Validation of post-treatment PET-based dosimetry software for hepatic radioembolization of Yttrium-90 microspheres

被引:20
作者
Maughan, Nichole M. [1 ]
Garcia-Ramirez, Jose [1 ]
Arpidone, Matt [2 ]
Swallen, Amy [2 ]
Laforest, Richard [3 ]
Goddu, S. Murty [1 ]
Parikh, Parag J. [4 ]
Zoberi, Jacqueline E. [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, Sch Med, St Louis, MO 63110 USA
[2] MIM Software, Cleveland, OH 44122 USA
[3] Washington Univ, Dept Radiol, Sch Med, St Louis, MO 63110 USA
[4] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
PET; MRI; post-implant dosimetry; radioembolization; Y-90; DOSE-RESPONSE; HEPATOCELLULAR-CARCINOMA; 3-DIMENSIONAL DOSIMETRY; QUANTIFICATION; LOCALIZATION; ALBUMIN;
D O I
10.1002/mp.13444
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Yttrium-90 (Y-90) microsphere radioembolization enables selective internal radiotherapy for hepatic malignancies. Currently, there is no standard postdelivery imaging and dosimetry of the microsphere distribution to verify treatment. Recent studies have reported utilizing the small positron yield of Y-90 (32 ppm) with positron emission tomography (PET) to perform treatment verification and dosimetry analysis. In this study, we validated a commercial dosimetry software, MIM SurePlan (TM) LiverY90 (MIM Software Inc., Cleveland, OH), for clinical use. Methods A MATLAB-based algorithm for Y-90 PET-based dosimetry was developed in-house and validated for the purpose of commissioning the commercial software. The algorithm is based on voxel S values and dosimetry formalism reported in MIRD Pamphlet 17. We validated the in-house algorithm to establish it as the ground truth by comparing results from a digital point phantom and a digital uniform cylinder to manual calculations. Once we validated our in-house MATLAB-based algorithm, we used it to perform acceptance testing and commissioning of the commercial dosimetry software, MIM SurePlan, which uses the same dosimetry formalism. A 0.4 cm/5% gamma test was performed on PET-derived dose maps from each algorithm of uniform digital and nonuniform physical phantoms filled with Y-90 chloride solution. Average dose (D-avg) and minimum dose to 70% (D-70) of a given volume of interest (VOI) were compared for the digital phantom, the physical phantom, and five patient cases (27 tumor VOIs), representing different clinical scenarios. Results The gamma-pass rates were 97.26% and 97.66% for the digital and physical phantoms, respectively. The differences between D-avg and D-70 were 0.076% and 0.10% for the digital phantom, respectively, and <5.2% for various VOIs in the physical phantom. In the clinical cases, 96.3% of the VOIs had a difference <5% for D-avg, and 88.9% of the VOIs had a difference Dose calculation results from MIM SurePlan were found to be in good agreement with our in-house algorithm. This indicates that MIM SurePlan performs as it should and, hence, can be deemed accepted and commissioned for clinical use for post-implant PET-based dosimetry of Y-90 radioembolization. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:2394 / 2402
页数:9
相关论文
共 28 条
  • [1] Dose calculations for external photon beams in radiotherapy
    Ahnesjö, A
    Aspradakis, MM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) : R99 - R155
  • [2] Characterization of 90Y-SPECT/CT self-calibration approaches on the quantification of voxel-level absorbed doses following 90Y-microsphere selective internal radiation therapy
    Balagopal, Anjali
    Kappadath, S. Cheenu
    [J]. MEDICAL PHYSICS, 2018, 45 (02) : 875 - 883
  • [3] Bolch WE, 1999, J NUCL MED, V40, p11S
  • [4] Prospective Trial Using Internal Pair-Production Positron Emission Tomography to Establish the Yttrium-90 Radioembolization Dose Required for Response of Hepatocellular Carcinoma
    Chan, Keith T.
    Alessio, Adam M.
    Johnson, Guy E.
    Vaidya, Sandeep
    Kwan, Sharon W.
    Monsky, Wayne
    Wilson, Ann E.
    Lewis, David H.
    Padia, Siddharth A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 101 (02): : 358 - 365
  • [5] PET/MRI of Hepatic 90Y Microsphere Deposition Determines Individual Tumor Response
    Fowler, Kathryn J.
    Maughan, Nichole M.
    Laforest, Richard
    Saad, Nael E.
    Sharma, Akash
    Olsen, Jeffrey
    Speirs, Christina K.
    Parikh, Parag J.
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2016, 39 (06) : 855 - 864
  • [6] Internal Pair Production of 90Y Permits Hepatic Localization of Microspheres Using Routine PET: Proof of Concept
    Gates, Vanessa L.
    Esmail, Abdulredha A. H.
    Marshall, Karen
    Spies, Stewart
    Salem, Riad
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (01) : 72 - 76
  • [7] Correlation of Technetium-99m Macroaggregated Albumin and Yttrium-90 Glass Microsphere Biodistribution in Hepatocellular Carcinoma: A Retrospective Review of Pretreatment Single Photon Emission CT and Posttreatment Positron Emission Tomography/CT
    Haste, Paul
    Tann, Mark
    Persohn, Scott
    LaRoche, Thomas
    Aaron, Vasantha
    Mauxion, Thibault
    Chauhan, Nikhil
    Dreher, Matthew R.
    Johnson, Matthew S.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 28 (05) : 722 - 730
  • [8] Post-radioembolization yttrium-90 PET/CT - part 2: dose-response and tumor predictive dosimetry for resin microspheres
    Kao, Yung-Hsiang
    Steinberg, Jeffrey D.
    Tay, Young-Soon
    Lim, Gabriel K. Y.
    Yan, Jianhua
    Townsend, David W.
    Budgeon, Charley A.
    Boucek, Jan A.
    Francis, Roslyn J.
    Cheo, Timothy S. T.
    Burgmans, Mark C.
    Irani, Farah G.
    Lo, Richard H. G.
    Tay, Kiang-Hiong
    Tan, Bien-Soo
    Chow, Pierce K. H.
    Satchithanantham, Somanesan
    Tan, Andrew E. H.
    Ng, David C. E.
    Goh, Anthony S. W.
    [J]. EJNMMI RESEARCH, 2013, 3 : 1 - 27
  • [9] Post-radioembolization yttrium-90 PET/CT - part 1: diagnostic reporting
    Kao, Yung-Hsiang
    Steinberg, Jeffrey D.
    Tay, Young-Soon
    Lim, Gabriel K. Y.
    Yan, Jianhua
    Townsend, David W.
    Takano, Angela
    Burgmans, Mark C.
    Irani, Farah G.
    Teo, Terence K. B.
    Yeow, Tow-Non
    Gogna, Apoorva
    Lo, Richard H. G.
    Tay, Kiang-Hiong
    Tan, Bien-Soo
    Chow, Pierce K. H.
    Satchithanantham, Somanesan
    Tan, Andrew E. H.
    Ng, David C. E.
    Goh, Anthony S. W.
    [J]. EJNMMI RESEARCH, 2013, 3 : 1 - 28
  • [10] Hepatocellular Carcinoma Tumor Dose Response After 90Y-radioembolization With Glass Microspheres Using 90Y-SPECT/CT-Based Voxel Dosimetry
    Kappadath, S. Cheenu
    Mikell, Justin
    Balagopal, Anjali
    Baladandayuthapani, Veera
    Kaseb, Ahmed
    Mahvash, Armeen