Critical computational aspects of near infrared circular tomographic imaging: Analysis of measurement number, mesh resolution and reconstruction basis

被引:45
作者
Yalavarthy, Phaneendra K. [1 ]
Dehghani, Hamid
Pogue, Brian W.
Paulsen, Keith D.
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Univ Exeter, Sch Phys, Exeter EX4 4QL, Devon, England
来源
OPTICS EXPRESS | 2006年 / 14卷 / 13期
关键词
D O I
10.1364/OE.14.006113
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The image resolution and contrast in Near-Infrared (NIR) tomographic image reconstruction are affected by parameters such as the number of boundary measurements, the mesh resolution in the forward calculation and the reconstruction basis. Increasing the number of measurements tends to make the sensitivity of the domain more uniform reducing the hypersensitivity at the boundary. Using singular-value decomposition (SVD) and reconstructed images, it is shown that the numbers of 16 or 24 fibers are sufficient for imaging the 2D circular domain for the case of 1% noise in the data. The number of useful singular values increases as the logarithm of the number of measurements. For this 2D reconstruction problem, given a computational limit of 10 sec per iteration, leads to choice of forward mesh with 1785 nodes and reconstruction basis of 30 x 30 elements. In a three-dimensional (3D) NIR imaging problem, using a single plane of data can provide useful images if the anomaly to be reconstructed is within the measurement plane. However, if the location of the anomaly is not known, 3D data collection strategies are very important. Further the quantitative accuracy of the reconstructed anomaly increased approximately from 15% to 89% as the anomaly is moved from the centre to boundary, respectively. The data supports the exclusion of out of plane measurements may be valid for 3D NIR imaging.
引用
收藏
页码:6113 / 6127
页数:15
相关论文
共 26 条
  • [11] Singular-value analysis and optimization of experimental parameters in fluorescence molecular tomography
    Graves, EE
    Culver, JP
    Ripoll, J
    Weissleder, R
    Ntziachristos, V
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2004, 21 (02) : 231 - 241
  • [12] A submillimeter resolution fluorescence molecular imaging system for small animal imaging
    Graves, EE
    Ripoll, J
    Weissleder, R
    Ntziachristos, V
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 901 - 911
  • [13] Three-dimensional optical tomography of the premature infant brain
    Hebden, JC
    Gibson, A
    Yusof, RM
    Everdell, N
    Hillman, EMC
    Delpy, DT
    Arridge, SR
    Austin, T
    Meek, JH
    Wyatt, JS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (23) : 4155 - 4166
  • [14] Three-dimensional time-resolved optical tomography of a conical breast phantom
    Hebden, JC
    Veenstra, H
    Dehghani, H
    Hillman, EMC
    Schweiger, M
    Arridge, SR
    Delpy, DT
    [J]. APPLIED OPTICS, 2001, 40 (19) : 3278 - 3287
  • [15] Optical image reconstruction using frequency-domain data: Simulations and experiments
    Jiang, HB
    Paulsen, KD
    Osterberg, UL
    Pogue, BW
    Patterson, MS
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1996, 13 (02) : 253 - 266
  • [16] Improved continuous light diffusion imaging in single- and multi-target tissue-like phantoms
    Jiang, HB
    Paulsen, KD
    Osterberg, UL
    Patterson, MS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (03) : 675 - 693
  • [17] Optimal imaging with adaptive mesh refinement in electrical impedance tomography
    Molinari, M
    Blott, BH
    Cox, SJ
    Daniell, GJ
    [J]. PHYSIOLOGICAL MEASUREMENT, 2002, 23 (01) : 121 - 128
  • [18] Instrumentation and design of a frequency-domain diffuse optical tomography imager for breast cancer detection
    Pogue, Brian W.
    Testorf, Markus
    McBride, Troy
    Osterberg, Ulf
    Paulsen, Keith
    [J]. OPTICS EXPRESS, 1997, 1 (13): : 391 - 403
  • [19] Three-dimensional simulation of near-infrared diffusion in tissue:: boundary condition and geometry analysis for finite-element image reconstruction
    Pogue, BW
    Geimer, S
    McBride, TO
    Jiang, SD
    Österberg, UL
    Paulsen, KD
    [J]. APPLIED OPTICS, 2001, 40 (04) : 588 - 600
  • [20] Electrode configurations for improved spatial resolution in electrical impedance tomography
    Polydorides, N
    McCann, H
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (12) : 1862 - 1870