Evaluation of capacitive micromachined ultrasonic transducers for passive monitoring of microbubble-assisted ultrasound therapies

被引:8
|
作者
Dauba, Ambre [1 ]
Goulas, Jordane [1 ]
Colin, Laurent [2 ]
Jourdain, Laurene [1 ]
Larrat, Benoit [3 ]
Gennisson, Jean-Luc [1 ]
Certon, Dominique [2 ]
Novell, Anthony [1 ]
机构
[1] Univ Paris Saclay, Serv Hosp Frederic Joliot, INSERM, CEA,CNRS,BioMaps, F-91401 Orsay, France
[2] Univ Tours, INSA Ctr Val Loire, GREMAN CNRS UMR 7347, Tours, France
[3] Univ Paris Saclay, NeuroSpin, CEA, CNRS,Baobab, F-91191 Gif Sur Yvette, France
来源
关键词
BLOOD-BRAIN-BARRIER; FOCUSED ULTRASOUND; CAVITATION; DISRUPTION; PROBE;
D O I
10.1121/10.0002096
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Passive cavitation detection can be performed to monitor microbubble activity during brain therapy. Microbubbles under ultrasound exposure generate a response characterized by multiple nonlinear emissions. Here, the wide bandwidth of capacitive micromachined ultrasonic transducers (CMUTs) was exploited to monitor the microbubble signature through a rat skull and a macaque skull. The intrinsic nonlinearity of the CMUTs was characterized in receive mode. Indeed, undesirable nonlinear components generated by the CMUTs must be minimized as they can mask the microbubble harmonic response. The microbubble signature at harmonic and ultra-harmonic components (0.5-6 MHz) was successfully extracted through a rat skull using moderate bias voltage.
引用
收藏
页码:2248 / 2255
页数:8
相关论文
共 50 条
  • [21] Acoustic lens for capacitive micromachined ultrasonic transducers
    Chang, Chienliu
    Firouzi, Kamyar
    Park, Kwan Kyu
    Sarioglu, Ali Fatih
    Nikoozadeh, Amin
    Yoon, Hyo-Seon
    Vaithilingam, Srikant
    Carver, Thomas
    Khuri-Yakub, Butrus T.
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2014, 24 (08)
  • [22] An accurate model for capacitive micromachined ultrasonic transducers
    Caronti, A
    Caliano, G
    Iula, A
    Pappalardo, M
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (02) : 159 - 168
  • [23] Experimental characterization of capacitive micromachined ultrasonic transducers
    Olcum, Selim
    Atalar, Abdullah
    Koymen, Hayerttin
    Oguz, Kagan
    Senlik, Muhammed N.
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 2131 - 2134
  • [24] Capacitive Micromachined Ultrasonic Transducers: Technology and Application
    Salim, Muhammed Sabri
    Malek, M. F. Abd
    Heng, R. B. W.
    Juni, K. M.
    Sabri, Naseer
    JOURNAL OF MEDICAL ULTRASOUND, 2012, 20 (01) : 8 - 31
  • [25] Capacitive micromachined ultrasonic transducers: Theory and technology
    Ergun, AS
    Yaralioglu, GG
    Khuri-Yakub, BT
    JOURNAL OF AEROSPACE ENGINEERING, 2003, 16 (02) : 76 - 84
  • [26] Surface micromachined capacitive ultrasonic immersion transducers
    Jin, XC
    Ladabaum, I
    Khuri-Yakub, BT
    MICRO ELECTRO MECHANICAL SYSTEMS - IEEE ELEVENTH ANNUAL INTERNATIONAL WORKSHOP PROCEEDINGS, 1998, : 649 - 654
  • [27] Rotational Capacitive Micromachined Ultrasonic Transducers (cMUTs)
    Kuntzman, Michael L.
    Hall, Neal A.
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2014, 23 (01) : 1 - 3
  • [28] The characterization of capacitive micromachined ultrasonic transducers in air
    McIntosh, JS
    Hutchins, DA
    Billson, DR
    Robertson, TJ
    Noble, RA
    Jones, ADR
    ULTRASONICS, 2002, 40 (1-8) : 477 - 483
  • [29] Capacitive Micromachined Ultrasonic Transducers for Medical Application
    Li, Yan
    Zhang, Pei-yu
    2015 INTERNATIONAL CONFERENCE ON ELECTRICAL AND ELECTRONICS: TECHNIQUES AND APPLICATIONS (EETA 2015), 2015, : 111 - 116
  • [30] A new class of capacitive micromachined ultrasonic transducers
    Ahrens, O
    Hohlfeld, D
    Buhrdorf, A
    Glitza, O
    Binder, J
    2000 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2000, : 939 - 942