tumor. These results suggest that QUS techniques have the potential to be used for non-invasive monitoring of HIFU exposure.
引用
收藏
页码:239 / 255
页数:17
相关论文
共 50 条
  • [31] High-intensity focused ultrasound in ovarian cancer xenografts
    Wu, Rong
    Hu, Bing
    Jiang, Li-Xin
    Hung, Ying
    Kuang, Sheng-Li
    Zhang, Bulin
    ADVANCES IN THERAPY, 2008, 25 (08) : 810 - 819
  • [32] Liver hemostasis using high-intensity focused ultrasound
    Vaezy, S
    Martin, R
    Schmiedl, U
    Caps, M
    Taylor, S
    Beach, K
    Carter, S
    Kaczkowski, P
    Keilman, G
    Helton, S
    Chandler, W
    Mourad, P
    Rice, M
    Roy, R
    Crum, L
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (09) : 1413 - 1420
  • [33] High-intensity focused ultrasound for treating prostate cancer
    Acher, Peter L.
    Hodgson, Dominic J.
    Murphy, Declan G.
    Cahill, Declan J.
    BJU INTERNATIONAL, 2007, 99 (01) : 28 - 32
  • [34] High-Intensity Focused Ultrasound in Dentistry: A Literature Review
    Tran, Minh Dien
    Ngo, Hien
    Fawzy, Amr
    INTERNATIONAL DENTAL JOURNAL, 2024, 74 (05) : 1168 - 1173
  • [35] Focal High-Intensity Focused Ultrasound Ablation of the Prostate
    Abrams, Lauren R.
    Koch, Michael O.
    Bahler, Clinton D.
    JOURNAL OF ENDOUROLOGY, 2021, 35 : S24 - S32
  • [36] High-Intensity Focused Ultrasound in the Treatment of Breast Cancer
    Feril, Loreto B., Jr.
    Fernan, Rodney L.
    Tachibana, Katsuro
    CURRENT MEDICINAL CHEMISTRY, 2021, 28 (25) : 5179 - 5188
  • [37] High-intensity focused ultrasound in the management of prostate cancer
    Chaussy, Christian
    Thueroff, Stefan
    EXPERT REVIEW OF MEDICAL DEVICES, 2010, 7 (02) : 209 - 217
  • [38] In Vitro and In Vivo High-Intensity Focused Ultrasound Thrombolysis
    Wright, Cameron
    Hynynen, Kullervo
    Goertz, David
    INVESTIGATIVE RADIOLOGY, 2012, 47 (04) : 217 - 225
  • [39] Current Status of Clinical High-Intensity Focused Ultrasound
    Hwang, Joo Ha
    Crum, Lawrence A.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 130 - 133
  • [40] High-intensity focused ultrasound in ovarian cancer xenografts
    Rong Wu
    Bing Hu
    Li-Xin Jiang
    Ying Hung
    Sheng-Li Kuang
    Bulin Zhang
    Advances in Therapy, 2008, 25 : 810 - 819