Experimental study of bone-implant contact with a parabolic acetabular component (Hofer-Imhof)

被引:4
作者
Effenberger, H
Böhm, G
Huber, M
Lintner, F
Hofer, H
机构
[1] Gen Hosp Gmunden, Dept Orthopaed, A-4810 Gmunden, Austria
[2] Salzburg Gen Hosp, Dept Orthopaed, Salzburg, Austria
[3] Inst Pathol Bacteriol, Vienna, Austria
关键词
Acetabular Component; Human Cadaver; Digitize Measurement; Bone Contact; Sclerotic Bone;
D O I
10.1007/s004020050035
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Incomplete bone contact with threaded acetabular components is usually attributable to the tapered thread design and sclerotic bone, and might be the reason for early loosening. The self-tapping flat threads of the Hofer-Imhof cup allow accurate insertion without incarceration, even in sclerotic bone. To demonstrate the effectiveness of this threaded cup design based on digitized measurements of bone contact area, six Hofer-Imhof cups were implanted into three human cadavers. The bone contact area at the threads was digitally analyzed on cadaver-explanted cups cut into horizontal and coronal sections. In cups fixed with ideal form fit, bone contact was 31.6% (range 8-55%) at the threads and 28.5% (range 0-49%) be tween the threads on average. In cups with incomplete bone contact secondary to insufficient reaming, the bone contact area averaged 27.4% (range 10-44%) at the threads. In sclerotic bone, the highest bone contact at the threads was 69%. The threads were the first part of the cup to achieve primary bone contact. Factors contributing to poor contact were incomplete reaming, sections which passed through the incisura of the acetabulum, and incomplete contact with the medial aspect of the acetabulum. The aim was to achieve complete bone contact. Because of the higher contact areas observed in sclerotic bone, some sclerotic bone should remain after reaming.
引用
收藏
页码:160 / 165
页数:6
相关论文
共 15 条
[1]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[2]  
BOHM G, 1995, INT J SURG PATHOL S, V2, P9
[3]  
Delaunay CP, 1997, CLIN ORTHOP RELAT R, P130
[4]  
EFFENBERGER H, 1997, ORTHOP T, V21, P223
[5]  
EFFENBERGER HH, 1991, NONCEMENTED TOTAL HI, P275
[6]   CEMENTLESS ACETABULAR COMPONENTS [J].
ENGH, CA ;
GRIFFIN, WL ;
MARX, CL .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :53-59
[7]   Loosening of threaded acetabular cups in arthroplasty of the hip - The association with different types of coxarthrosis [J].
HernandezVaquero, D ;
SuarezVazquez, A ;
FernandezCorona, C ;
MenendezVinuela, G ;
AlegreMateo, R ;
GarciaSandoval, MA .
INTERNATIONAL ORTHOPAEDICS, 1996, 20 (02) :70-74
[8]   MECHANICAL FUNCTION OF SUBCHONDRAL BONE AS EXPERIMENTALLY DETERMINED ON ACETABULUM OF HUMAN PELVIS [J].
JACOB, HAC ;
HUGGLER, AH ;
DIETSCHI, C ;
SCHREIBER, A .
JOURNAL OF BIOMECHANICS, 1976, 9 (10) :625-&
[9]   HISTOLOGY OF TISSUE ADJACENT TO AN HAC-COATED FEMORAL PROSTHESIS - A CASE-REPORT [J].
LINTNER, F ;
BOHM, G ;
HUBER, M ;
SCHOLZ, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :824-830
[10]  
Lintner F, 1986, J Arthroplasty, V1, P183, DOI 10.1016/S0883-5403(86)80029-8