A Systematic Review and Meta-Analysis of Perioperative Parameters in Robot-Guided, Navigated, and Freehand Thoracolumbar Pedicle Screw Instrumentation

被引:27
作者
Siccoli, Alessandro [1 ,2 ]
Klukowska, Anita M. [1 ,3 ]
Schroder, Marc L. [1 ]
Staartjes, Victor E. [1 ,2 ,4 ]
机构
[1] Bergman Clin, Dept Neurosurg, Amsterdam, Netherlands
[2] Univ Zurich, Fac Med, Zurich, Switzerland
[3] Univ Nottingham, Sch Med, Nottingham, England
[4] Vrije Univ Amsterdam, Amsterdam UMC, Amsterdam Movement Sci, Neurosurg, Amsterdam, Netherlands
关键词
Instrumentation; Navigation; Outcome; Perioperative; Robot; Spinal fusion; INTRAOPERATIVE COMPUTED-TOMOGRAPHY; FLUOROSCOPY-BASED NAVIGATION; CONVENTIONAL FLUOROSCOPY; LEARNING-CURVE; PLACEMENT ACCURACY; LUMBAR FUSION; SPINE; INSERTION; SURGERY; ASSISTANCE;
D O I
10.1016/j.wneu.2019.03.196
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Robotic guidance (RG) and navigation (NV) have been shown to reduce radiologic and clinically relevant pedicle screw malpositions. It remains unknown if there are any additional benefits to these techniques in intraoperative and perioperative end points. METHODS: We conducted a systematic review in MEDLINE, Embase, Scopus, and the Cochrane Library and identified controlled studies comparing RG, NV, and freehand (FH) thoracolumbar pedicle screw insertion and carried out random-effects meta-analyses. RESULTS: Thirty-two studies (24,008 patients) were included. Only 8 studies (26%) were randomized, and study quality was rated as very low or low in 24 cases (77%). Compared with NV, FH procedures showed longer length of hospital stay (Delta, 0.7 days; 95% confidence interval, 0.2-1.2; P = 0.006) and more overall complications (odds ratio, 1.6; 95% confidence interval, 1.3-1.9; P < 0.001). No statistically significant differences among RG and FH were identified, likely because of lack in statistical power (all P> 0.05). In particular, both RG and NV did not show increased intraoperative radiation use, as determined by seconds of fluoroscopy, compared with FH (both P > 0.05). CONCLUSIONS: It seems that navigation may offer potential benefits in perioperative outcomes such as length of hospital stay and overall complications, without significant increase in intraoperative radiation, which cannot yet be said for robotic guidance. The findings must be interpreted with caution, because the evidence is severely limited in both quantity and quality. Further evaluation will establish any demonstrable intraoperative or perioperative benefits to computer assistance, which may warrant the high costs often associated with these devices.
引用
收藏
页码:576 / +
页数:17
相关论文
共 64 条
[1]   Robot guidance for percutaneous minimally invasive placement of pedicle screws for pyogenic spondylodiscitis is associated with lower rates of wound breakdown compared to conventional fluoroscopy-guided instrumentation [J].
Alaid, Awad ;
von Eckardstein, Kajetan ;
Smoll, Nicolas Roydon ;
Solomiichuk, Volodymyr ;
Rohde, Veit ;
Martinez, Ramon ;
Schatlo, Bawarjan .
NEUROSURGICAL REVIEW, 2018, 41 (02) :489-496
[2]   Comparison of 30-Day Complications Between Navigated and Conventional Single-level Instrumented Posterior Lumbar Fusion [J].
Bovonratwet, Patawut ;
Nelson, Stephen J. ;
Ondeck, Nathaniel T. ;
Geddes, Benjamin J. ;
Grauer, Jonathan N. .
SPINE, 2018, 43 (06) :447-453
[3]   Minimally invasive transforaminal lumbosacral interbody fusion [J].
Chang, Peng-Yuan ;
Wang, Michael Y. .
Neurosurgical Focus, 2016, 41
[4]   Evaluation of Screw Placement Accuracy in Circumferential Lumbar Arthrodesis Using Robotic Assistance and Intraoperative Flat-Panel Computed Tomography [J].
Chenin, Louis ;
Capel, Cyrille ;
Fichten, Anthony ;
Peltier, Johann ;
Lefranc, Michel .
WORLD NEUROSURGERY, 2017, 105 :86-94
[5]   Minimally invasive transforaminal lumbar interbody fusion with the ROSATM Spine robot and intraoperative flat-panel CT guidance [J].
Chenin, Louis ;
Peltier, Johann ;
Lefranc, Michel .
ACTA NEUROCHIRURGICA, 2016, 158 (06) :1125-1128
[6]   Application of Intraoperative Computed Tomography With or Without Navigation System in Surgical Correction of Spinal Deformity A Preliminary Result of 59 Consecutive Human Cases [J].
Cui, Geng ;
Wang, Yan ;
Kao, Ting-Hsien ;
Zhang, Yonggang ;
Liu, Zhengsheng ;
Liu, Baowei ;
Li, Jie ;
Zhang, Xifeng ;
Zhu, Shourong ;
Lu, Ning ;
Mao, Keya ;
Wang, Zheng ;
Zhang, Xuesong ;
Yuan, Xincheng ;
Dong, Tianxiang ;
Xiao, Songhua .
SPINE, 2012, 37 (10) :891-900
[7]   Accuracy of Pedicle Screw Insertion Among 3 Image-Guided Navigation Systems: Systematic Review and Meta-Analysis [J].
Du, Jin Peng ;
Fan, Yong ;
Wu, Qi Ning ;
Wang, Dai Hua ;
Zhang, Jing ;
Hao, Ding Jun .
WORLD NEUROSURGERY, 2018, 109 :24-30
[8]   Revision Rate of Misplaced Pedicle Screws of the Thoracolumbar Spine-Comparison of Three-Dimensional Fluoroscopy Navigation with Freehand Placement: A Systematic Analysis and Review of the Literature [J].
Fichtner, Jens ;
Hofmann, Nicole ;
Rienmueller, Anna ;
Buchmann, Niels ;
Gempt, Jens ;
Kirschke, Jan S. ;
Ringel, Florian ;
Meyer, Bernhard ;
Ryang, Yu-Mi .
WORLD NEUROSURGERY, 2018, 109 :E24-E32
[9]   Iso-C/3-Dimensional Neuronavigation versus Conventional Fluoroscopy for Minimally Invasive Pedicle Screw Placement in Lumbar Fusion [J].
Fraser, J. ;
Gebhard, H. ;
Irie, D. ;
Parikh, K. ;
Haertl, R. .
MINIMALLY INVASIVE NEUROSURGERY, 2010, 53 (04) :184-190
[10]   Pedicle screw insertion: computed tomography versus fluoroscopic image guidance [J].
Fu, Tsai-Sheng ;
Wong, Chak-Bor ;
Tsai, Tsung-Ting ;
Liang, Yen-Chiu ;
Chen, Lih-Huei ;
Chen, Wen-Jer .
INTERNATIONAL ORTHOPAEDICS, 2008, 32 (04) :517-521