Selective In Vivo Bone Imaging With Long-T2 Suppressed PETRA MRI

被引:17
作者
Li, Cheng [1 ]
Magland, Jeremy F. [1 ]
Zhao, Xia [1 ]
Seifert, Alan C. [1 ]
Wehrli, Felix W. [1 ]
机构
[1] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
基金
美国国家卫生研究院;
关键词
long-T-2; suppression; inversion recovery; short T-2; phase-modulated RF excitation; image reconstruction; ultrashort echo time; zero echo time; CORTICAL BONE; MAGNETIC-RESONANCE; FIELD STRENGTHS; PORE-WATER; RECONSTRUCTION; EXCITATION; DENSITY; NMR;
D O I
10.1002/mrm.26178
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To design and evaluate an optimized PETRA (point-wise encoding time reduction with radial acquisition) sequence with long-T-2 suppression at 3 Tesla. Methods: An adiabatic inversion recovery-based scheme was used to null the long-T-2 signal. To minimize scan time, the signal was sampled multiple times after each inversion with variable excitation flip angles designed to yield constant short-T-2 signal amplitude. The excitation pulses were phase-modulated, allowing for increased flip angle and higher signal-to-noise ratio (SNR). A fast, noniterative image reconstruction algorithm was designed to minimize image artifacts due to nonuniform excitation profile. Results: Phase-modulated pulse excitation, along with the noniterative reconstruction algorithm, allows the use of larger radiofrequency pulse flip angles, resulting in effective suppression of long-T-2 protons and improved image SNR without causing image artifacts. Midtibia images representative of collagen-bound water yielded SNR of 15 at 1-mm isotropic resolution in 6.5 minutes with a standard extremity coil. Further, the technology is shown to be suited for generating multi-angle projection images of bone akin to X-ray images displaying subtle anatomic detail. Conclusion: Optimized long-T-2 suppressed PETRA allows imaging of bone matrix water unencumbered by long-T-2 soft tissue and pore water protons, opening up new possibilities for anatomic bone imaging at isotropic resolution and quantification in clinically practical scan times. (C) 2016 International Society for Magnetic Resonance in Medicine.
引用
收藏
页码:989 / 997
页数:9
相关论文
共 50 条
  • [21] Craniofacial Imaging of Pediatric Patients by Ultrashort Echo-Time Bone- Selective MRI in Comparison to CT
    Kamona, Nada
    Ng, Jinggang J.
    Kim, Yohan
    Vu, Brian-Tinh D.
    Vossough, Arastoo
    Wagner, Connor S.
    Cordray, Holly
    Lee, Hyunyeol
    Villavisanis, Dillan F.
    Rajapakse, Chamith S.
    Bartlett, Scott P.
    Wehrli, Felix W.
    [J]. ACADEMIC RADIOLOGY, 2024, 31 (11) : 4629 - 4642
  • [22] The effect of excitation and preparation pulses on nonslice selective 2D UTE bicomponent analysis of bound and free water in cortical bone at 3T
    Li, Shihong
    Chang, Eric Y.
    Bae, Won C.
    Chung, Christine B.
    Hua, Yanqing
    Zhou, Yi
    Du, Jiang
    [J]. MEDICAL PHYSICS, 2014, 41 (02)
  • [23] MRI of the Hip at 7T: Feasibility of Bone Microarchitecture, High-Resolution Cartilage, and Clinical Imaging
    Chang, Gregory
    Deniz, Cem M.
    Honig, Stephen
    Egol, Kenneth
    Regatte, Ravinder R.
    Zhu, Yudong
    Sodickson, Daniel K.
    Brown, Ryan
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (06) : 1384 - 1393
  • [24] Identification of ultrasound imaging markers to quantify long bone regeneration in a segmental tibial defect sheep model in vivo
    Tang, Songyuan
    Shajudeen, Peer
    Tasciotti, Ennio
    Righetti, Raffaella
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [25] Simultaneous time of flight-MRA and T2*imaging for cerebrovascular MRI
    Lanzman, Bryan A.
    Huang, Yuhao
    Lee, Edward H.
    Iv, Michael
    Moseley, Michael E.
    Holdsworth, Samantha J.
    Yeom, Kristen W.
    [J]. NEURORADIOLOGY, 2021, 63 (02) : 243 - 251
  • [26] Quantitative Assessment of MRI T2 Response to Kainic Acid Neurotoxicity in Rats in vivo
    Liachenko, Serguei
    Ramu, Jaivijay
    Konak, Tetyana
    Paule, Merle G.
    Hanig, Joseph
    [J]. TOXICOLOGICAL SCIENCES, 2015, 146 (01) : 183 - 191
  • [27] Nearly automatic segmentation of hippocampal subfields in in vivo focal T2-weighted MRI
    Yushkevich, Paul A.
    Wang, Hongzhi
    Pluta, John
    Das, Sandhitsu R.
    Craige, Caryne
    Avants, Brian B.
    Weiner, Michael W.
    Mueller, Susanne
    [J]. NEUROIMAGE, 2010, 53 (04) : 1208 - 1224
  • [28] Ultrashort TE MR Imaging of Bovine Cortical Bone: The Effect of Water Loss on the T1 and T2* Relaxation Times
    Kokabi, Nima
    Bae, Won
    Diaz, Eric
    Chung, Christine B.
    Bydder, Graeme M.
    Du, Jiang
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (02) : 476 - 482
  • [29] In vivo MRI of submillisecond T2 species with two-dimensional and three-dimensional radial sequences and applications to the measurement of cortical bone water
    Techawiboonwong, Aranee
    Song, Hee Kwon
    Wehrli, Felix W.
    [J]. NMR IN BIOMEDICINE, 2008, 21 (01) : 59 - 70
  • [30] Artifact-Suppressed Optimal Three-Dimensional T1- and T2*-Weighted Dual-Echo Imaging
    Do, Won-Joon
    Kim, Ki Hwan
    Choi, Seung Hong
    Park, Sung-Hong
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (05) : 1504 - 1511