Quantitative Ultrasound Imaging for Monitoring In Situ High-Intensity Focused Ultrasound Exposure

被引:17
作者
Ghoshal, Goutam [1 ,2 ]
Kemmerer, Jeremy P. [2 ]
Karunakaran, Chandra [2 ]
Abuhabsah, Rami [2 ]
Miller, Rita J. [2 ]
Sarwate, Sandhya [2 ]
Oelze, Michael L. [2 ]
机构
[1] Acoust MedSyst Inc, Savoy, IL USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
关键词
quantitative ultrasound imaging; HIFU; effective scatterer diameter; effective scatterer concentration; parametric imaging; NONINVASIVE TEMPERATURE ESTIMATION; TISSUE TEMPERATURE; PROPAGATION SPEED; VIVO; ATTENUATION; BACKSCATTER; DEPENDENCE; FEASIBILITY; THERAPY; PHANTOM;
D O I
10.1177/0161734614524179
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Quantitative ultrasound (QUS) imaging is hypothesized to map temperature elevations induced in tissue with high spatial and temporal resolution. To test this hypothesis, QUS techniques were examined to monitor high-intensity focused ultrasound (HIFU) exposure of tissue. In situ experiments were conducted on mammary adenocarcinoma tumors grown in rats and lesions were formed using a HIFU system. A thermocouple was inserted into the tumor to provide estimates of temperature at one location. Backscattered time-domain waveforms from the tissue during exposure were recorded using a clinical ultrasonic imaging system. Backscatter coefficients were estimated using a reference phantom technique. Two parameters were estimated from the backscatter coefficient (effective scatterer diameter (ESD) and effective acoustic concentration (EAC). The changes in the average parameters in the regions corresponding to the HIFU focus over time were correlated to the temperature readings from the thermocouple. The changes in the EAC parameter were consistently correlated to temperature during both heating and cooling of the tumors. The changes in the ESD did not have a consistent trend with temperature. The mean ESD and EAC before exposure were 120 +/- 16 mm and 32 +/- 3 db/cm(3), respectively, and changed to 144 +/- 9 mm and 51 +/- 7 db/cm(3), respectively, just before the last HIFU pulse was delivered to the tissue. After the tissue cooled down to 37 degrees C, the mean ESD and EAC were 126 +/- 8 mm and 35 +/- 4 db/cm(3), respectively. Peak temperature in the range of 50-60 degrees C was recorded by a thermocouple placed just behind the tumor. These results suggest that QUS techniques have the potential to be used for non-invasive monitoring of HIFU exposure.
引用
收藏
页码:239 / 255
页数:17
相关论文
共 50 条
  • [21] Real-time visualization of high-intensity focused ultrasound treatment using ultrasound imaging
    Vaezy, S
    Shi, XG
    Martin, RW
    Chi, E
    Nelson, PI
    Bailey, MR
    Crum, LA
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (01) : 33 - 42
  • [22] Improving high-intensity focused ultrasound beam imaging via a backscattering suppression algorithm
    Matsui, Kazuhiro
    Azuma, Takashi
    Fujiwara, Keisuke
    Takeuchi, Hideki
    Itani, Kazunori
    Wang, Junchen
    Iwahashi, Toshihide
    Kobayashi, Etsuko
    Sakuma, Ichiro
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (05)
  • [23] High-Intensity Focused Ultrasound Treatment for Advanced Pancreatic Cancer
    Zhou, Yufeng
    GASTROENTEROLOGY RESEARCH AND PRACTICE, 2014, 2014
  • [24] High-intensity focused ultrasound ablation of uterine localized adenomyosis
    Dong, Xiaojing
    Yang, Zhu
    CURRENT OPINION IN OBSTETRICS & GYNECOLOGY, 2010, 22 (04) : 326 - 330
  • [25] Applications of High-Intensity Focused Ultrasound in the Treatment of Different Pathologies
    Manzoor, Iqra
    Bacha, Raham
    Gilani, Syed Amir
    JOURNAL OF DIAGNOSTIC MEDICAL SONOGRAPHY, 2021, 37 (02) : 171 - 178
  • [26] High-Intensity Focused Ultrasound in the Treatment of Abdominal Wall Endometriosis
    Shi, Suhua
    Ni, Guantai
    Ling, Li
    Ding, Huafeng
    Zhou, Yihui
    Ding, Zhimin
    JOURNAL OF MINIMALLY INVASIVE GYNECOLOGY, 2020, 27 (03) : 704 - 711
  • [27] Ex Vivo and In Vivo Monitoring and Characterization of Thermal Lesions by High-Intensity Focused Ultrasound and Microwave Ablation Using Ultrasonic Nakagami Imaging
    Zhang, Siyuan
    Shang, Shaoqiang
    Han, Yuqiang
    Gu, Chunming
    Wu, Shan
    Liu, Sihao
    Niu, Gang
    Bouakaz, Ayache
    Wan, Mingxi
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (07) : 1701 - 1710
  • [28] Gel phantom for use in high-intensity focused ultrasound dosimetry
    Lafon, C
    Zderic, V
    Noble, ML
    Yuen, JC
    Kaczkowski, PJ
    Sapozhnikov, OA
    Chavrier, F
    Crum, LA
    Vaezy, S
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (10) : 1383 - 1389
  • [29] Treatment of localized prostate cancer with high-intensity focused ultrasound
    Baumunk, D.
    Schostak, M.
    UROLOGE, 2015, 54 (02): : 183 - 190
  • [30] Development and Characterization of a Tissue-Mimicking Material for High-Intensity Focused Ultrasound
    King, Randy L.
    Liu, Yunbo
    Maruvada, Subha
    Herman, Bruce A.
    Wear, Keith A.
    Harris, Gerald R.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (07) : 1397 - 1405