Biomechanical evaluation of the interfacial strength of a chemically modified sandblasted and acid-etched titanium surface

被引:169
作者
Ferguson, S. J. [1 ]
Broggini, N.
Wieland, M.
de Wild, M.
Rupp, F.
Geis-Gerstorfer, J.
Cochran, D. L.
Buser, D.
机构
[1] Univ Bern, MEM Res Ctr Orthopaed Surg, Inst Surg Technol & Biomech, CH-3012 Bern, Switzerland
[2] Univ Bern, Sch Dent Med, Dept Oral Surg & Stomatol, CH-3012 Bern, Switzerland
[3] Univ Tubingen Hosp, Dept Prosthet Dent, Sect Med Mat & Technol, Tubingen, Germany
[4] Univ Texas, Hlth Sci Ctr, Dept Periodont, San Antonio, TX 78284 USA
关键词
dental implant; osseointegration; biomechanical testing; surface chemistry; interface; wettability;
D O I
10.1002/jbm.a.30678
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The functional capacity of osseointegrated dental implants to bear load is largely dependent on the quality of the interface between the bone and implant. Sandblasted and acid-etched (SLA) surfaces have been previously shown to enhance bone apposition. In this study, the SLA has been compared with a chemically modified SLA (modSLA) surface. The increased wettability of the modSLA surface in a protein solution was verified by dynamic contact angle analysis. Using a well-established animal model with a split-mouth experimental design, implant removal torque testing was performed to determine the biomechanical properties of the bone-implant interface. All implants had an identical cylindrical shape with a standard thread configuration. Removal torque testing was performed after 2, 4, and 8 weeks of bone healing (n = 9 animals per healing period, three implants per surface type per animal) to evaluate the interfacial shear strength of each surface type. Results showed that the modSLA surface was more effective in enhancing the interfacial shear strength of implants in comparison with the conventional SLA surface during early stages of bone healing. Removal torque values of the modSLA-surfaced implants were 8-21% higher than those of the SLA implants (p = 0.003). The mean removal torque values for the modSLA implants were 1.485 N m at 2 weeks, 1.709 N m at 4 weeks, and 1.345 N m at 8 weeks; and correspondingly, 1.231 N m, 1.585 N m, and 1.143 N m for the SLA implants. The bone-implant interfacial stiffness calculated from the torque-rotation curve was on average 9-14% higher for the modSLA implants when compared with the SLA implants (p = 0.038). It can be concluded that the modSLA surface achieves a better bone anchorage during early stages of bone healing than the SLA surface; chemical modification of the standard SLA surface likely enhances bone apposition and this has a beneficial effect on the interfacial shear strength. (c) 2006 Wiley Periodicals, Inc.
引用
收藏
页码:291 / 297
页数:7
相关论文
共 41 条
[1]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[2]  
[Anonymous], BONE ENG
[3]  
[Anonymous], 2001, ENG MAT SER
[4]  
[Anonymous], 1996, INTERFACIAL PHENOMEN
[5]  
[Anonymous], 2000, BONE ENG
[6]  
Bornstein MM, 2003, INT J ORAL MAX IMPL, V18, P659
[7]   Early loading of non-submerged titanium implants with a sandblasted and acid-etched surface [J].
Bornstein, MM ;
Schmid, B ;
Belser, UC ;
Lussi, A ;
Buser, D .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (06) :631-638
[8]  
Buser D, 1998, INT J ORAL MAX IMPL, V13, P611
[9]  
Buser D, 1999, J BIOMED MATER RES, V45, P75, DOI 10.1002/(SICI)1097-4636(199905)45:2<75::AID-JBM1>3.0.CO
[10]  
2-P