Validation of the angular measurements of a new inertial-measurement-unit based rehabilitation system: comparison with state-of-the-art gait analysis

被引:67
|
作者
Leardini, Alberto [1 ]
Lullini, Giada [1 ]
Giannini, Sandro [1 ]
Berti, Lisa [1 ]
Ortolani, Maurizio [1 ]
Caravaggi, Paolo [1 ]
机构
[1] Ist Ortoped Rizzoli, Movement Anal Lab, Bologna, Italy
来源
JOURNAL OF NEUROENGINEERING AND REHABILITATION | 2014年 / 11卷
关键词
Inertial measurement unit; Gait analysis; Rehabilitation; Knee; Hip; Thorax; Joint flexion; Audio-visual bio-feedback; Video-games; JOINT ANGLE MEASUREMENT; KINEMATICS; MOTION; SENSORS; BIOFEEDBACK; OUTCOMES; GAMES;
D O I
10.1186/1743-0003-11-136
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Several rehabilitation systems based on inertial measurement units (IMU) are entering the market for the control of exercises and to measure performance progression, particularly for recovery after lower limb orthopaedic treatments. IMU are easy to wear also by the patient alone, but the extent to which IMU's malpositioning in routine use can affect the accuracy of the measurements is not known. A new such system (Riablo (TM), CoRehab, Trento, Italy), using audio-visual biofeedback based on videogames, was assessed against state-of-the-art gait analysis as the gold standard. Methods: The sensitivity of the system to errors in the IMU's position and orientation was measured in 5 healthy subjects performing two hip joint motion exercises. Root mean square deviation was used to assess differences in the system's kinematic output between the erroneous and correct IMU position and orientation. In order to estimate the system's accuracy, thorax and knee joint motion of 17 healthy subjects were tracked during the execution of standard rehabilitation tasks and compared with the corresponding measurements obtained with an established gait protocol using stereophotogrammetry. Results: A maximum mean error of 3.1 +/- 1.8 deg and 1.9 +/- 0.8 deg from the angle trajectory with correct IMU position was recorded respectively in the medio-lateral malposition and frontal-plane misalignment tests. Across the standard rehabilitation tasks, the mean distance between the IMU and gait analysis systems was on average smaller than 5 degrees. Conclusions: These findings showed that the tested IMU based system has the necessary accuracy to be safely utilized in rehabilitation programs after orthopaedic treatments of the lower limb.
引用
收藏
页数:7
相关论文
共 28 条
  • [21] Evaluating Gait Abnormalities in Asian Elephants Using Inertial Measurement Unit-Based Vertical Movement Symmetry Analysis: A Pilot Study
    Khammesri, Siripat
    Wantanajittikul, Kittichai
    Namwongprom, Kittikul
    Kittisirikul, Narueporn
    Ueangpaibool, Pichamon
    Thitaram, Chatchote
    Brown, Janine L.
    Kongsawasdi, Siriphan
    VETERINARY SCIENCES, 2025, 12 (02)
  • [22] Validation of foot pitch angle estimation using inertial measurement unit against marker-based optical 3D motion capture system
    Sharif Bidabadi S.
    Murray I.
    Lee G.Y.F.
    Biomedical Engineering Letters, 2018, 8 (3) : 283 - 290
  • [23] Design, Validity, and Reliability of a New Test, Based on an Inertial Measurement Unit System, for Measuring Cervical Posture and Motor Control in Children with Cerebral Palsy
    Carmona-Perez, Cristina
    Perez-Ruiz, Alberto
    Garrido-Castro, Juan L.
    Vidal, Francisco Torres
    Alcaraz-Clariana, Sandra
    Garcia-Luque, Lourdes
    Rodrigues-de-Souza, Daiana Priscila
    Alburquerque-Sendin, Francisco
    DIAGNOSTICS, 2020, 10 (09)
  • [24] A Nonproprietary Movement Analysis System (MoJoXlab) Based on Wearable Inertial Measurement Units Applicable to Healthy Participants and Those With Anterior Cruciate Ligament Reconstruction Across a Range of Complex Tasks: Validation Study
    Islam, Riasat
    Bennasar, Mohamed
    Nicholas, Kevin
    Button, Kate
    Holland, Simon
    Mulholland, Paul
    Price, Blaine
    Al-Amri, Mohammad
    JMIR MHEALTH AND UHEALTH, 2020, 8 (06):
  • [25] Restenosis of Drug-Eluting Stents A New Classification System Based on Disease Mechanism to Guide Treatment and State-of-the-Art Review
    Shlofmitz, Evan
    Iantorno, Micaela
    Waksman, Ron
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2019, 12 (08)
  • [26] Analysis of Inertial Measurement Unit Data for an AI-Based Physical Function Assessment System Using In-Clinic-like Movements
    Kouno, Nobuji
    Takahashi, Satoshi
    Takasawa, Ken
    Komatsu, Masaaki
    Ishiguro, Naoaki
    Takeda, Katsuji
    Matsuoka, Ayumu
    Fujimori, Maiko
    Yokoyama, Kazuki
    Yamamoto, Shun
    Honma, Yoshitaka
    Kato, Ken
    Obama, Kazutaka
    Hamamoto, Ryuji
    BIOENGINEERING-BASEL, 2024, 11 (12):
  • [27] Comparison of a New Inertial Sensor Based System with an Optoelectronic Motion Capture System for Motion Analysis of Healthy Human Wrist Joints
    Wirth, Michael Alexander
    Fischer, Gabriella
    Verdu, Jorge
    Reissner, Lisa
    Balocco, Simone
    Calcagni, Maurizio
    SENSORS, 2019, 19 (23)
  • [28] Using New Camera-Based Technologies for Gait Analysis in Older Adults in Comparison to the Established GAITRite System
    Steinert, Anika
    Sattler, Igor
    Otte, Karen
    Roehling, Hanna
    Mansow-Model, Sebastian
    Mueller-Werdan, Ursula
    SENSORS, 2020, 20 (01)