Determination of thiamine in blood serum and urine by high-performance liquid chromatography with direct injection and post-column derivatization

被引:25
|
作者
Bohrer, D [1 ]
do Nascimento, PC
Ramirez, AG
Mendonça, JKA
de Carvalho, LM
Pomblum, SCG
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
[2] Univ Fed Santa Maria, Dept Anal Clin & Toxicol, BR-97105900 Santa Maria, RS, Brazil
关键词
SPE; polyethylene; serum; urine; vitamin B-1;
D O I
10.1016/j.microc.2004.03.013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Thiamine (vitamin B1) was determined in human serum and urine by HPLC with fluorimetric detection of its oxidation product, thiochrome. The samples were injected directly into the chromatographic system without previous treatment or dilution. A column filled with an ultra-high molecular weight surface-modified polyethylene (PE) was able to separate matrix components from analyte and also to allow a good chromatographic resolution of thiamine. The interaction of thiamine, thiocrome and both matrices (serum and urine) with PE was studied off- and on-line to determine the optimal procedure for vitamin B1 determination. When carried off-line, matrix adsorption yield was 49 mg serum proteins/g polymer and components of 1000 mul urine/g polymer. In an on-line arrangement, the yield dropped to 10 mg/g and 150 mul/g, respectively. The matrix/analyte separation was carried out in an on-line procedure on a 50x4.6-mm, 25-mum PE column, using a water-sodium phosphate-methanol gradient elution. Part of the matrix was eluted within the first 2 min and thiamine after 3.8 min. The rest of the matrix retained on the column was eluted after thiamine at the last step of the gradient elution. Analysis time was 12 min. The within-day and day-to-day precision gave C.V. varying from 3.6% to 14.5% and recoveries from spiked samples were in the range of 84.8-98.8%. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:71 / 76
页数:6
相关论文
共 50 条