The most inclusive parameter describing the performance of an imaging detector may be considered detective quantum efficiency (DQE) expressed in function of the spatial frequencies. In this paper our effort is addressed to formulating a general and absolute expression of the DQE in the frequency domain, coherent with the condition DQE less than or equal to QE. The case of a flat-field input image and of integrating detectors is also considered. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:163 / 167
页数:5
相关论文
共 13 条
[1]  
[Anonymous], 2006, J ICRU
[2]  
Barrett HH., 1981, RADIOLOGICAL IMAGING
[3]   DQE measurement in a scintillating glass optical fiber detector for X-ray imaging [J].
Costa, S ;
Ottonello, P ;
Rottigni, GA ;
Zanella, G ;
Zannoni, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (03) :568-571
[4]  
Dainty JC., 1974, IMAGE SCI
[5]  
Jones R. C., 1959, ADVAN ELECTRONICS EL, V11, P87
[6]   ANALYSIS OF THE SPATIAL-FREQUENCY-DEPENDENT DQE OF OPTICALLY COUPLED DIGITAL MAMMOGRAPHY DETECTORS [J].
MAIDMENT, ADA ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (06) :721-729
[7]   The concept of spatial frequency depending DQE and its application to a comparison of two detectors used in transvenous coronary angiography [J].
Menk, RH ;
Thomlinson, W ;
Gmur, N ;
Zhong, Z ;
Chapman, D ;
Arfelli, F ;
Dix, WR ;
Graeff, W ;
Lohmann, M ;
Illing, G ;
Schildwachter, L ;
Reime, B ;
Kupper, W ;
Hamm, C ;
Giacomini, JC ;
Gordon, HJ ;
Rubenstein, E ;
Dervan, J ;
Besch, HJ ;
Walenta, AH .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 398 (2-3) :351-367
[8]   TOWARD CONSENSUS ON QUANTITATIVE ASSESSMENT OF MEDICAL IMAGING-SYSTEMS [J].
METZ, CE ;
WAGNER, RF ;
DOI, K ;
BROWN, DG ;
NISHIKAWA, RM ;
MYERS, KJ .
MEDICAL PHYSICS, 1995, 22 (07) :1057-1061
[9]  
Rose A, 1946, J SOC MOTION PIC ENG, V47, P273
[10]   RELATIVE EFFICIENCIES OF ENERGY TO PHOTOGRAPHIC DENSITY CONVERSIONS IN TYPICAL SCREEN-FILM SYSTEMS [J].
VYBORNY, CJ ;
METZ, CE ;
DOI, K .
RADIOLOGY, 1980, 136 (02) :465-471