Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space

被引:2
作者
Zhang, J.
Bao, Z. R.
Huang, X. T.
Jiang, D. Z.
Xie, C. H.
Zhou, Y. F.
Liu, H. [1 ]
机构
[1] Wuhan Univ, Zhongnan Hosp, Dept Radiat & Med Oncol, Wuhan, Hubei, Peoples R China
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2019年 / 17卷 / 02期
关键词
Image quality; CBCT; three-dimensional; IMAGE-GUIDED RADIOTHERAPY; QUALITY-ASSURANCE PROGRAM; FLAT-PANEL IMAGER; RADIATION-THERAPY; CT; SYSTEM; KV; LOCALIZATION; PROJECTIONS; ULTRASOUND;
D O I
10.18869/acadpub.ijrr.17.2.189
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Material and Methods: Compared to the traditional methods with only two-dimensional axial slice evaluated, the new methods were designed to evaluate the CBCT image quality not only in the axial slices but also along three dimensions and three sections. Image parameters were evaluated by scanning the corresponding phantom. Results: The new methods demonstrated the high spatial resolution, the image uniformity, the high CT numbers linearity, geometric accuracy and relatively poor performance in low contrast resolution, compared to the results of plan CT. Additionally, the spatial resolution, the CT number linearity and the image uniformity proved to be more superior along the z direction and in the cross-section. Furthermore, the spatial resolution and the CT value linearity decreased gradually when deviating from the center of rotation axis and accelerates the reduction towards the periphery area, suggesting that locating the target within the range of 4 cm around the isocenter during radiotherapy ensure the optimal spatial resolution and the CT value linearity. Conclusions: Results indicate that the proposed three-dimensional image quality evaluation holds the potential to complement existing standards to provide sufficient information for the CBCT image performance and thereby guiding the clinical applications of CBCT.
引用
收藏
页码:189 / 202
页数:14
相关论文
共 50 条
[31]   Evaluation and choice of imaging protocols on the Elekta XVI® kilovoltage cone-beam computed tomography imaging system [J].
Werle, F. ;
Buffard, E. ;
Brune, K. ;
Perrin, B. ;
Atlani, D. .
CANCER RADIOTHERAPIE, 2014, 18 (01) :47-54
[32]   Radiation dose measurements during kilovoltage-cone beam computed tomography imaging in radiotherapy [J].
Kumar, A. Sathish ;
Singh, I. Rabi Raja ;
Sharma, Sunil Dutt ;
John, Subhashini ;
Ravindran, B. Paul .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2016, 12 (02) :858-863
[33]   Impact of voxel size and scan time on the accuracy of three-dimensional radiological imaging data from cone-beam computed tomography [J].
Dach, Eva ;
Bergauer, Bastian ;
Seidel, Anna ;
Von Wilmowsky, Cornelius ;
Adler, Werner ;
Lell, Michael ;
Wichmann, Manfred ;
Matta, Ragai-Edward .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2018, 46 (12) :2190-2196
[34]   REFINEMENT OF TREATMENT SETUP AND TARGET LOCALIZATION ACCURACY USING THREE-DIMENSIONAL CONE-BEAM COMPUTED TOMOGRAPHY FOR STEREOTACTIC BODY RADIOTHERAPY [J].
Wang, Zhiheng ;
Nelson, John W. ;
Yoo, Sua ;
Wu, Jackie ;
Kirkpatrick, John P. ;
Marks, Lawrence B. ;
Yin, Fang-Fang .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02) :571-577
[35]   Cone-beam computed tomography performance in measuring periodontal bone loss [J].
Yang, Jingmei ;
Li, Xinyi ;
Duan, Dingyu ;
Bai, Lin ;
Zhao, Lei ;
Xu, Yi .
JOURNAL OF ORAL SCIENCE, 2019, 61 (01) :61-66
[36]   Effects of rapid maxillary expansion on upper airway volume: A three-dimensional cone-beam computed tomography study [J].
Abdalla, Yousef ;
Brown, Louise ;
Sonnesen, Liselotte .
ANGLE ORTHODONTIST, 2019, 89 (06) :917-923
[37]   Three-dimensional evaluation of angular, linear, and resorption features of maxillary impacted canines on cone-beam computed tomography [J].
İlhan Metin Dağsuyu ;
Fatih Kahraman ;
Rıdvan Okşayan .
Oral Radiology, 2018, 34 :66-72
[38]   Quantitative breast density measurement based on three-dimensional images: a study on cone-beam breast computed tomography [J].
Liu, Aidi ;
Yin, Lu ;
Ma, Yue ;
Han, Peng ;
Wu, Yalin ;
Wu, Yaopan ;
Ye, Zhaoxiang .
ACTA RADIOLOGICA, 2022, 63 (08) :1023-1031
[39]   Three-dimensional quantification of mandibular asymmetry through cone-beam computerized tomography [J].
Cevidanes, Lucia H. S. ;
Alhadidi, Abeer ;
Paniagua, Beatriz ;
Styner, Martin ;
Ludlow, John ;
Mol, Andre ;
Turvey, Timothy ;
Proffit, William R. ;
Rossouw, Paul Emile .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2011, 111 (06) :757-770
[40]   Low-dose Cone-Beam Computed Tomography Reconstruction through a fast Three-Dimensional Compressed Sensing Method Based on the Three-Dimensional Pseudo-polar Fourier Transform [J].
Teyfouri, N. ;
Rabbani, Hossein ;
Jabbari, I .
JOURNAL OF MEDICAL SIGNALS & SENSORS, 2022, 12 (01) :8-24