Determination of 11-deoxycortisol (Reichstein's compound S) in human plasma by clinical isotope dilution mass spectrometry using benchtop gas chromatography-mass selective detection

被引:17
|
作者
Wudy, SA
Hartmann, M
Homoki, J
机构
[1] Univ Giessen, Steroid Res Unit, Childrens Hosp, Steroid Lab, D-35392 Giessen, Germany
[2] Univ Ulm, Childrens Hosp, Steroid Lab, Ulm, Germany
关键词
gas chromatography; mass spectrometry; steroid; 11-deoxycortisol; compound S; stable isotope;
D O I
10.1016/S0039-128X(02)00052-1
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A first assay based on stable isotope dilution/gas chromatography-mass spectrometry (ID/GC-MS) has been developed for plasma 11-deoxycortisol (Reichstein's compound S), the leading hormonal marker of 11beta-hydroxylase deficiency. A suitable internal standard being unavailable, we synthesized dideuterated 11-deoxycortisol according to a newly devised synthetic procedure. 17,21-Dihydroxy-pregna-1,4-diene-3,20-dione underwent selective deuteration using Wilkinson's catalyst. Our product [1alpha,2alpha-H-2(2)]11-deoxycortisol was obtained in good yield (35.6%) and high isotopic purity (0.1% H-2(0), 99.9% H-2(2)). Structural confirmation was done by MS and NMR. Our plasma work up consisted of equilibration of plasma with internal standard ([1alpha,2alpha-H-2(2)]11-deoxyconisol), solid phase extraction with Extrelut(R) NT columns, a clean up step using Sephadex LH-20 mini columns and preparation of heptafluorobutyrates as derivatives. Quantification was achieved by selected ion monitoring of m/z 465.40 (analyte) and m/z 467.40 (internal standard). One hundred twenty picograms of 11-deoxycortisol gave a signal to noise ratio of 10. Calibration plot was linear. Spiking experiments showed good accuracy with relative errors <3.0%. Intraassay precision CV was 4.78% and interassay precision CV was 4.56%. We succeeded in integrating our new analyte into our already existing multisteroid ID/GC-MS plasma assay, which now, in its expanded version, is capable of determining all major diagnostic steroids of androgen related disorders in a single profile: 11-deoxycortisol, 17alpha-hydroxyprogesterone, testosterone, 4-androstenedione, 17alpha-hydroxypregnenolone, dehydroepiandrosterone, androstanediol and 5alpha-dihydrotestosterone. The diagnostic potential of our multisteroid ID/GC-MS assay, the small amounts of plasma (0.5 ml) required, the rapid and convenient sample work up, the application of benchtop GC-MS instrumentation, and highest specificity offered by mass spectrometric detection prove our assay suitable for routine clinical use, especially in pediatric endocrinology. (C) 2002 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:851 / 857
页数:7
相关论文
共 38 条
  • [11] Pipette tip solid-phase extraction and gas chromatography-mass spectrometry for the determination of mequitazine in human plasma
    Kumazawa, Takeshi
    Hasegawa, Chika
    Lee, Xiao-Pen
    Marumo, Akemi
    Shimmen, Natsuko
    Ishii, Akira
    Seno, Hiroshi
    Sato, Keizo
    TALANTA, 2006, 70 (02) : 474 - 478
  • [12] Interference from a glucuronide metabolite in the determination of ramipril and ramiprilat in human plasma and urine by gas chromatography-mass spectrometry
    Persson, BA
    Fakt, C
    Ervik, M
    Ahnoff, M
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 40 (03) : 794 - 798
  • [13] Determination of Formetanate Hydrochloride in Fruit Samples Using Liquid Chromatography-Mass Selective Detection or-Tandem Mass Spectrometry
    Podhorniak, Lynda V.
    Kamel, Alaa
    Rains, Diane M.
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2010, 58 (10) : 5862 - 5867
  • [14] METHOD VALIDATION FOR ORGANOCHLORINE INSECTICIDES DETERMINATION IN GINSENG BY USING ISOTOPE-DILUTION GAS CHROMATOGRAPHY-MASS SPECTROMETRY (ID-GC-MS)
    Aristiawan, Yosi
    Putri Ramadhaningtyas, Dillani
    Komalasari, Isna
    Styarini, Dyah
    Hamim, Nuryatini
    PERIODICO TCHE QUIMICA, 2020, 17 (34): : 362 - 372
  • [15] Comparison of liquid chromatography with fluorescence detection to liquid chromatography-mass spectrometry for the determination of fluoxetine and norfluoxetine in human plasma
    Kovacevic, I
    Miljkovic, B
    Jovanovic, D
    Prostran, M
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 830 (02): : 372 - 376
  • [16] Quantitative determination of phenothiazine derivatives in human plasma using monolithic silica solid-phase extraction tips and gas chromatography-mass spectrometry
    Kumazawa, Takeshi
    Hasegawa, Chika
    Uchigasaki, Seisaku
    Lee, Xiao-Pen
    Suzuki, Osamu
    Sato, Keizo
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (18) : 2521 - 2527
  • [17] Comparison of orthogonal liquid and gas chromatography-mass spectrometry platforms for the determination of amino acid concentrations in human plasma
    McGaw, Elizabeth A.
    Phinney, Karen W.
    Lowenthal, Mark S.
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (37) : 5822 - 5831
  • [18] Rapid quantitation of digoxin in human plasma and urine using isotope dilution liquid chromatography-tandem mass spectrometry
    Salvador, A
    Sagan, C
    Denouel, J
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (13) : 1917 - 1932
  • [19] Detection of allantoin in clinical samples using hydrophilic liquid chromatography with stable isotope dilution negative ion tandem mass spectrometry
    Turner, Rufus
    Stamp, Lisa K.
    Kettle, Anthony J.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 891 : 85 - 89
  • [20] Simultaneous determination of fentanyl and midazolam in plasma using direct solid-phase microextraction before gas chromatography-mass spectrometry analysis
    Dufresne, C
    Favetta, P
    Gonin, R
    Bureau, J
    Guitton, J
    ANALYTICAL LETTERS, 2002, 35 (09) : 1575 - 1590