High resolution fast digital photoacoustic CT for breast cancer diagnosis

被引:8
|
作者
Xiang Liang-Zhong
Xing Da [1 ]
Guo Hua
Yang Si-Hua
机构
[1] S China Normal Univ, MOE Key Lab Laser Life Sci, Guangzhou 510631, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
photoacoustic computer tomography; modified LFBP algorithm; focusing by acoustical lens; breast cancer detection; IN-VIVO; BLOOD; TOMOGRAPHY;
D O I
10.7498/aps.58.4610
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Photoacoustic imaging combines advantages of both ultrasound imaging and optical imaging to provide high ultrasonic resolution and high optical contrast in noninvasive medical images. A photoacoustic computer tomography (PCT) system with a focused digital phased array transducer is proposed. In particular each element of the transducer array has a thin cylindrical ultrasonic lens to select 2D image plane and suppress the out - of - plane signals to realize photoacoustic computed tomography. A fast full - view photoacoustic imaging system for visualizing tissue structures using a linear transducer array with combined scan, whies; detects photoacoustic signals by combined electronic and mechanical scanning. An improved limited - field filtered back projection algorithm with directivity factors was applied to reconstruct the optical absorption distribution and further improve the lateral resolution. In the experiment, the spatial resolution of this imaging system is determined as 0.2 mm, and the slice resolution of Z axis is 1.5 mm. Mechanical step scanning was performed, leading to a measurement time of 6.4 seconds per position, and a total scan duration of 150 seconds to obtain the two-dimensional photoacoustic image. High spatial resolution of 2D photoacoustic vascular image and quasi-3D photoacoustic CT for breast cancer imaging are achicned. The experimental results demonstrate that this method and imaging system can realize the high resolution photoacoustic tomography wtich can be applied to noninvasive detection of early breast cancer.
引用
收藏
页码:4610 / 4617
页数:8
相关论文
共 18 条
  • [1] Photoacoustic tomography imaging based on an acoustic lens imaging system
    Chen Zhan-Xu
    Tang Zhi-Lie
    Wan Wei
    He Yong-Heng
    [J]. ACTA PHYSICA SINICA, 2006, 55 (08) : 4365 - 4370
  • [2] EMELIANOV SY, 2006, P IEEE ULTRASONICS S, V405
  • [3] FOCUSING PROPERTIES OF SOLID ULTRASONIC CYLINDRICAL LENSES
    FOLDS, DL
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1973, 53 (03): : 826 - 834
  • [4] Gu HM, 2006, PROG BIOCHEM BIOPHYS, V33, P431
  • [5] Imaging of tumor angiogenesis in rat brains in vivo by photoacoustic tomography
    Ku, G
    Wang, XD
    Xie, XY
    Stoica, G
    Wang, LHV
    [J]. APPLIED OPTICS, 2005, 44 (05) : 770 - 775
  • [6] Optoacoustic imaging of blood for visualization and diagnostics of breast cancer
    Oraevsky, AA
    Savateeva, EV
    Solomatin, SV
    Karabutov, AA
    Andreev, VG
    Gatalica, Z
    Khamapirad, T
    Henrichs, PM
    [J]. BIOMEDICAL OPTOACOUSTICS III, 2002, 4618 : 81 - 94
  • [7] Su YX, 2006, CHINESE PHYS LETT, V23, P512, DOI 10.1088/0256-307X/23/2/066
  • [8] Suzuki K, 1996, J Biomed Opt, V1, P330, DOI 10.1117/12.239902
  • [9] Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain
    Wang, XD
    Pang, YJ
    Ku, G
    Xie, XY
    Stoica, G
    Wang, LHV
    [J]. NATURE BIOTECHNOLOGY, 2003, 21 (07) : 803 - 806
  • [10] Photoacoustic imaging with deconvolution algorithm
    Wang, Y
    Xing, D
    Zeng, YG
    Chen, Q
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14): : 3117 - 3124