Midterm Results of Fourth-Generation Ceramic-On-Ceramic Total Hip Arthroplasty

被引:2
作者
Cho, Kyu-Jin [1 ]
Park, Kyung-Soon [1 ]
Jong, Won-Young [2 ]
Park, Chun-Gon [2 ]
Yoon, Taek-Rim [1 ]
机构
[1] Chonnam Natl Univ, Ctr Joint Dis, Hwasun Hosp, 160 Ilsimri, Jeonnam 519809, South Korea
[2] Chonnam Natl Univ Hosp, Orthopaed Dept, Gwangju, South Korea
关键词
Ceramic-on-ceramic; delta ceramic; liner fracture; midterm result; total hip arthroplasty; REVISION; HEAD; BEARINGS; FAILURE; FRACTURES; ARTHRITIS; OUTCOMES;
D O I
10.4103/ortho.IJOrtho_192_18
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: There are limited reports for the results of the fourth-generation ceramic-on-ceramic (CoC) articulation total hip arthroplasty (THA). And, throughout the surgical experience, we encountered some cases of liner pulling-out phenomenon after liner fixation and femoral preparation. The objective of this study was to evaluate the incidence, risk factors of delta ceramic liner or head fractures, and also the clinical and radiological results of using the fourth-generation CoC articulation in THA. Patients and Methods: We retrospectively reviewed 242 patients (263 hips) who underwent primary THA using the fourth-generation CoC articulation with a minimum followup of 2 years. Demographic data, Harris Hip Score (HHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Patient's satisfactory level were recorded. The radiological evaluation was used to evaluate the implant fixation and complications. Mean followup duration was 5.2 years. Results: Mean HHS and WOMAC score were significantly (P < 0.05) improved at the last followup. About 98.5% of the patients were satisfied with results of the surgery. All acetabular components were placed in adequate position and there was no osteolysis on acetabular or femoral components and subsidence of femoral stem. Four patients showed complications including one-liner fracture. Conclusion: Our midterm study demonstrated excellent clinical and radiological results with only one ceramic liner fracture. Moreover, the results of this study indicate that one possible cause of pulling-out phenomenon is the resonance effect during implantation in Dorr type A patients with the thick cortex. If the surgeon is aware of the liner malposition throughout the operation, the fourth-generation CoC articulation THA could be an outstanding treatment.
引用
收藏
页码:630 / 636
页数:7
相关论文
共 27 条
  • [1] Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head - A case report
    Allain, J
    Goutallier, D
    Voisin, MC
    Lemouel, S
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) : 1355 - 1360
  • [2] Midterm Outcomes of the Recently FDA Approved Ceramic on Ceramic Bearing in Total Hip Arthroplasty Patients Under 65 Years of Age
    Aoude, Ahmed A.
    Antoniou, John
    Epure, Laura M.
    Huk, Olga L.
    Zukor, David J.
    Tanzer, Michael
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (08) : 1388 - 1392
  • [3] BELLAMY N, 1988, J RHEUMATOL, V15, P1833
  • [4] Ceramic-on-ceramic bearings in total hip arthroplasty
    Bierbaum, BE
    Nairus, J
    Kuesis, D
    Morrison, JC
    Ward, D
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) : 158 - 163
  • [5] RESONANCE, TACOMA NARROWS BRIDGE FAILURE, AND UNDERGRADUATE PHYSICS TEXTBOOKS
    BILLAH, KY
    SCANLAN, RH
    [J]. AMERICAN JOURNAL OF PHYSICS, 1991, 59 (02) : 118 - 124
  • [6] Third-Generation Ceramic-On-Ceramic Total Hip Arthroplasty in Patients Younger Than 30 Years with Osteonecrosis of Femoral Head
    Byun, Jae-Wook
    Yoon, Taek-Rim
    Park, Kyung-Soon
    Seon, Jong-Keun
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (07) : 1337 - 1343
  • [7] Large-diameter Delta Ceramic-on-ceramic Versus Common-sized Ceramic-on-polyethylene Bearings in THA
    Cai, Pengde
    Hu, Yihe
    Xie, Jie
    [J]. ORTHOPEDICS, 2012, 35 (09) : E1307 - E1313
  • [8] RESULTS OF REVISION FOR MECHANICAL FAILURE AFTER CEMENTED TOTAL HIP-REPLACEMENT, 1979 TO 1982 - A 2 TO 5-YEAR FOLLOW-UP
    CALLAGHAN, JJ
    SALVATI, EA
    PELLICCI, PM
    WILSON, PD
    RANAWAT, CS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (07) : 1074 - 1085
  • [9] Capello William N, 2005, Instr Course Lect, V54, P171
  • [10] Ceramic Bearings for Total Hip Arthroplasty Have High Survivorship at 10 Years
    D'Antonio, James A.
    Capello, William N.
    Naughton, Marybeth
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (02) : 373 - 381