The effect of five mechanical instrumentation protocols on implant surface topography and roughness: A scanning electron microscope and confocal laser scanning microscope analysis

被引:49
|
作者
Cha, Jae-Kook [1 ]
Paeng, Kyeongwon [1 ]
Jung, Ui-Won [1 ]
Choi, Seong-Ho [1 ]
Sanz, Mariano [2 ,3 ]
Sanz-Martin, Ignacio [2 ]
机构
[1] Yonsei Univ, Dept Periodontol, Coll Dent, Res Inst Periodontal Regenerat, 50-1 Yonsei Ro, Seoul 120752, South Korea
[2] Univ Complutense Madrid, Fac Odontol, Sect Grad Periodontol, Madrid, Spain
[3] Univ Complutense Madrid, ETEP Etiol & Therapy Periodontal Dis Res Grp, Madrid, Spain
关键词
confocal microscope; dental implants; peri-implantitis; scanning electron microscope; surface decontamination; surface macrotopography; titanium; TITANIUM DENTAL IMPLANTS; ULTRASONIC SCALER TIPS; HUMAN GINGIVAL FIBROBLASTS; PERI-IMPLANTITIS; PROLIFERATION; ATTACHMENT; DECONTAMINATION; DEBRIDEMENT; VIABILITY; REMOVAL;
D O I
10.1111/clr.13446
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective To evaluate in vitro the changes in implant surface topography and roughness of commercial implants after instrumentation with five decontamination protocols. Material and methods Seventy-two titanium implants with a sandblasted and acid-etched (SLA) surface were placed 5 mm supra-crestally. Five groups of twelve implants were instrumented with the following protocols: a metal scaler tip (SCAL), a thermoplastic scaler tip (PEEK), a round titanium brush (RBRU), a tufted brush with titanium bristles (TNBRU), and a glycine-based air-powder abrasive (GLYC). A sixth group with untreated implants was used as control. Scanning electron microscope and confocal laser scanning microscope were utilized to evaluate the changes in the implant surfaces. Results The SCAL caused pronounced macroscopic alterations and damage of the implant surface, the PEEK left remnants of the plastic tip in the implant surface, and both titanium brush groups flattened the thread profile, while minimal alterations were observed in the GLYC. When compared to the control group, the roughness parameters (Sa) in the buccal aspect increased in the thread area of SCAL, and a minor reduction was observed in the PEEK while in the other groups, these values remained unchanged. In the valley areas, however the RBRU, TNBRU, and GLYC experienced a significant reduction (smoothening) indicating different accessibility of the decontamination protocols to the thread and valley. Similarly, the buccal aspects had more pronounced changes than those in the palatal aspect. Conclusion Within the limitations of this in vitro investigation, the tested protocols induced different macroscopic alterations and surface roughness changes that varied in the thread and valley area.
引用
收藏
页码:578 / 587
页数:10
相关论文
共 50 条
  • [31] Analysis of plant tissue images obtained by confocal tandem scanning reflected light microscope
    Gancarz, M.
    Konstankiewicz, K.
    Pawlak, K.
    Zdunek, A.
    INTERNATIONAL AGROPHYSICS, 2007, 21 (01) : 49 - 53
  • [32] Hollow implants retrieved for fracture: A light and scanning electron microscope analysis of 4 cases
    Piattelli, A
    Scarano, A
    Piattelli, M
    Vaia, E
    Matarasso, S
    JOURNAL OF PERIODONTOLOGY, 1998, 69 (02) : 185 - 189
  • [33] Area scanning method for 3D surface profilometry based on an adaptive confocal microscope
    Luo, Ding
    Taphanel, Miro
    Claus, Daniel
    Boettcher, Tobias
    Osten, Wolfgang
    Laengle, Thomas
    Beyerer, Juergen
    OPTICS AND LASERS IN ENGINEERING, 2020, 124
  • [34] Assessment of the Inner Surface Microstructure of Decellularized Cortical Bone by a Scanning Electron Microscope
    Jeong, Heonuk
    Asai, Jungo
    Ushida, Takashi
    Furukawa, Katsuko S.
    BIOENGINEERING-BASEL, 2019, 6 (03):
  • [35] Scanning-electron-microscope image processing for accurate analysis of line-edge and line-width roughness
    Hiraiwa, Atsushi
    Nishida, Akio
    METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XXVI, PTS 1 AND 2, 2012, 8324
  • [36] Scanning electron microscope charging effect model for chromium/quartz photolithography masks
    Seeger, Adam
    Duci, Alessandro
    Haussecker, Horst
    SCANNING, 2006, 28 (03) : 179 - 186
  • [37] Sidewall effect of photomask by scanning electron microscope and optical critical dimension metrology
    Yamane, T
    Hirano, T
    JOURNAL OF MICROLITHOGRAPHY MICROFABRICATION AND MICROSYSTEMS, 2005, 4 (03):
  • [38] 1.06 μm laser irradiation on high reflection coatings inside a scanning electron microscope
    Poulingue, M
    Dijon, J
    Garrec, P
    Lyan, P
    LASER-INDUCED DAMAGE IN OPTICAL MATERIALS: 1998, 1999, 3578 : 188 - 195
  • [39] Evaluation of Correlation between Dissolution Rates of Loxoprofen Tablets and Their Surface Morphology Observed by Scanning Electron Microscope and Atomic Force Microscope
    Yoshikawa, Shinichi
    Murata, Ryo
    Shida, Shigenari
    Uwai, Koji
    Suzuki, Tsuneyoshi
    Katsumata, Shunji
    Takeshita, Mitsuhiro
    CHEMICAL & PHARMACEUTICAL BULLETIN, 2010, 58 (01) : 34 - 37
  • [40] The Effects of Ultrasonic Scaling and Air-Abrasive Powders on the Topography of Implant Surfaces: Scanning Electron Analysis and In Vitro Study
    Gianfreda, Francesco
    Marenzi, Gaetano
    Nicolai, Eleonora
    Muzzi, Maurizio
    Bari, Monica
    Bernardini, Sergio
    Adamo, Daniela
    Miniello, Alessandra
    Sammartino, Gilberto
    Bollero, Patrizio
    EUROPEAN JOURNAL OF DENTISTRY, 2024,