Use of an automated fluorescent microsphere method to measure regional blood flow in the fetal lamb

被引:22
作者
Tan, WP [1 ]
Riggs, KW [1 ]
Thies, RL [1 ]
Rurak, DW [1 ]
机构
[1] UNIV BRITISH COLUMBIA, FAC PHARMACEUT SCI, BRITISH COLUMBIA RES INST CHILD & FAMILY HLTH, VANCOUVER, BC V5Z 4H4, CANADA
关键词
fluorescent microspheres; organ and tissue blood flow; fluorescent microplate reader; spillover correction;
D O I
10.1139/cjpp-75-8-959
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
We have developed a method for measuring regional blood flow by means of fluorescent microspheres in all organs and tissues of the fetal lamb, including brain, heart, lung, liver, gut, spleen, kidney, adrenal, brown fat, skin, muscle, bone, and placenta. Five different fluorescent-labeled microspheres were used: blue (B), yellow-green (Y), orange (O), red (R), and crimson (C). An automated, 96-well microplate fluorescent reader (bottom reading) was chosen for the assay because of the rapidity and high throughput that it offers. Tissue samples were digested by 4 M ethanolic KOH. The sedimentation method and dye extraction with Cellosolve acetate, as previously reported by others, were used for the sample processing. The bones were crushed and allowed to directly soak in Cellosolve acetate to extract the dye. The relationship between microsphere number and fluorescent intensity was linear over a broad range of microsphere numbers (80 - 20 000/mL). The coefficients of variation of within-run and between-run precision were 3.39 +/- 1.10% and 4.54 +/- 1.10%, respectively. Recovery of microspheres from tissues and blood averaged 94.3 +/- 2.5% and was not dependent on microsphere number. The spillover of the fluorescent signals into adjacent colors was 4.0 +/- 0.1% for O to Y, 8.1 +/- 0.4% for O to R, and 9.1 +/- 0.5% for R to C, and these values were constant over a wide range in concentrations of the microsphere pairs. No evidence was obtained for quenching of the emission of one fluorophore via photon absorption by another fluorophore. The measurements of regional blood flow obtained with fluorescent microspheres in three chronically instrumented fetal lambs at similar to 140 days gestation were similar to the flow estimates obtained using radioactive microspheres in four other fetal lambs at the same gestational age. The fluorescent method is thus a viable alternative to the radioactive technique for the measurement of regional blood flow to all fetal organs and tissues, particularly when an automated fluorescent microplate reader is employed to reduce analysis time.
引用
收藏
页码:959 / 968
页数:10
相关论文
共 31 条
[1]  
ABEL FL, 1993, CIRC SHOCK, V41, P156
[2]  
AN D, 1996, MANUAL USING FLUORES
[3]  
Austin G E, 1993, Am J Cardiovasc Pathol, V4, P352
[4]  
BUCHWALDER LF, 1996, J SOC GYNECOL INVE S, V3, pA258
[5]   SOME SOURCES OF ERROR IN MEASURING REGIONAL BLOOD FLOW WITH RADIOACTIVE MICROSPHERES [J].
BUCKBERG, GD ;
LUCK, JC ;
PAYNE, DB ;
HOFFMAN, JIE ;
ARCHIE, JP ;
FIXLER, DE .
JOURNAL OF APPLIED PHYSIOLOGY, 1971, 31 (04) :598-&
[6]   FLUORESCENT VS RADIOACTIVE MICROSPHERE MEASUREMENT OF REGIONAL MYOCARDIAL BLOOD-FLOW [J].
CHIEN, GL ;
ANSELONE, CG ;
DAVIS, RF ;
VANWINKLE, DM .
CARDIOVASCULAR RESEARCH, 1995, 30 (03) :405-412
[7]   CARDIOVASCULAR-RESPONSES TO HYPOXEMIA AND ACIDEMIA IN FETAL LAMBS [J].
COHN, HE ;
SACKS, EJ ;
HEYMANN, MA ;
RUDOLPH, AM .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1974, 120 (06) :817-824
[8]  
COURT DJ, 1984, J DEV PHYSIOL, V6, P349
[9]   RELATIVE ERROR AND VARIABILITY IN BLOOD-FLOW MEASUREMENTS WITH RADIOLABELED MICROSPHERES [J].
DOLE, WP ;
JACKSON, DL ;
ROSENBLATT, JI ;
THOMPSON, WL .
AMERICAN JOURNAL OF PHYSIOLOGY, 1982, 243 (03) :H371-H378
[10]   VALIDATION OF FLUORESCENT-LABELED MICROSPHERES FOR MEASUREMENT OF REGIONAL ORGAN PERFUSION [J].
GLENNY, RW ;
BERNARD, S ;
BRINKLEY, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1993, 74 (05) :2585-2597