A comparison of conventional tube and EndoFlex tube for tracheal intubation in patients with a cervical spine immobilisation

被引:4
|
作者
Gaszynska, Ewelina [1 ]
Stankiewicz-Rudnicki, Michal [3 ]
Wieczorek, Andrzej [3 ]
Gaszynski, Tomasz [2 ]
机构
[1] Med Univ Lodz, Dept Hyg & Hlth Promot, Lodz, Poland
[2] Med Univ Lodz, Dept Emergency Med & Disaster Med, Lodz, Poland
[3] Med Univ Lodz, Dept Anesthesiol & Intens Therapy, Lodz, Poland
来源
SCANDINAVIAN JOURNAL OF TRAUMA RESUSCITATION & EMERGENCY MEDICINE | 2013年 / 21卷
关键词
Intubation; Cervical collar; EndoFlex; ENDOTRACHEAL-TUBE;
D O I
10.1186/1757-7241-21-79
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The EndoFlex is a new type of tracheal tube with an adjustable distal tip that can be bent without the use of a stylet. The aim of this study was to compare a standard endotracheal tube with the EndoFlex tracheal tube for intubation in patients with simulated cervical spine injury. Methods: A group of 60 patients without any kind of the cervical spine injury, classified as the ASA physiological scale I or II and qualified for elective surgery procedures were intubated with the use of classical Macintosh laryngoscope, and either a standard endotracheal tube with the intubation stylet in it or EndoFlex tube without stylet. The subjects were randomized into two subgroups. All patients have had the cervical collar placed on their neck for the simulation of intubation procedure in case of the spinal injury. Results: The intubation procedure was performed by 16 anesthetists with different experience (5-19 yrs). Time of intubation with the use of EndoFlex tube was similar to that with a the use of standard endotracheal tube and intubation stylet: Me (median) 19.5 s [IQR (interquatile range) 18-50] vs. Me 20 s [IQR 17-60] respectively (p = 0.9705). No significant additional maneuvers were necessary during intubation with the use of EndoFlex tube in comparison with standard endotracheal tube (70% vs. 56.6%) (p = 0.4220). Subjective assessment of the usability of both tubes revealed that more anesthesiologists found intubations with the use of EndoFlex more demanding than intubation with conventional tracheal tube and intubation stylet. The assessment of usability: very easy 3.3% vs. 20%, easy 83.4% vs. 56.7%, difficult 10% vs. 20% and very difficult 3.3% vs. 3.3% for standard endotracheal tube with stylet and EndoFlex, respectively. Conclusion: In conclusion we asses, that the EndoFlex tube does not improve intubation success rate, in fact it requires more maneuvers facilitating intubation and was found to be more difficult to use.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] A comparison of conventional tube and EndoFlex tube for tracheal intubation in patients with a cervical spine immobilisation
    Ewelina Gaszynska
    Michal Stankiewicz-Rudnicki
    Andrzej Wieczorek
    Tomasz Gaszynski
    Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 21
  • [2] A comparison of the intubating laryngeal mask tracheal tube with a standard tracheal tube for fibreoptic intubation
    Lucas, DN
    Yentis, SM
    ANAESTHESIA, 2000, 55 (04) : 358 - 361
  • [3] Videolaryngoscopy vs. flexible fibrescopy for tracheal intubation in patients with cervical spine immobilisation: a randomised controlled trial
    Choi, S.
    Yoo, H. K.
    Shin, K. W.
    Kim, Y. J.
    Yoon, H. K.
    Park, H. P.
    Oh, H.
    ANAESTHESIA, 2023, 78 (08) : 970 - 978
  • [4] Tracheal intubation in patients with cervical spine immobilization: a comparison of the Airwayscope®, LMA CTrach®, and the Macintosh laryngoscopes
    Malik, M. A.
    Subramaniam, R.
    Churasia, S.
    Maharaj, C. H.
    Harte, B. H.
    Laffey, J. G.
    BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (05) : 654 - 661
  • [5] Tracheal intubation in a simulated cervical spine immobilisation: The Macintosh laryngoscope versus supraglottic airway devices - A manikin study
    Aleksandrowicz, Dawid
    Gaszynski, Tomasz
    TRENDS IN ANAESTHESIA AND CRITICAL CARE, 2018, 21 : 53 - 56
  • [6] A comparison of a flexometallic tracheal tube with the intubating laryngeal mask tracheal tube for nasotracheal fibreoptic intubation using the two-scope technique
    Rai, M. R.
    Scott, S. H.
    Marfin, A. G.
    Popat, M. T.
    Pandit, J. J.
    ANAESTHESIA, 2009, 64 (12) : 1303 - 1306
  • [7] FIBEROPTIC INTUBATION - INFLUENCE OF TRACHEAL TUBE TIP DESIGN
    JONES, HE
    PEARCE, AC
    MOORE, P
    ANAESTHESIA, 1993, 48 (08) : 672 - 674
  • [8] Validation of tracheal intubation of wire-reinforced endotracheal tube with ultrasonography
    Inangil, Gokhan
    Deniz, Suleyman
    Temircan, Sedat
    Bakal, Omer
    Sen, Huseyin
    Ozkan, Sezai
    SIGNA VITAE, 2018, 14 (02) : 20 - 23
  • [9] Comparison of the C-MAC®, Airtraq®, and Macintosh laryngoscopes in patients undergoing tracheal intubation with cervical spine immobilization
    McElwain, J.
    Laffey, J. G.
    BRITISH JOURNAL OF ANAESTHESIA, 2011, 107 (02) : 258 - 264
  • [10] Endotrol tracheal tube and McGrath Mac are an effective combination for oral tracheal intubation
    Shimazaki, Azusa
    Ueshima, Hironobu
    Otake, Hiroshi
    SAUDI JOURNAL OF ANAESTHESIA, 2018, 12 (01) : 72 - 76